  "results": [
    {
      "status": "Ongoing",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88588",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2446-2021",
      "product_description": "syngo.CT VA20A, VA30A and VA40A in:   syngo.via RT Image Suite / Sim&Go    SOMATOM go.Now   SOMATOM go.Up   SOMATOM go.All   SOMATOM go.Top   SOMATOM go.Sim  SOMATOM go.Open Pro and  SOMATOM X.cite    Material Numbers:  11061620  11061628  11061630  11061640  11061660  11061670",
      "product_quantity": "44 systems",
      "reason_for_recall": "Potential for data loss when using the contouring or patient marking workflow when editing a structure set after saving and reopening a study, resulting in an incomplete structure set sent to the treatment planning system (TPS).",
      "recall_initiation_date": "20210816",
      "center_classification_date": "20210910",
      "report_date": "20210922",
      "code_info": "Serial Numbers: 128051 130051 128065 128035 130057 128086 128079 128041 128019 128078 128060 128068 112045 128085 128075 128022 128033 130012 128016 130044 130048 130049 128018 119216 108085 128015 130019 130055 130054 128020 128039 130052 130036 130061 128049 117756 128029 130021 130045 130039 128063 128057 128045 128046  ***UPDATED 12/9/21*** 128114 128113 130058  ***Updated 1/31/22: Serial Number 128123 130015 130081 112261 112071 128124 128132 128125 130079 130077  Expanded Recall 3/12/22: 11061620 S/N: 111805 11061660  S/N: 128144, 128135, 128142, 128136, 128133 11061670 S/N: 130088, 130094, 130090, 130087, 130091, 130092, 130089",