  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Tempe",
      "address_1": "1810 W Drake Dr",
      "reason_for_recall": "Thrombogenicity test results demonstrated that both the test devices (SSS) and control devices (OM) did not meet the internal specification of  non-thrombogenic .",
      "address_2": "",
      "product_quantity": "2,722 units",
      "code_info": "2669978  2669987  2688793  2688794  2611427  2683820  2715724  2715729  2715730  2715757  2717434  2717435  2540303  2540308  2669971  2676061  2676062  2676063  2676066  2718603  2718604  2520710  2579520  2597159  2676070  2455042  2698584  2698585  2708150  2708152  2683539  2728373  2728374  2728471  2728473  2698582  2650600  2671503  2673627  2673753  2676052  2723516  2719375  2728372  2667503  2667504  2667508  2711911  2711913  2714099  2634425  2667502  2463266  2711909  2718456  2718457  2727804  2718454  2718455  2653972  2657505  2660189  2662668  2669911  2669912  2669913  2677027  2677030  2677032  2711065  2711067  2717567  2717568  2717569  2718453  2452247  2563722  2627797  2231657  2650352  2652204  2428771  2431261  2455437  2718392  2487373  2492934  2492935  2492936  2530007  2573328  2575113  2577649  2599284  2632405  2632435  2637254  2641646  2642451  2648809  2455438  2455439  2636330  2645840  2649865  2663765  2663767  2673768  2676056  2676060  2676064  2717822  2711322  2711323  2711324  2711325  2711326  2711096  2711100  2711320  2711059  2711062  2421530  2590901  2629625  2630695  2635829  2728880  2711094  2707509  2707514  2707515  2707516  2444646  2444649  2444650  2594332  2609784  2613319  2613320  2625109  2641491  2641725  2644024  2644025  2708370  2708371  2710839  2728685  2728686  2728688  2728691  2728801  2487850  2629618  2631616  2680821  2713852  2705060  2705297  2705298  2705359  2705364  2690033  2690034  2690035  2690036  2690037  2690038  2699613  2699614  2699697  2700978  2700980  2700981  2642970  2644574  2644575  2670437  2680816  2700987  2700988  2705024  2705041  2705042  2559785  2642964  2684922  2690027  2690028  2690029  2690030  2690031  2570031  2629620  2647031  2654846  2719368  2719384  2728682  2728683  2668243  2668396  2668397  2689336  2689340  2694434  2696789  2696790  2696795  2696999  2697473  2698842  2698843  2702901  2702902  2654855  2692041  2424988  2450374  2457907  2659011  2659019  2684914  2421500  2497497  2498945  2498951  2718694  2503754  2511938  2517832  2533215  2563463  2570030  2668242  2231718  2448728  2461582  2461924  2468439  2473883  2478014  2526468  2553995  2454287  2458190  2552269  2596754  2596758  2609786  2625150  2635962  2642607  2642608  2668241  2631690  2631692  2635592  2635593  2636538  2640019  2640021  2640136  2640137  2640163  2668482  2680696  2680698  2680701  2680702  2680704  2680705  2698854  2701261  2627023  2627502  2424040  2431391  2439893  2443246  2443317  2443323  2444878  2448523  2518755  2537237  2634733  2636601  2636602  2636603  2636604  2544410  2562543  2565814  2565887  2554357  2554359  2566031  2566045  2617492  2629188  2629189  2631446  2631447  2631558  2627021  2627022  2530222  2530247  2530409  2530413  2530414  2530552  2530555  2530765  2531385  2531934  2531935  2531936  2533884  2533885  2535628  2535783  2537590  2538439  2544238  2544239  2544240  2544241  2626433  2464582  2464765  2465652  2465666  2468559  2469151  2469154  2469156  2469158  2469160  2469683  2626437  2626438  2626482  2626973  2626975  2626976  2470323  2470324  2470951  2473138  2473275  2473405  2473618  2473620  2474469  2477975  2477977  2478068  2480317  2482591  2482592  2482743  2482744  2482781  2484348  