{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bedford",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63180",
      "recalling_firm": "Hologic, Inc.",
      "address_1": "35 Crosby Dr",
      "address_2": "N/A",
      "postal_code": "01730-1450",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia,Belgium, Canada, China, Germany, Ecuador, Spain, France, Great Britain, Greece, Hong Kong, Italy, Japan Korea, Lithuvania, Neherlands, qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovenia, Thailand, Taiwan, South Africa, and Switzerland.",
      "recall_number": "Z-2452-2012",
      "product_description": "Hologic Fluoroscan Mini C-Arm Model with InSight2, Software version 4.0 and 4.0.1     The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.",
      "product_quantity": "414 units",
      "reason_for_recall": "InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180¿around vertical axis.  InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180¿ around vertical axis as well as 180¿ around a horizontal axis.",
      "recall_initiation_date": "20120914",
      "center_classification_date": "20120925",
      "termination_date": "20140110",
      "report_date": "20121003",
      "code_info": "Software version 4.0 and 4.0.1",
      "more_code_info": ""
    }
  ]
}