      "country": "United States",
      "city": "Mounds View",
      "address_1": "8200 Coral Sea St Ne",
      "reason_for_recall": "Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage.  Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert.  Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.",
      "address_2": "",
      "product_quantity": "5057 units",
      "code_info": "Product UPN/GTIN:  00613994425027  Serial Numbers:  PNT608737S  PNT608738S  PNT608739S  PNT608740S  PNT608742S  PNT608743S  PNT608744S  PNT608745S  PNT608746S  PNT608748S  PNT608749S  PNT608750S  PNT608751S  PNT608752S  PNT608753S  PNT608754S  PNT608755S  PNT608756S  PNT608757S  PNT608758S  PNT608759S  PNT608760S  PNT608761S  PNT608762S  PNT608763S  PNT608764S  PNT608765S  PNT608766S  PNT608767S  PNT608768S  PNT608769S  PNT608770S  PNT608771S  PNT608772S  PNT608773S  PNT608774S  PNT608775S  PNT608776S  PNT608777S  PNT608778S  PNT608779S  PNT608780S  PNT608781S  PNT608782S  PNT608783S  PNT608784S  PNT608785S  PNT608786S  PNT608787S  PNT608788S  PNT608789S  PNT608790S  PNT608791S  PNT608792S  PNT608793S  PNT608794S  PNT608795S  PNT608796S  PNT608798S  PNT608799S  PNT608800S  PNT608801S  PNT608802S  PNT608803S  PNT608804S  PNT608805S  PNT608806S  PNT608807S  PNT608808S  PNT608809S  PNT608810S  PNT608830S  PNT608924S  PNT608926S  PNT608927S  PNT608930S  PNT608931S  PNT608932S  PNT608934S  PNT608938S  PNT608939S  PNT608940S  PNT608944S  PNT608945S  PNT608946S  PNT608947S  PNT608948S  PNT608949S  PNT608951S  PNT608952S  PNT608953S  PNT608954S  PNT608955S  PNT608956S  PNT608957S  PNT608958S  PNT608959S  PNT608960S  PNT608961S  PNT608962S  PNT608963S  PNT608964S  PNT608965S  PNT608966S  PNT608967S  PNT608968S  PNT608969S  PNT608970S  PNT608971S  PNT608972S  PNT608973S  PNT608974S  PNT608975S  PNT608976S  PNT608977S  PNT608978S  PNT608979S  PNT608980S  PNT608981S  PNT608982S  PNT608983S  PNT608984S  PNT608985S  PNT608986S  PNT608987S  PNT608988S  PNT608989S  PNT608990S  PNT608991S  PNT608992S  PNT608993S  PNT608994S  PNT608995S  PNT608996S  PNT608997S  PNT608998S  PNT608999S  PNT609000S  PNT609001S  PNT609073S  PNT609074S  PNT609075S  PNT609076S  PNT609077S  PNT609078S  PNT609079S  PNT609081S  PNT609082S  PNT609083S  PNT609115S  PNT609116S  PNT609117S  PNT609118S  PNT609119S  PNT609120S  PNT609121S  PNT609122S  PNT609123S  PNT609124S  PNT609125S  PNT609127S  PNT609128S  PNT609129S  PNT609130S  PNT609131S  PNT609132S  PNT609133S  PNT609134S  PNT609135S  PNT609137S  PNT609138S  PNT609139S  PNT609140S  PNT609141S  PNT609142S  PNT609143S  PNT609144S  PNT609145S  PNT609146S  PNT609147S  PNT609148S  PNT609149S  PNT609150S  PNT609151S  PNT609152S  PNT609153S  PNT609154S  PNT609155S  PNT609157S  PNT609158S  PNT609159S  PNT609160S  PNT609161S  PNT609162S  PNT609163S  PNT609164S  