2484353  2521922  2521924  2524558  2530203  2530207  2530211  2452117  2452118  2452216  2452217  2452218  2618892  2622075  2622076  2452220  2452221  2452315  2461597  2464247  2464249  2622079  2439787  2451258  2451844  2452040  2472722  2472753  2549347  2572282  2572286  2572287  2572289  2581440  2591345  2591347  2618891  2452245  2463091  2463291  2463292  2463599  2463601  2463607  2463626  2444653  2444655  2444659  2444882  2445974  2445975  2447422  2469330  2469336  2481253  2481254  2481265  2485917  2485918  2485920  2434967  2439402  2439786  2439809  2231961  2432273  2718497  2439813  2696912  2696918  2696921  2696923  2698078  2698080  2698081  2424862  2424869  2424922  2424924  2424925  2701011  2429015  2624332  2691805  2708977  2713269  2713372  2717882  2717883  2718207  2718208  2718209  2718210  2718425  2727692  2727700  2696898  2696899  2696905  2632412  2668225  2668280  2689746  2689747  2689749  2586986  2624498  2628367  2689751  2694669  2696670  2696885  2696892  2696896  2628974  2706486  2713100  2717589  2508055  2527913  2717591  2717600  2710937  2710939  2711093  2711095  2711098  2711102  2711210  2711211  2711214  2711215  2712799  2712809  2713086  2713087  2713120  2717520  2575107  2586991  2588902  2614378  2642606  2696284  2696403  2698003  2698006  2698090  2429754  2698091  2698092  2698093  2698094  2698095  2698096  2698097  2628972  2628973  2635776  2635883  2636808  2685638  2700153  2685640  2691057  2691069  2691071  2706516  2706517  2706518  2706519  2706520  2706521  2700154  2700358  2434473  2612254  2642605  2706739  2708244  2708246  2710780  2710891  2623015  2628128  2700152  2581083  2629643  2629816  2632034  2635961  2700142  2700143  2700144  2585301  2669684  2706192  2707842  2707884  2710409  2619674  2449060  2549560  2549561  2688029  2688032  2688046  2688047  2688048  2693690  2697532  2697578  2697579  2697588  2700129  2700130  2631612  2669206  2636356  2669210  2704747  2705758  2705760  2669682  2669683  2676118  2681437  2681800  2681808  2681852  2681853  2681854  2681941  2681942  2683556  2683557  2683586  2684976  2684978  2705360  2705361  2705362  2705363  2705482  2705483  2508693  2624202  2710047  2710050  2710051  2710052  2684969  2684972  2664702  2664741  2664954  2664955  2664956  2664957  2664959  2664960  2664962  2664963  2664964  2664965  2671728  2671874  2672011  2672012  2672013  2672014  2673866  2673868  2674176  2674177  2674398  2674405  2674413  2649214  2649218  2651104  2655369  2444090  2544179  2590964  2594358  2596153  2627065  2627067  2627983  2439482  2688335  2424987  2430724  2434995  2688336  2688337  2692763  2702450  2705066  2705313  2710028  2669361  2684956  2684958  2684959  2684963  2684966  2684968  2705314  2705436  2705439  2705441  2705444  2688333  2688334  2596539  2699389  2702199  2635779  2640622  2660526  2670081  2670085  2670086  2671997  2673968  2674042  2674043  2679652  2679840  2679841  2679842  2679843  2683540  2683541  2688254  2688255  2688332  2610277  2610279  2613326  2618917  2622850  2622853  2622922  2630508  2668872  2668874  2694748  2694749  2694750  2694751  2606806  2608499  2610224  2599968  2696667  2434996  2443743  2443744  2449069  2449072  2456882  2458196  2459505  2463137  2684055  2530170  2530175  2532637  2540326  2543727  2567832  2577602  2579862  2587700  2696668  2597278  2610085  2667731  2684715  2698142  2683821  2684053  2668327  2668329  2696664  2621577  2673895  2668308  2668312  2668316  2674232  2674233  2674235  2683815  2683816  2696422  2696427  2696428  2696509  2698373  2698420  2661229  2663365  2671631  2671632  2671637  2671638  