  "results": [
    {
      "status": "Terminated",
      "city": "Wilmington",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71556",
      "recalling_firm": "Ferno-Washington Inc",
      "address_1": "70 Weil Way",
      "address_2": "N/A",
      "postal_code": "45177-9371",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Distributed in the states of AZ, CO, IA, MD, OH, PA & WA, and the countries of Australia, Canada & United Kingdom.",
      "recall_number": "Z-2461-2015",
      "product_description": "Wheel caster set screw for Ferno iNX Wheeled Stretcher.  Ferno iNX Wheeled Stretcher for pre-hospital patient transport.",
      "product_quantity": "26 units",
      "reason_for_recall": "The firm discovered an improperly installed set screw in one caster arm assembly on a unit in production. The screw assists in preventing the bearing cup on the caster assembly from backing out of the caster arm during use.",
      "recall_initiation_date": "20150501",
      "center_classification_date": "20150825",
      "termination_date": "20151106",
      "report_date": "20150902",
      "code_info": "Serial numbers: 15F001703, 15F001704, 15F001705, 15F001706, 15F001707, 15F001708, 15F001709, 15F001710, 15F001711, 15F001712, 15F001713, 15F001714, 15F001715, 15F001716, 15F001717, 15F001719, 15F001720, 15F001721, 15F001722, 15F001723, 15F001724, 15F001725, 15F001726, 15F001727, 15F001729  & 15F001730."
    }
  ]
}