{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Miami Lakes",
      "state": "FL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88372",
      "recalling_firm": "Cordis Corporation",
      "address_1": "14201 NW 60th Ave",
      "address_2": "N/A",
      "postal_code": "33014-2802",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Canada, Europe.",
      "recall_number": "Z-2431-2021",
      "product_description": "Cordis SUPER TORQUE MB 5F PIG, Angiographic Catheter, REF 532-598B",
      "product_quantity": "71330 units",
      "reason_for_recall": "Cordis has identified that the SUPER TORQUE MB Angiographic Catheter (with Marker Bands) are being used in a manner that may allow entrapment of the catheter between endovascular devices and the vessel wall, which can lead to marker band movement or dislodgement.",
      "recall_initiation_date": "20210721",
      "center_classification_date": "20210915",
      "termination_date": "20240909",
      "report_date": "20210922",
      "code_info": "all codes",
      "more_code_info": ""
    }
  ]
}