2679217  2679219  2679220  2679224  2679227  2679429  2679430  2692895  2692897  2698524  2698840  2698852  2604859  2667567  2702903  2702940  2702942  2702943  2702944  2702945  2671644  2671646  2689460  2698899  2701059  2673894  2619666  2667844  2697141  2660714  2660726  2660751  2661226  2630697  2635325  2647156  2649209  2651083  2651085  2422260  2422262  2429660  2454282  2472701  2484553  2486246  2486247  2486250  2486253  2486259  2486316  2487892  2491915  2491916  2492240  2692846  2692847  2692849  2695061  2493370  2493553  2688589  2688591  2693689  2507068  2524731  2642237  2693696  2697580  2700045  2700046  2520082  2615824  2624447  2693560  2697879  2700359  2692695  2692696  2679820  2679821  2679855  2679948  2679949  2679950  2679951  2688586  2616951  2679758  2679759  2687748  2692692  2692693  2679817  2679818  2679819  2635801  2651732  2653272  2653273  2655911  2660650  2683371  2683402  2683403  2683438  2687743  2687744  2606995  2624604  2635799  2687563  2687564  2687565  2424899  2450938  2521676  2580885  2581530  2599957  2670821  2671428  2672452  2674988  2678875  2678987  2678998  2679000  2679002  2679094  2679105  2679106  2679108  2683207  2687562  2610207  2616950  2679564  2458440  2480743  2496725  2555023  2565939  2571039  2608493  2618913  2661382  2661383  2670707  2434992  2445981  2447230  2448038  2454331  2432026  2433496  2434704  2424483  2430774  2681109  2681114  2582060  2587257  2670326  2670328  2679550  2679552  2499019  2581786  2582058  2629968  2677145  2681105  2683817  2683822  2498943  2683985  2683986  2684058  2640100  2640102  2658052  2658056  2658134  2658140  2674361  2674369  2674371  2674373  2674385  2674499  2674695  2674700  2674787  2674789  2676255  2672454  2672455  2672456  2672923  2672925  2672926  2672927  2674358  2617319  2622996  2631566  2631578  2631583  2678296  2683768  2487676  2508129  2586273  2611430  2614638  2511840  2517768  2603610  2603611  2677476  2427114  2427115  2427296  2492294  2510697  2581457  2597275  2597279  2648600  2680917  2680921  2682369  2682820  2682822  2682825  2656125  2656126  2672265  2674920  2683226  2683227  2683370  2692894  2695331  2673239  2673240  2673241  2673242  2673243  2673305  2673362  2673658  2674550  2674574  2674575  2677647  2677923  2677924  2677925  2680906  2640950  2642150  2646054  2648104  2527309  2632389  2635603  2603164  2660494  2678120  2573333  2576828  2594278  2602934  2608498  2612298  2619175  2627066  2434969  2497005  2627069  2651010  2665733  2670834  2672781  2672784  2672788  2673044  2673046  2498955  2511556  2511833  2517509  2436339  2673460  2445453  2445456  2445457  2446208  2496723  2516245  2505562  2538398  2591943  2597277  2664891  2671359  2673394  2583692  2649211  2649212  2673456  2673457  2654597  2658595  2660489  2660491  2661653  2664161  2664166  2424838  2458954  2523618  2424039  2544171  2544175  2544366  2545315  2581459  2581462  2590792  2590968  2599519  2651530  2651532  2664797  2664798  2664801  2664802  2664879  2664882  2664883  2644150  2644151  2647177  2429306  2433673  2433674  2434815  2518922  2585298  2585299  2663619  2663621  2663764  2663766  2677681  2677682  2657137  2657138  2658392  2658395  2658514  2677683  2677684  2459507  2463106  2578816  2582214  2644267  2644268  2504530  2504531  2649714  2649715  2657429  2657431  2657435  2657981  2657983  2657984  2658000  2658048  2658049  2658050  2658051  2658055  2658855  2658859  2658886  2672575  2672576  2672577  2672586  2675481  2675482  2676160  2676161  2676162  2676163  2676164  2676165  2676166  2676167  2676257  2676258  2676259  2653337  