  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92491",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide - US Nationwide distribution.",
      "recall_number": "Z-2435-2023",
      "product_description": "Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65",
      "product_quantity": "9175 units",
      "reason_for_recall": "Users were reporting instances of \"Gas Loss in IAB Circuit\" and \"Gas Gain in IAB Circuit\" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing).  Blood in the circuit, kins and abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms.  The HHE also identified hardware issues related to these alarms.",
      "recall_initiation_date": "20230731",
      "center_classification_date": "20230830",
      "report_date": "20230906",
      "code_info": "All Lot Numbers. Model: 0998-00-0800-31, UDI: 10607567109053; Model: 0998-UC-0800-31, UDI: N/A; Model: 0998-00-0800-32,UDI: 10607567111117; Model: 0998-00-0800-33, UDI: 10607567109008; Model: 0998-UC-0800-33, UDI: N/A; Model: 0998-00-0800-34, UDI: 10607567111940; Model: 0998-00-0800-35, UDI: 10607567109107; Model: 0998-00-0800-45, UDI: 10607567108421;Model: 0998-00-0800-52, UDI: 10607567108438; Model: 0998-UC-0800-52, UDI: N/A; Model: 0998-00-0800-53, UDI:10607567108391; Model: 0998-UC-0800-53, UDI: N/A; Model: 0998-00-0800-55, UDI: 10607567108414; Model: 0998-UC-0800-55,UDI: N/A; Model: 0998-00-0800-65, UDI: 10607567113432; Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449"
    }
  ]
}