  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71848",
      "recalling_firm": "Carestream Health Inc",
      "address_1": "150 Verona St",
      "address_2": "N/A",
      "postal_code": "14608-1733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and in the countries: Australia, Canada, Chile, China, India, Korea, Malaysia, New Zealand, Singapore, Switzerland, Taiwan and Thailand.",
      "recall_number": "Z-2449-2015",
      "product_description": "CARESTREAM DRX-EVOLUTION X-Ray System     Product Usage:  The device is a permanently installed diagnostic X-ray system for general radiographic x-ray imaging including tomography. The tomography feature is not to be used for imaging pediatric patients.",
      "product_quantity": "Domestic: 76 units, Foreign: 160 units",
      "reason_for_recall": "Software defect that impacts image alignment when using the automatic stitching option for long length images. The defect could cause stitching inconsistencies which may go undetected on a radiograph.",
      "recall_initiation_date": "20150807",
      "center_classification_date": "20150821",
      "termination_date": "20160706",
      "report_date": "20150902",
      "code_info": "Software versions 5.6 and greater and 5.7 (current releases); Serial Numbers: 5029, 5569, 4067, 4754, 4131, 4100, 4096, 4076, 4569, 4570, 4588, 4670, 4767, 4879, 5266, 5393, 5408, 5442, 5572, 5573, 5555, 5556, 5563, 4146, 4679, 4702, 4703, 4704, 4801, 5090, 5154, 4787, 4069, 4074, 4344, 4085, 4058, 4084, 5470, 5158, 4086, 4099, 4073, 4665, 5241, 5303, 5304, 4639, 5350, 5352, 5429, 5508, 4696, 4763, 5182, 4291, 5037, 5554, 4192, 4117, 4172, 4207, 4225, 4239, 4278, 4301, 4310, 4328, 4333, 4338, 4366, 4384, 4421, 4427, 4442, 4443, 4447, 4465, 4499, 4523, 4534, 4538, 4539, 4692, 4822, 4861, 4875, 5128, 5131, 5174, 5553, 5579, 5580, 5597, 5628, 4320, 5309, 5042, 4807, 4808, 4809, 4810, 5493, 4266, 4711, 5327, 4125, 4143, 4097, 4156, 4282, 4287, 4289, 4258, 4742, 4964, 4784, 5050, 5551, 4233, 4644, 4082, 4113, 4196, 4683, 4309, 4315, 4064, 4256, 4987, 4624, 4585, 4882, 5362, 4004, 5377, 5418, 4300, 4308, 4627, 4638, 5138, 5251, 5360, 4947, 5620, 4985, 5403, 5495, 5560, 5544, 5545, 5621, 5622, 5451, 5282, 5645, 4724, 4298, 4269, 4756, 4418, 5273, 5178, 5179, 5443, 5197, 5389, 5584, 5473, 4899, 5334, 5431, 4990 --- NOTE: Additional serial numbers at foreign consignees to be determined upon reporting part usage."
    }
  ]
}