  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Plainsboro",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62711",
      "recalling_firm": "Integra Limited",
      "address_1": "311 Enterprise Dr",
      "address_2": "N/A",
      "postal_code": "08536-3344",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DC, GA, IA, IL, IN, NC, and NY and the countries of: Canada and Belgium.",
      "recall_number": "Z-2458-2012",
      "product_description": "The MAYFIELD¿ Horseshoe Gel Pads are used in conjunction with the MAYFIELD¿ Base Units and MAYFIELD¿ Swivel Adaptors for neurosurgeries in the prone or supine positions.  Those devices provide cranial stability and immobility for various craniotomies and spinal procedures.  The Gel Pads are supplied to users in labeled plastic bags, 1 per bag.      Product Usage:  The MAYFIELD¿ Horseshoe Gel Pads are used in conjunction with the MAYFIELD¿ Base Units and MAYFIELD¿ Swivel Adaptors for neurosurgeries in the prone or supine positions. They are plastic gel filled devices that provide cushioning between the head and other metallic part of the MAYFIELD¿ system. The Left and Right Pads are mirror images.",
      "product_quantity": "18 from Lot 093984 for Right Gel Pads",
      "reason_for_recall": "An investigation of customer complaints identified that for two product lots, Left Horseshoe Gel Pads were packaged as a Right Horseshoe Gel Pads and vice versa. None of the complaints resulted in a report of patient injuries or an increase in the time to complete the surgical procedure. A physician was consulted to assess patient risk and concluded that there is no potential harm to patients, onl",
      "recall_initiation_date": "20120727",
      "center_classification_date": "20120925",
      "termination_date": "20130220",
      "report_date": "20121003",
      "code_info": "The Lot No's for the affected Gel Pads are Lot 093984 of part number# 440C1011 for Right Gel Pads",
      "more_code_info": ""
    }
  ]
}