  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Fort Worth",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71941",
      "recalling_firm": "Alcon Research, Ltd.",
      "address_1": "6201 South Fwy",
      "address_2": "N/A",
      "postal_code": "76134-2099",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries Argentina, Australia, Austria, Belarus, Belgium, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Dominican Republic, France, Germany, Greece, Guatemala, Hungary, India, Italy, Japan, Jordan, Korea, Lebanon, Malaysia, Mexico, Mongolia, New Zealand/Fiji, Oman, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Switzerland, Taiwan, Thailand, Turkey, Ukraine and United Kingdom.",
      "recall_number": "Z-2448-2015",
      "product_description": "LX3 Floor Stand for Endure Luxite Envision Optical System; Endure LuxOR Q-Vue.    Product Usage:  The LX3 Microscope Floor Stand is an accessory to the LuxOR Opthalmic Microscope and the LuxOR Opthalmic Microscope with Q-VUE 3-D Assistant Visualization and provides a portable suspension system for the microscope that integrates an AC power module and microscope controls (Primary Control Panel and connectivity of foot controllers)",
      "product_quantity": "688 units",
      "reason_for_recall": "Alcon is initiating a voluntary recall of all LX3 Microscope Floor Stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules, potentially causing the system to display a system message, and/or resulting in illumination failure.",
      "recall_initiation_date": "20150811",
      "center_classification_date": "20150821",
      "termination_date": "20161104",
      "report_date": "20150902",
      "code_info": "All units LX3 Microscope Floor Stands manufactured prior to July 2015 (Serial Numbers 1402548501X to 1502857606X)"
    }
  ]
}