  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Thornwood",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80418",
      "recalling_firm": "Zeiss, Carl Inc",
      "address_1": "1 Zeiss Dr",
      "address_2": "",
      "postal_code": "10594-1939",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2441-2018",
      "product_description": "Disinfection  unit UV (432332-9020-000)  for the Celldiscoverer 7 microscope",
      "product_quantity": "8",
      "reason_for_recall": "Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.",
      "recall_initiation_date": "20180529",
      "center_classification_date": "20180719",
      "termination_date": "20240726",
      "report_date": "20180725",
      "code_info": "432332-9020-000",
      "more_code_info": ""
    }
  ]
}