  "results": [
    {
      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71786",
      "recalling_firm": "Roche Diagnostics Operations, Inc.",
      "address_1": "9115 Hague Rd",
      "address_2": "N/A",
      "postal_code": "46256-1025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states of  LA, SC, CA, TX, NE, VA and WV including:  Puerto Rico.",
      "recall_number": "Z-2447-2015",
      "product_description": "cobas b 123 Fluid Pack COOX REF 05170036001 400    Product Usage:  The cobas b 123 POC system is a fully automated POC system for whole blood in vitro measurement of pH, blood gases (B3G), electrolytes NAT, K+, iCaWt (ISE), hematocrit (THct), metabolites (GLu, Lac) total hemoglobin (tHb), hemoglobin derivatives (O2Hb, HHb, COHb, MetHb), and oxygen saturation (SO2). The cobas b 123 POC system calculates derived parameters. Depending on the equipment configuration of the instrument, the Sensor Cartridge and the Fluid Pack used, the following parameters are measured in human whole blood and QC materials.",
      "product_quantity": "82 Fluid Packs",
      "reason_for_recall": "Low PO2 results.   QC failures of the PO2 parameter, affecting primarily Levels 1 and 2, caused by a calibration issue with the PO2 parameter.  This issue may not be detected since QC results can be below mean values, but still within 2 standard deviations (SD) limits.  Potential for erroneously low PO2 results in patient samples, especially in blood samples with PO2 values below 50 mmHg",
      "recall_initiation_date": "20150714",
      "center_classification_date": "20150821",
      "termination_date": "20160119",
      "report_date": "20150902",
      "code_info": "cobas b 123 Fluid Pack COOx Catalog 05170036001, lot numbers 21446113 (exp 9/5/2015), and  21446153( exp 9/24/2015)"
    }
  ]
}