  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Munich",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85505",
      "recalling_firm": "LivaNova Deutschland GmbH",
      "address_1": "Lindberghstr. 25",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US:  FL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA  OUS:  Canada  -\tUzbekistan  -\tTurkey  -\tTunisia  -\tTaiwan  -\tSouth Africa  -\tSerbia  -\tSaudi Arabia  -\tRussia  -\tPoland  -\tMexico  -\tKorea(Rep.  of)  -\tIndia  -\tHong Kong  -\tColombia  -\tChina  -\tBrazil  -\tBahrain  -\tArgentina  -\tAlgeria  -\tItaly  -\tGermany",
      "recall_number": "Z-2440-2020",
      "product_description": "Control Panel REF - 60-02-15:  LivaNova SCP Pump Control Panel. REF 60-02-15, IPX2, RxOnly, UDI:    04033817900986",
      "product_quantity": "24",
      "reason_for_recall": "The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder.  The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.",
      "recall_initiation_date": "20190826",
      "center_classification_date": "20200626",
      "termination_date": "20210422",
      "report_date": "20200708",
      "code_info": "Serial Number:  60S14838  60S14839  60S14829  60S14724  60S14850  60S14858  60S14732  60S14737  60S14844  60S14846  60S14854  60S14855  60S14777  60S14778  60S14725  60S14727  60S14784  60S14785  60S14845  60S14774  60S14775  60S14786  60S14830  60S14776  60S14882"
    }
  ]
}