2662589  2663617  2677679  2677680  2658153  2502867  2502869  2502870  2448399  2652203  2663832  2663833  2663835  2446202  2481631  2481634  2481635  2481636  2481637  2481704  2481707  2481708  2602689  2603020  2654225  2656510  2656511  2656513  2643947  2643948  2652604  2652605  2652606  2652607  2652608  2652610  2652614  2644266  2658135  2658136  2658137  2658145  2658146  2658151  2424770  2583875  2629626  2641498  2643945  2656258  2656259  2656260  2444903  2656262  2602634  2642448  2642449  2642450  2653974  2444883  2629619  2444888  2449073  2452928  2455474  2540141  2545204  2545205  2565943  2599708  2603022  2649177  2649178  2649180  2649183  2649184  2649186  2653406  2662666  2662667  2429434  2580395  2595857  2602635  2603023  2429220  2511188  2629616  2631948  2632036  2632046  2640488  2505518  2505519  2510545  2510548  2444893  2447087  2510551  2567192  2567193  2577645  2580889  2586232  2660206  2492939  2581623  2487367  2660190  2660191  2660192  2660193  2492937  2605525  2605526  2492299  2505466  2515148  2631629  2574808  2584134  2596569  2596570  2648806  2648807  2648811  2648812  2660188  2650887  2650888  2652662  2507039  2507040  2572966  2652663  2655478  2431566  2636809  2499660  2516589  2516590  2516706  2644769  2644770  2647172  2648570  2648571  2650886  2471472  2591665  2621659  2637256  2641391  2641402  2642185  2644532  2644533  2644534  2106038  2424478  2424709  2645961  2647611  2647613  2648052  2648299  2655143  2655274  2427342  2429445  2429446  2444836  2628975  2636806  2632416  2641352  2511560  2515144  2635789  2515150  2612209  2642005  2650271  2650279  2650329  2650330  2650348  2231579  2444078  2559110  2614404  2629416  2444282  2516246  2520708  2635773  2635784  2632407  2632409  2632410  2632413  2632415  2445752  2487849  2512210  2517757  2517759  2498958  2520083  2527268  2554975  2527916  2619109  2619110  2619111  2626989  2626992  2626993  2634379  2634384  2634385  2636279  2636280  2636285  2636321  2636340  2636345  2636346  2636347  2636357  2631484  2632404  2650350  2644020  2644022  2644023  2444997  2450731  2621579  2498926  2498949  2635890  2640853  2640854  2431259  2444885  2545203  2552235  2552248  2629812  2632433  2632437  2646519  2612256  2614184  2629639  2629640  2619671  2619672  2626409  2632159  2436866  2438005  2438011  2573304  2575109  2575111  2637028  2637029  2637034  2597523  2617423  2635309  2635316  2596610  2624333  2617424  2566884  2577549  2583013  2421886  2424895  2431734  2444881  2454273  2552268  2631928  2631931  2631934  2631936  2632026  2632044  2432245  2432249  2439922  2444854  2563793  2642850  2642861  2645806  2647547  2647548  2635303  2624499  2624500  2625404  2635797  2640608  2640618  2618918  2618922  2630498  2630500  2630502  2630506  2631568  2631650  2614634  2614639  2631655  2631660  2631661  2631666  2631667  2631669  2631671  2631674  2634422  2634424  2635795  2434806  2463596  2517762  2618916  2444291  2569464  2575409  2606949  2613596  2624206  2471484  2471487  2501424  2539835  2581503  2581509  2591937  2630442  2424775  2523082  2570385  2612259  2612260  2591939  2608736  2608741  2618914  2618915  2436338  2443064  2443784  2445611  2612301  2627979  2627982  2619171  2619174  2621147  2621158  2614386  2443745  2448102  2450924  2450925  2455394  2457693  2460590  2463294  2608497  2614797  2614799  2502062  2614801  2616286  2555031  2586989  2511946  2515618  2610158  2594330  2613655  2597274  2620635  2625857  2633297  2633298  2438331  2597280  2511186  2610280  2454709  2444891  2449065  2519650  2455471  2581455  2581458  2434874  2445052  