  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "San Diego",
      "address_1": "11075 Roselle St Ste 200",
      "reason_for_recall": "The fuel gauge, the component that reads and reports the battery parameters to the pump, could provide inaccurate readings, which present to the user in one of two ways:   A) by triggering a Malfunction 4 Alarm, or   B) by triggering a succession of notifications prior to the pump shutting off, including Alert 2 (Battery Low), Alert 3 (Battery Very Low) and Alarm 12 (Battery Very Low). The alarms in both scenarios notify the user that the pump has stopped delivering insulin.",
      "address_2": "",
      "product_quantity": "55",
      "code_info": "Serial Numbers: 483731,  483930,  483992,  484013,  492055,  492368,  501157,  501177,  501178,  501187,  501191,  501200,  501202,    501205,    501207,    501268,    501285,    501364,    501368,    501370,    501376,    501385,    501386,    501387,    501432,  501435,  501440,  501443,  501444,  501445,  501466,  501479,  501488,  501490,  501491,  501495,  501503,  501508,  501509,  501519,  501522,  501523,  501526,  501535,  501560,  504522,  504531,  504535,  504540,   504753,   504972,   504990,   505025,   90472671,   90488584",
      "center_classification_date": "20180718",
      "distribution_pattern": "U.S. Consignees: IN, AZ, UT, TX, NY, NJ, MD, MN, IL, CO, AL, WI,   CA, PA, OH, OR, MS, FL, NM, ID, NV, KY, VA,",
      "state": "CA",
      "product_description": "t:slim G4 Insulin Pump with Dexcom G4 Platinum CGM    Software version: 4.3.4.3, Firmware version: 004722, Firmware, Fuel Gauge, Binary Rev B.",
      "report_date": "20180725",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Tandem Diabetes Care Inc",
      "recall_number": "Z-2471-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79887",
      "more_code_info": "",
      "recall_initiation_date": "20180423",
      "postal_code": "92121-1231",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}