  "results": [
    {
      "status": "Ongoing",
      "city": "Pittsburgh",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92748",
      "recalling_firm": "Cardiac Assist, Inc",
      "address_1": "240 Alpha Dr",
      "address_2": "N/A",
      "postal_code": "15238-2906",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution in the states of IL, FL.",
      "recall_number": "Z-2465-2023",
      "product_description": "LifeSPARC Pump (part number 5800-0000) contained within TandemLife LifeSPARC Priming Tray part number: 5800-1000.     Note: The 5800-1000 model number is an individually packaged component contained within the Kit numbers:   5830-2916  LS TANDEMLIFE KIT   DL29  5840-2417  LS TANDEMLIFE KIT - V24, A17",
      "product_quantity": "2 units",
      "reason_for_recall": "Epoxy used during manufacturing of the LifeSPARC  Pump may have uncured epoxy. if undetectable curing may result in added risk of early pump failure,",
      "recall_initiation_date": "20230630",
      "center_classification_date": "20230825",
      "report_date": "20230906",
      "code_info": "S/N and UDI: (1)Kit 5830-2916  Pump SN 00876971  UDI: ( 01)00814112020609(11)230411(17)250219(21)00876971 (2)Kit 5840-2417 Pump SN 00876972  UDI: (01)00814112020609(11)230411(17)250219(21)00876972"
    }
  ]
}