  "results": [
    {
      "status": "Terminated",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88546",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of ID, MA, MI, TX, VA.",
      "recall_number": "Z-2475-2021",
      "product_description": "The Allura Xper series is intended for use on human patients to perform:  \"Vascular, cardiovascular and neurovascular imaging applications, including diagnostic,  interventional and minimally invasive procedures.  This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs,  stent placements, embolisations and thrombolysis.\"  \"Cardiac imaging applications including diagnostics, interventional and minimally invasive  procedures (such as PTCA, stent placing, atherectomies), pacemaker implanatations, and  electrophysiology (EP).  Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures.\"",
      "product_quantity": "1 unit",
      "reason_for_recall": "Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU.     The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails.    When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.",
      "recall_initiation_date": "20210813",
      "center_classification_date": "20210914",
      "termination_date": "20240311",
      "report_date": "20210922",
      "code_info": "Serial number: 7",
      "more_code_info": ""
    }
  ]
}