2581460  2591780  2591788  2614427  2614428  2448099  2448101  2450930  2587381  2594274  2594275  2594277  2444995  2448247  2613607  2456661  2505443  2508699  2429755  2520094  2604866  2604867  2587375  2610087  2610092  2587371  2587374  2419917  2429704  2454687  2596053  2606928  2607189  2574774  2574776  2574778  2616952  2454690  2465096  2465009  2574779  2596520  2612566  2617128  2581533  2581535  2582212  2594354  2594359  2594360  2610204  2616949  2586751  2442861  2539844  2584413  2584416  2459508  2459558  2584417  2584420  2594078  2596075  2446285  2463090  2599971  2521938  2530121  2447423  2559009  2563056  2594077  2581156  2421418  2481547  2524734  2571198  2589334  2589336  2590736  2602688  2606270  2606277  2606280  2421490  2472676  2501430  2606314  2589338  2590790  2505467  2581108  2590791  2590962  2590969  2599506  2429417  2452991  2455464  2455498  2458991  2540145  2564789  2579274  2581106  2585785  2603074  2576829  2581084  2581085  2606269  2565801  2443872  2605527  2567186  2532636  2606257  2606264  2606265  2539837  2539839  2444840  2444892  2586442  2586444  2586451  2606256  2586081  2589257  2590321  2606037  2573336  2573340  2424740  2458204  2458439  2577545  2577548  2577583  2580886  2583014  2592510  2597522  2573331  2431951  2443881  2445963  2562846  2573312  2580389  2580392  2580393  2580394  2586059  2603035  2429723  2434862  2439476  2551330  2579515  2577607  2565940  2572810  2448112  2453400  2457568  2574146  2575833  2446980  2559784  2559786  2624208  2439212  2566886  2572725  2572727  2572728  2580062  2557117  2559781  2559782  2559783  2566882  2455445  2527308  2527310  2445812  2455365  2482887  2482890  2527299  2527301  2527303  2444286  2480744  2480746  2518700  2444057  2540566  2540570  2553834  2570496  2527304  2527305  2527306  2527307  2526311  2527266  2552140  2424766  2532886  2539816  2552256  2518914  2522595  2444120  2515860  2563618  2443148  2563621  2563723  2506514  2480742  2509684  2613600  2613602  2455462  2563612  2563614  2563615  2424776  2424861  2431957  2563548  2563788  2563789  2563792  2499196  2559135  2559137  2527914  2527922  2554067  2559033  2559134  2533213  2434860  2434863  2541792  2541793  2544167  2544169  2520704  2520705  2520706  2520707  2553907  2434814  2503866  2520709  2520711  2520712  2520713  2520714  2520715  2544172  2544177  2544178  2544180  2544210  2544215  2544217  2544226  2518643  2527788  2554058  2554060  2554062  2515615  2515616  2515617  2515619  2445985  2445992  2526778  2541787  2541789  2541790  2541791  2436028  2511555  2532569  2511557  2511952  2231710  2429724  2431929  2522605  2522608  2522609  2522611  2522612  2511936  2511941  2498928  2498935  2498939  2498952  2498957  2498961  2517769  2231578  2231582  2559074  2559111  2499021  2499023  2511825  2511827  2231641  2231656  2511828  2511829  2511837  2424857  2424858  2424859  2551598  2551599  2570383  2517767  2545311  2545313  2545316  2525630  2434875  2448343  2517765  2517766  2543562  2543563  2556366  2437084  2516287  2516288  2516289  2525728  2528072  2528073  2517763  2424713  2424745  2424747  2434872  2606944  2434873  2436849  2543561  2517758  2517760  2511185  2420918  2421501  2421529  2423939  2509924  2509963  2511153  2517755  2550393  2550401  2555024  2555028  2555029  2555033  2424712  2523489  2523490  2523619  2523621  2523915  2531337  2540322  2487677  2508127  2538659  2538661  2540306  2540309  2540315  2540317  2519653  2519654  2519655  2537236  2478735  2520093  2520146  2520156  2519651  2605272  2605273  2605275  2463290  2463295  2463296  2517608  2520088  2520089  2488812  