  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Munich",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85505",
      "recalling_firm": "LivaNova Deutschland GmbH",
      "address_1": "Lindberghstr. 25",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US:  FL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA  OUS:  Canada  -\tUzbekistan  -\tTurkey  -\tTunisia  -\tTaiwan  -\tSouth Africa  -\tSerbia  -\tSaudi Arabia  -\tRussia  -\tPoland  -\tMexico  -\tKorea(Rep.  of)  -\tIndia  -\tHong Kong  -\tColombia  -\tChina  -\tBrazil  -\tBahrain  -\tArgentina  -\tAlgeria  -\tItaly  -\tGermany",
      "recall_number": "Z-2439-2020",
      "product_description": "Control Panel REF - 60-02-50:  LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI:   04033817900993",
      "product_quantity": "8",
      "reason_for_recall": "The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder.  The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.",
      "recall_initiation_date": "20190826",
      "center_classification_date": "20200626",
      "termination_date": "20210422",
      "report_date": "20200708",
      "code_info": "Serial Number:  60S14735  60S14877  60S14715  60S14716  60S14736  60S14755  60S14707  60S14708"
    }
  ]
}