{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bronshoj",
      "state": "N/A",
      "country": "Denmark",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85634",
      "recalling_firm": "Radiometer Medical ApS",
      "address_1": "Akandevej 21",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AK, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID,  IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT,  VA, VT, WA, WI, WV, WY and the country of CA.",
      "recall_number": "Z-2468-2020",
      "product_description": "ABL80-FLEX CO-OX 393-841 Box Label, REF 393-841, IVD, CE, UDI:  05700693938417 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.",
      "product_quantity": "1981",
      "reason_for_recall": "Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number.  Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.",
      "recall_initiation_date": "20200511",
      "center_classification_date": "20200629",
      "termination_date": "20230224",
      "report_date": "20200708",
      "code_info": "All lots"
    }
  ]
}