2488814  2450928  2429016  2442862  2497031  2503265  2450931  2450939  2450946  2450947  2454958  2454959  2463287  2511314  2450730  2450888  2450927  2443739  2448100  2508697  2508892  2424736  2427025  2427373  2510567  2510568  2510569  2434997  2434998  2439411  2439479  2439480  2454944  2455161  2455293  2455318  2455343  2463590  2424814  2424815  2424835  2432347  2432384  2472703  2472752  2472757  2434959  2434960  2434963  2434964  2506764  2462706  2463089  2463126  2463130  2463138  2472700  2421498  2489667  2489734  2489735  2489737  2496087  2496418  2501688  2459520  2545314  2508056  2508057  2459521  2463124  2502300  2502865  2449068  2449063  2472668  2492083  2495250  2496132  2496133  2496518  2502061  2438135  2439916  2439917  2439918  2456223  2456227  2456536  2439920  2444087  2444088  2444089  2454276  2456883  2457064  2458202  2436531  2438196  2454286  2471641  2471642  2478890  2479479  2487848  2492082  2454281  2420822  2420829  2420831  2431953  2454285  2575407  2575408  2575410  2575411  2496516  2501840  2454279  2492282  2492283  2492275  2492276  2439924  2439925  2444852  2444880  2462606  2462608  2462609  2444908  2447342  2448532  2448533  2454274  2454277  2492268  2505437  2460548  2460591  2460592  2460593  2458926  2439949  2458927  2458987  2458989  2458990  2458992  2458993  2446940  2448536  2448634  2453780  2453782  2453783  2458217  2458219  2459263  2459266  2459267  2459269  2459339  2459340  2460547  2501432  2501433  2505436  2453511  2453513  2453515  2453517  2453521  2458925  2453374  2453376  2453383  2453384  2453385  2453386  2453398  2501425  2447232  2447233  2453311  2454626  2456670  2463003  2463159  2420814  2447126  2447128  2447231  2471466  2454602  2454606  2447089  2447093  2458555  2458557  2458559  2458592  2458593  2458594  2450489  2453230  2453231  2453233  2455518  2455522  2458436  2458438  2460640  2460644  2450463  2450486  2424991  2439406  2439408  2453304  2454561  2455519  2456129  2456178  2456471  2456646  2456648  2457670  2458246  2460665  2460666  2472667  2444847  2444848  2444849  2444850  2444853  2444855  2448525  2447256  2447258  2454245  2458192  2454247  2454254  2444843  2444846  2458194  2458206  2458207  2465005  2467500  2448520  2448521  2448522  2447254  2446608  2446610  2455366  2539833  2539834  2539836  2539838  2539841  2539842  2539843  2444905  2444907  2447217  2447253  2448542  2458215  2458218  2458245  2424989  2444837  2450371  2450372  2457908  2444864  2444865  2444886  2444894  2444895  2444896  2444902  2444904  2446085  2446086  2446094  2448429  2448430  2448431  2455463  2455475  2455483  2455484  2458211  2458212  2458213  2443490  2446874  2446986  2446994  2447198  2444246  2444247  2439464  2439465  2439466  2439470  2439474  2448727  2448732  2455477  2455480  2458560  2458561  2429706  2429722  2431925  2431950  2434855  2434857  2443485  2432340  2439896  2439923  2446703  2467505  2455392  2455393  2455395  2452930  2452931  2540465  2540466  2540468  2448397  2452906  2452907  2452908  2452925  2452927  2445993  2445994  2458535  2465006  2445996  2445997  2446001  2446002  2465008  2467502  2446204  2446205  2446207  2450476  2450493  2450494  2443871  2443873  2443875  2443876  2443444  2443445  2443470  2443471  2443472  2448028  2448030  2448033  2448037  2424898  2424901  2434990  2434993  2429705  2439335  2443230  2443244  2443245  2448250  2455403  2457653  2457654  2443321  2443325  2443326  2439334  2448246  2439333  2431959  2431960  2432244  2434805  2438328  2438329  2438346  2424773  2424774  2444064  2444566  2446050  2446059  2448244  2424777  2424781  2424782  