  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "The diluent may have an incomplete slit on the septum in the cap of the vial.  A diluent cap with a defective septum has the potential to cause erroneously elevated results if a sample is autodiluted onboard the system, due to a reduced volume of diluent being pipetted from the vial.",
      "address_2": "Po Box 6101",
      "product_quantity": "1195",
      "code_info": "lot 7CDA79, UDI # (UDI) 008427680324177CDA7918090110483586, exp 2018-09-01;  lot # 7DDA70, UDI # (UDI) 008427680324177DDA7018110110483586, exp 2018-11-01",
      "center_classification_date": "20180713",
      "distribution_pattern": "The products were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY.  The products were distributed to Guam, Saipan, and Virgin Islands.      The products were distributed to the following countries: Australia, Austria, Bahamas, Belgium, Brazil, Canada, Canary Islands, Croatia, Czech Republic, Denmark, France, Germany, Italy, Japan, Lebanon, Netherlands, Poland, Portugal, Qatar, Russian Federation, Slovakia, Slovenia, Saudi Arabia, South Korea, Spain, Switzerland, and United Arab Emirates.",
      "state": "DE",
      "product_description": "Dimension Vista¿ MULTI 2 SDIL, KD694, SMN# 10483586",
      "report_date": "20180725",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-2438-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "79784",
      "termination_date": "20200612",
      "more_code_info": "",
      "recall_initiation_date": "20180321",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}