  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "West Chester",
      "address_1": "1301 Goshen Pkwy",
      "reason_for_recall": "The wrap is for single use for nail sterilization but does not have sufficient testing for the device to be considered a multi-use item.",
      "address_2": "",
      "product_quantity": "1487",
      "code_info": "all lots of part numbers: 900.50, 900.51",
      "center_classification_date": "20150820",
      "distribution_pattern": "Worldwide Distribution - US Distribution including Puerto Rico and to the countries of : Canada, Switzerland and New Zealand.",
      "state": "PA",
      "product_description": "Synthes Tibial Nail Wrap Holds 11 Nails and Synthes Intramedullary Nail Wrap Holds 10 Nails; the Nail Wrap is intended to be used as a protective sheath specifically for Universal Femoral Nails, Universal Tibial Nails, Unreamed Tibial Nail, and the Stainless Steel Tibial Nail during steam sterilization by the hospital.",
      "report_date": "20150826",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Synthes (USA) Products LLC",
      "recall_number": "Z-2425-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71713",
      "termination_date": "20160818",
      "more_code_info": "",
      "recall_initiation_date": "20150701",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}