  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Mahwah",
      "address_1": "325 Corporate Dr",
      "reason_for_recall": "Stryker is recalling rHead, uHead, Sigmoid Notch, Remotion, Radio Capitellum because packaging integrity (sterile barrier) of the packaging type KIT I may potentially be compromised by transportation.",
      "address_2": "",
      "product_quantity": "16992 total",
      "code_info": "Part Number: UHA-S1 (Ulnar Stem Implant #1 Sterile Packed) - Lot #'s 17584, 17585, 22217301, 22217302, 23470801, 19066-S2, 19417-S2, 19418-S2, 1973-S2, 19732-S2, 19785-S2, 19815-S2 and $23470801;  Part Number: UHA-S120 (Recon Ulnar Stem Implant #1, 20mm Extension) - Lot #'s 16725, 16726, 18577, 18757, 18758, 18759, 18867, 23124001, 19056-S2, 19786-S2, 19816-S2, 19928-S2 and 19929-S2;  Part Number: UHA-S2 (Ulnar Stem Implant #2 Sterile Packed) - Lot #'s 16714, 16715, 18760, 22760101, 22760102, 23124101, 23124102, 23470901, 19669-S2, 19817-S2, 19866-S2 and R23470901;  Part Number: UHA-S220 (Recon Ulnar Stem Implant #2, 20mm Extention) - Lot #'s 16720, 16721, 18415, 18761, 18762, 18763, 18764, 18765, 18868, 23124201, 19122-S2, 19787-S2, 19818-S2, 19930-S2 and 19932-S2;  Part Number: UHA-S3 (Ulnar Stem Implant #3 Sterile Packed) - Lot #'s 16716, 16717, 22760201, 22760202, 22760203, 23124301, 23124302, 19788-S2, 19789-S2, 19818-S2, 19820-S2, 19821-S2, 19867-S2 and 19868-S2;  Part Number: UHA-S320 (Recon Ulnar Stem Implant #3, 20mm Extention) - Lot #'s16722, 16723, 18869, 19065-S2, 19157-S2, 19733-S2, 19790-S2 and 19933-S2;  Part Number: UHA-S4 (Ulnar Stem Implant #4 Sterile Packed) - Lot #'s 16718, 16719, 18766, 18865, 18866, 19156-S2, 19670-S2, 19671-S2, 19672-S2, 19734-S2, 19822-S2 and 19869-S2;  Part Number: UHA-S420 (Recon Ulnar Stem Implant #4, 20mm Extention) - Lot #'s 16724, 18578, 18767, 18768, 18870, 18871, 22760301, 19072-S2, 19100-S2, 19158-S2, 19791-S2, 19823-S2, 19934-S2 and 19935-S2",
      "center_classification_date": "20150820",
      "distribution_pattern": "Worldwide Distribution.",
      "state": "NJ",
      "product_description": "uHead, Ulnar Implant, Standard Stem, Size 1, Sterile, Rx Only. Intended for replacement of the distal radioulnar joint following ulnar head resection arthroplasty.",
      "report_date": "20150826",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Howmedica Osteonics Corp.",
      "recall_number": "Z-2427-2015",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "71676",
      "termination_date": "20160429",
      "more_code_info": "",
      "recall_initiation_date": "20150624",
      "postal_code": "07430-2006",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}