  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83410",
      "recalling_firm": "Alere San Diego, Inc.",
      "address_1": "9975 Summers Ridge Rd",
      "address_2": "N/A",
      "postal_code": "92121-2997",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution in the states of AK, A, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NY< OH, OR, PA, RI, SC, SD, TN, TX, VA, VT, WA, WI.    No OUS consignees",
      "recall_number": "Z-2475-2019",
      "product_description": "Alere Cholestech LDX Analyzer, model # 14-874 containing power supplies with the identifying codes of 1837d and 1843d - Product Usage: The Alere Cholestech LDX System is a small, portable analyzer and test cassette system. The System is for in vitro diagnostic use only. The Lipid Profile GLU Cassette is for the quantitative determination of total cholesterol, HDL (high-density lipoprotein) cholesterol, triglycerides and glucose in whole blood. A TC/HDL (total cholesterol/HDL cholesterol) ratio and estimated values for LDL (low-density lipoprotein) and non-HDL cholesterol are calculated by the Alere Cholestech LDX¿ Analyzer.",
      "product_quantity": "186 analyzer kits",
      "reason_for_recall": "The polarity of the output of  power supply was reversed.  Due to polarity of the output of power supplies being reversed, it renders the power supplies nonfunctional and the analyzer inoperable.",
      "recall_initiation_date": "20190215",
      "center_classification_date": "20190905",
      "termination_date": "20231212",
      "report_date": "20190911",
      "code_info": "Lot/Unit Codes:  Lot codes of Meter Kits: 418197, 418198, 418199, 418252, & 418253  Lot codes of Power Supplies: 724599, 724604, & 725914    Serial Numbers: AA184455, AA184344, AA184328, AA184388, AA184304, AA184349,  AA184396, AA184339, AA184421, AA184280, AA184435, AA184427,  AA184395, AA183401, AA184394, AA184710, AA184345, AA184039,  AA184463, AA184609, AA184383, AA184415, AA184094, AA184070,  AA184412, AA183861, AA184278, AA184283, AA184711, AA184392,  AA184343, AA184368, AA184708, AA184350, AA184674, AA184378,  AA184433, AA184366, AA184365, AA184354, AA184369, AA184067,  AA184353, AA184664, AA184279, AA184355, AA184270, AA184285,  AA184017, AA184272, AA184340, AA184346, AA184347, AA184336,  AA184342, AA184269, AA184376, AA184949, AA184386, AA184351,  AA184341, AA184413, AA184298, AA183707, AA184595, AA184364,  AA184287, AA184618, AA184971, AA184334, AA184845, AA184267,  AA184296, AA184333, AA184608, AA184651, AA184319, AA184348,  AA184317, AA184699, AA184430, AA184432, AA184363, AA184030,  AA184338, AA184371, AA184375, AA184393, AA184306, AA184447,  AA183418, AA184323, AA184401, AA184438, AA184594, AA184281,  AA184461, AA184312, AA184321, AA184288, AA184357, AA184380,  AA184617, AA184352, AA184273, AA184303, AA184424, AA184290,  AA184300, AA184916, AA184965, AA184963, AA184980, AA184964,  AA184955, AA185069, AA185009, AA184904, AA184993, AA184973,  AA184991, AA184995, AA184900, AA185046, AA184951, AA185927,  AA184468, AA185050, AA184967, AA184853, AA184877, AA184830,  AA184961, AA184794, AA184930, AA184459, AA184887, AA184869,  AA185020, AA184785, AA184985, AA184901, AA184979, AA184990,  AA184865, AA184988, AA184872, AA184952, AA184946, AA184924,  AA182949, AA184918, AA184928, AA184886, AA184957, AA184974,  AA184862, AA185022, AA184657, AA184880, AA185021, AA184922,  AA184875, AA185013, AA184943, AA184873, AA184989, AA184742,  AA184876, AA184301, AA184367, AA184370, AA184359, AA184434,  AA184615, AA185004, AA184678, AA184329, AA184377, AA184326,  AA184284, AA184384, AA184795, AA184297, AA183549, AA184318"
    }
  ]
}