  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Memphis",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98629",
      "recalling_firm": "Smith & Nephew, Inc.",
      "address_1": "1450 E Brooks Rd",
      "address_2": "",
      "postal_code": "38116-1804",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "International distribution to the countries of BE, DE, IT.",
      "recall_number": "Z-2456-2026",
      "product_description": "TRIGEN PROXIMAL STRAIGHT HUMERAL NAIL 8/7 X 16, REF: 71760816",
      "product_quantity": "5",
      "reason_for_recall": "Humeral nail due to a manufacturing error is missing distal holes and if implanted may lead to revision surgery but, if the missing holes are identified before implantation, an equivalent backup can be used, or a bent nail of the same size, which would require additional surgical steps to address soft tissue/bone preparation, which could lead to a surgical delay of greater than 30 minutes.",
      "recall_initiation_date": "20260316",
      "center_classification_date": "20260616",
      "report_date": "20260624",
      "code_info": "UDI-DI: 03596010514394. Lot/Batch: 25BM03616",
      "more_code_info": ""
    }
  ]
}