PNT609167S  PNT609170S  PNT609172S  PNT609174S  PNT609176S  PNT609177S  PNT609179S  PNT609183S  PNT609185S  PNT609186S  PNT609187S  PNT609188S  PNT609189S  PNT609190S  PNT609191S  PNT609192S  PNT609193S  PNT609194S  PNT609195S  PNT609197S  PNT609198S  PNT609199S  PNT609200S  PNT609201S  PNT609202S  PNT609203S  PNT609204S  PNT609205S  PNT609206S  PNT609208S  PNT609209S  PNT609210S  PNT609211S  PNT609212S  PNT609214S  PNT609215S  PNT609216S  PNT609223S  PNT609224S  PNT609225S  PNT609226S  PNT609227S  PNT609228S  PNT609230S  PNT609231S  PNT609232S  PNT609233S  PNT609234S  PNT609235S  PNT609236S  PNT609237S  PNT609238S  PNT609239S  PNT609241S  PNT609242S  PNT609243S  PNT609244S  PNT609246S  PNT609249S  PNT609250S  PNT609251S  PNT609252S  PNT609253S  PNT609255S  PNT609256S  PNT609257S  PNT609258S  PNT609259S  PNT609260S  PNT609262S  PNT609263S  PNT609264S  PNT609267S  PNT609268S  PNT609271S  PNT609283S  PNT609284S  PNT609285S  PNT609286S  PNT609297S  PNT609299S  PNT609302S  PNT609303S  PNT609304S  PNT609305S  PNT609306S  PNT609307S  PNT609308S  PNT609309S  PNT609310S  PNT609311S  PNT609312S  PNT609313S  PNT609314S  PNT609315S  PNT609316S  PNT609317S  PNT609318S  PNT609319S  PNT609320S  PNT609322S  PNT609323S  PNT609324S  PNT609325S  PNT609326S  PNT609327S  PNT609328S  PNT609329S  PNT609330S  PNT609331S  PNT609332S  PNT609333S  PNT609334S  PNT609335S  PNT609336S  PNT609337S  PNT609338S  PNT609339S  PNT609340S  PNT609341S  PNT609342S  PNT609343S  PNT609344S  PNT609345S  PNT609346S  PNT609347S  PNT609350S  PNT609351S  PNT609353S  PNT609354S  PNT609355S  PNT609356S  PNT609357S  PNT609358S  PNT609359S  PNT609360S  PNT609361S  PNT609362S  PNT609363S  PNT609364S  PNT609365S  PNT609366S  PNT609367S  PNT609368S  PNT609369S  PNT609370S  PNT609371S  PNT609372S  PNT609373S  PNT609374S  PNT609375S  PNT609376S  PNT609377S  PNT609378S  PNT609379S  PNT609380S  PNT609381S  PNT609382S  PNT609383S  PNT609384S  PNT609385S  PNT609386S  PNT609387S  PNT609388S  PNT609389S  PNT609390S  PNT609391S  PNT609392S  PNT609393S  PNT609394S  PNT609395S  PNT609396S  PNT609397S  PNT609398S  PNT609400S  PNT609401S  PNT609402S  PNT609403S  PNT609404S  PNT609405S  PNT609406S  PNT609407S  PNT609408S  PNT609409S  PNT609410S  PNT609419S  PNT609424S  PNT609468S  PNT609469S  PNT609470S  PNT609472S  PNT609473S  PNT609475S  PNT609476S  PNT609477S  PNT609478S  PNT609480S  PNT609481S  PNT609482S  PNT609487S  PNT609488S  PNT609489S  PNT609499S  PNT609503S  PNT609508S  PNT609510S  PNT609511S  PNT609515S  PNT609517S  PNT609518S  PNT609519S  PNT609520S  PNT609521S  PNT609526S  PNT609531S  PNT609533S  PNT609534S  PNT609535S  PNT609536S  PNT609537S  PNT609538S  PNT609539S  PNT609540S  PNT609541S  PNT609542S  PNT609543S  PNT609544S  PNT609545S  PNT609546S  PNT609547S  PNT609548S  PNT609549S  PNT609550S  