{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62110",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.",
      "recall_number": "Z-2475-2012",
      "product_description": "00789502801 VerSys Femoral Head Provisional 28mm -3.5  00789502802 VerSys Femoral Head Provisional 28mm +0  00789502803 VerSys Femoral Head Provisional 28mm +3.5  00789502804 VerSys Femoral Head Provisional 28mm +7  00789502805 VerSys Femoral Head Provisional 28mm +10.5  12/14 Neck Taper    The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.",
      "product_quantity": "231,641 all lots, all sizes",
      "reason_for_recall": "Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight.  Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp",
      "recall_initiation_date": "20110401",
      "center_classification_date": "20120928",
      "termination_date": "20140516",
      "report_date": "20121010",
      "code_info": "All lots distributed from lot 69998600 (2/20/1996) to present"
    }
  ]
}