      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92800",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide - Worldwide Distribution Foreign: Taiwan Uruguay Vietnam",
      "recall_number": "Z-2473-2023",
      "product_description": "Atellica CH Gamma-Glutamyl Transferase-IIn vitro diagnostic use in the quantitative determination of gamma-glutamyl transferase in human serum and plasma (lithium heparin  Test Code: GGT   Siemens Material Number (SMN): 11097597",
      "product_quantity": "19544 units",
      "reason_for_recall": "Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with  from Atellica¿CH Total Bilirubin_2 (TBil_2), Atellica¿ CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica¿ CH HDL Cholesterol (HDLC)",
      "recall_initiation_date": "20230710",
      "center_classification_date": "20230828",
      "report_date": "20230906",
      "code_info": "Unique Device Identification (UDI): 00630414596440 All Lot Numbers"
    }
  ]
}