      "country": "United States",
      "city": "Redwood City",
      "address_1": "900 Saginaw Dr",
      "reason_for_recall": "Following pressurization of the devices, damage to the device handle mechanism may occur, accompanied by a loud noise and potential dispersion of plastic fragments and components.",
      "address_2": "",
      "product_quantity": "400 umits",
      "code_info": "UDI: (01) 1 18 14900 00150 2  Lot numbers:  141118H,  151026C,   151118B,   160205A,   160805A,   160805B,   161128C,   161111D,   170222L,  170322J,   170712A,   170907D,   171201C,   171207C,  180103B.",
      "center_classification_date": "20180713",
      "distribution_pattern": "U.S. and international: Belgium, Germany, Italy and Netherlands. No gov accounts.",
      "state": "CA",
      "product_description": "AesDex, LLC   Cardica C-Port Flex-A PLUS Distal Anastomosis System, Catalog Number: FG-000150    The Cardica C-Port Flex-A PLUS Distal Anastomosis System, delivers a series of clips that create an anastomosis between a small target vessel and conduit. The stainless steel clips create a complete end-to-side anastomosis that is functionally equivalent to a hand-sutured interrupted stitch anastomosis. The system consists of one Anastomosis Device and one Retractor Mount. The system is designed with a flexible shaft to facilitate access to difficult to reach coronary targets.",
      "report_date": "20180725",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "AESDEX",
      "recall_number": "Z-2434-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80319",
      "termination_date": "20210427",
      "more_code_info": "",
      "recall_initiation_date": "20180402",
      "postal_code": "94063-4753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}