  "results": [
    {
      "country": "United States",
      "city": "Fajardo",
      "address_1": "Calle Igualdad Final #7",
      "reason_for_recall": "Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.",
      "address_2": "",
      "product_quantity": "5 lots/1110 units (multiple units per lot)",
      "code_info": "Product code 900-2471 ,5 lots  112041034  112051959  112072723  113078521  140513219",
      "center_classification_date": "20140829",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) in the state of  Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.",
      "state": "PR",
      "product_description": "BASIC PACK 1 (Ill) CUSTOMED  CONTENTS:\t  (1) TABLE COVER REINFORCED 44\" X 78\" (4) DRAPE UTILITY WffAPE  (2) TOWELS ABSORBENT 15\" X 20\" (1)  SUTURE BAG FLORAL  (1) DRAPE SHEET 42\" X 57\" SMS  (1) SHEET :X 57\" X 63\" REINFORCED SMS (1) MAYO STAND COVER REINFORCED    Product Usage:  EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.",
      "report_date": "20140910",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Customed, Inc",
      "recall_number": "Z-2461-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68536",
      "termination_date": "20170814",
      "more_code_info": "",
      "recall_initiation_date": "20140520",
      "postal_code": "00738",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}