  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Ann Arbor",
      "state": "MI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62322",
      "recalling_firm": "Terumo Cardiovascular Systems Corporation",
      "address_1": "6200 Jackson Road",
      "address_2": "N/A",
      "postal_code": "48103-9586",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and the countries of AUSTRALIA, Singapore, Jordan, Japan, Vietnam, Germany, Russia, Saudi Arabia, BELGIUM, Mexico, and CANADA.",
      "recall_number": "Z-2449-2012",
      "product_description": "Terumo Advanced Perfusion System 1 Flow Module    indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.",
      "product_quantity": "72 units",
      "reason_for_recall": "Terumo Cardiovascular Systems (CVS) received one report of a flow module for the Terumo Advanced Perfusion System 1 not displaying blood flow information on the Central Control Monitor or the Centrifugal Control Unit. The tricolor LED on the flow module was not illuminated. Terumo CVS' investigation found that the malfunction was caused by an improper component on a circuit board used in a populat",
      "recall_initiation_date": "20120920",
      "center_classification_date": "20120925",
      "termination_date": "20140113",
      "report_date": "20121003",
      "code_info": "Catalog number: 802018 and serial numbers: 00122,00132, 00137, 00321-00337, 00339, 00340, 00342-00389,00391-00439, 00441, 00442,  00444-00455, 00457, 00459, 00460, 00466, 00467, 00475, and 00488.",
      "more_code_info": ""
    }
  ]
}