{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Montreal",
      "state": "",
      "country": "Canada",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99006",
      "recalling_firm": "AMD Medicom Inc.",
      "address_1": "2555 Ch de L'Aviation",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: MT, NC, VA, CA, TX, OH, TN, IL, RI",
      "recall_number": "Z-2459-2026",
      "product_description": "Ritmed Neurological Sponges, 0.75x0.75in, Sterile, REF: 85995C",
      "product_quantity": "10",
      "reason_for_recall": "Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.",
      "recall_initiation_date": "20260508",
      "center_classification_date": "20260616",
      "report_date": "20260624",
      "code_info": "UDI-DI: 10686864044465, Lot: 17936",
      "more_code_info": ""
    }
  ]
}