2429978  2429988  2431900  2431952  2439029  2439031  2439296  2439297  2439382  2439390  2443063  2443143  2445020  2448029  2434876  2439028  2444284  2516236  2516244  2427357  2431392  2431393  2431394  2431896  2434866  2434867  2434870  2434871  2431317  2431321  2431323  2431324  2431388  2431390  2424750  2431316  2445050  2439401  2424710  2424711  2424714  2424741  2424742  2442856  2442857  2443032  2445751  2424851  2424902  2434986  2442855  2442854  2429760  2431235  2431254  2431258  2431262  2431265  2431267  2431269  2431270  2455345  2421923  2444285  2443883  2443885  2445965  2448202  2431271  2432858  2432860  2434470  2434474  2434476  2434477  2434633  2434636  2434701  2434702  2439147  2429756  2429758  2424708  2427365  2427366  2427368  2427369  2427371  2427372  2427385  2429752  2422484  2422485  2422486  2424229  2424230  2424232  2424233  2424234  2447685  2424235  2443767  2437139  2445048  2445049  2445051  2439392  2455119  2455120  2445494  2447682  2447683  2447684  2433676  2436581  2432262  2432263  2432264  2432265  2432266  2432267  2432268  2432270  2434864  2439387  2448026  2448027  2436584  2436832  2436833  2436834  2431452  2431453  2434807  2434809  2434813  2431172  2444289  2453920  2453924  2453925  2429655  2429661  2429662  2431171  2429451  2429450  2431922  2431903  2431904  2431905  2429449  2424227  2424481  2424482  2429211  2429212  2429447  2429448  2431898  2437482  2438007  2438009  2438012  2438013  2438092  2438194  2438537  2445607  2445610  2420689  2421060  2421061  2421063  2421654  2422018  2422019  2422020  2424226  2445612  2427374  2427380  2427383  2445753  2437072  2424754  2424755  2431678  2431889  2431897  2437074  2437075  2437079  2437083  2437113  2437481  2437067  2419754  2433610  2433611  2433612  2436611  2436612  2437066  2422481  2424707  2428974  2429726  2429727  2431559  2431563  2431189  2431190  2431191  2433608  2433609  2480748  2480749  2482889  2482899  2482903  2438197  2438198  2438201  2447574  2430722  2427107  2427108  2435709  2436927  2436929  2427163  2429219  2429222  2429433  2423917  2423995  2423996  2423997  2424042  2424043  2427079  2427081  2427106  2421884  2421885  2421797  2421794  2421795  2421796  2420876  2421781  2421788  2421790  2424130  2427021  2421792  2470362  2420873  2420875  2419968  2420837  2420839  2421499  2421527  2421627  2423916  2423937  2424129  2419612  2445458  2419613  2421888  2420817  2420827  2420830  2420835  2436029  2436299  2436301  2436404  2436405  2445455  2424148  2428428  2428506  2428512  2433068  2433578  2436026  2412447  2421713  2421716  2428431  2428717  2420743  2420744  2421362  2421364  2455359  2455361  2455364  2455370  2455372  2445810  2427356  2421925  2421927",
      "center_classification_date": "20180712",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and Canada",
      "state": "AZ",
      "product_description": "Coronary Sinus (CS) Uni-Directional Diagnostic Electrophysiology (EP) Catheter, catalog numbers D135303 and D135304.    Product Usage:  The Reprocessed CS Diagnostic EP Catheter is indicated for electrophysiological mapping of cardiac structures; i.e., stimulation and recording only. The catheter is designed for use in the coronary sinus.",
      "report_date": "20180718",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Sustainability Solutions",
      "recall_number": "Z-2426-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80324",
      "termination_date": "20200728",
      "more_code_info": "",
      "recall_initiation_date": "20180329",
      "postal_code": "85283-4327",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}