  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Stockholm",
      "state": "N/A",
      "country": "Sweden",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71414",
      "recalling_firm": "RAYSEARCH LABORATORIES AB",
      "address_1": "Sveavaegen 9",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed in CO, IL, MI, NC, OH,TN, and WA.",
      "recall_number": "Z-2457-2015",
      "product_description": "RayStation 4.7, version 4.7.1, used with Structure Definition or Automatic Breast Planning.  Radiation Therapy Treatment Planning System, for treatment planning and analysis of radiation therapy.",
      "product_quantity": "126 programs",
      "reason_for_recall": "Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior. These contraction distances are, in error, pairwise reversed, i.e. right is interpreted as left, inferior is interpreted as superior, and posterior is interpreted as anterior.",
      "recall_initiation_date": "20150527",
      "center_classification_date": "20150825",
      "termination_date": "20170922",
      "report_date": "20150902",
      "code_info": "software build numbers are 4.7.1.10"
    }
  ]
}