PNT609551S  PNT609554S  PNT609555S  PNT609556S  PNT609557S  PNT609558S  PNT609559S  PNT609560S  PNT609561S  PNT609562S  PNT609563S  PNT609565S  PNT609566S  PNT609567S  PNT609568S  PNT609569S  PNT609570S  PNT609571S  PNT609572S  PNT609573S  PNT609574S  PNT609575S  PNT609576S  PNT609577S  PNT609580S  PNT609584S  PNT609586S  PNT609588S  PNT609589S  PNT609591S  PNT609592S  PNT609593S  PNT609594S  PNT609595S  PNT609596S  PNT609599S  PNT609604S  PNT609605S  PNT609608S  PNT609612S  PNT609617S  PNT609618S  PNT609619S  PNT609622S  PNT609623S  PNT609632S  PNT609642S  PNT609650S  PNT609651S  PNT609670S  PNT609675S  PNT610280S  PNT610284S  PNT610297S  PNT610373S  PNT610374S  PNT610375S  PNT610376S  PNT610377S  PNT610378S  PNT610379S  PNT610380S  PNT610381S  PNT610382S  PNT610383S  PNT610384S  PNT610385S  PNT610386S  PNT610387S  PNT610388S  PNT610389S  PNT610390S  PNT610391S  PNT610392S  PNT610393S  PNT610395S  PNT610396S  PNT610397S  PNT610398S  PNT610399S  PNT610400S  PNT610401S  PNT610402S  PNT610403S  PNT610404S  PNT610406S  PNT610407S  PNT610408S  PNT610410S  PNT610411S  PNT610412S  PNT610414S  PNT610415S  PNT610416S  PNT610417S  PNT610418S  PNT610419S  PNT610420S  PNT610421S  PNT610422S  PNT610423S  PNT610424S  PNT610425S  PNT610426S  PNT610427S  PNT610437S  PNT610438S  PNT610439S  PNT610440S  PNT610441S  PNT610442S  PNT610443S  PNT610444S  PNT610445S  PNT610446S  PNT610447S  PNT610448S  PNT610449S  PNT610450S  PNT610451S  PNT610452S  PNT610453S  PNT610454S  PNT610455S  PNT610456S  PNT610457S  PNT610458S  PNT610459S  PNT610460S  PNT610461S  PNT610462S  PNT610463S  PNT610464S  PNT610465S  PNT610466S  PNT610467S  PNT610468S  PNT610469S  PNT610470S  PNT610471S  PNT610472S  PNT610474S  PNT610475S  PNT610476S  PNT610477S  PNT610478S  PNT610479S  PNT610480S  PNT610481S  PNT610482S  PNT610483S  PNT610484S  PNT610485S  PNT610486S  PNT610487S  PNT610488S  PNT610489S  PNT610490S  PNT610491S  PNT610492S  PNT610493S  PNT610494S  PNT610495S  PNT610496S  PNT610497S  PNT610498S  PNT610499S  PNT610500S  PNT610501S  PNT610502S  PNT610503S  PNT610504S  PNT610505S  PNT610506S  PNT610507S  PNT610508S  PNT610509S  PNT610510S  PNT610511S  PNT610512S  PNT610513S  PNT610514S  PNT610515S  PNT610516S  PNT610517S  PNT610518S  PNT610519S  PNT610520S  PNT610521S  PNT610523S  PNT610524S  PNT610525S  PNT610526S  PNT610527S  PNT610528S  PNT610529S  PNT610530S  PNT610531S  PNT610532S  PNT610533S  PNT610534S  PNT610535S  PNT610536S  PNT610537S  PNT610538S  PNT610539S  PNT610540S  PNT610541S  PNT610542S  PNT610543S  PNT610549S  PNT610550S  PNT610551S  PNT610552S  PNT610553S  PNT610557S  PNT610558S  PNT610559S  PNT610560S  PNT610562S  PNT610563S  PNT610564S  PNT610565S  PNT610566S  PNT610567S  PNT610568S  PNT610569S  PNT610570S  PNT610571S  PNT610572S  PNT610573S  PNT610574S  PNT610575S  PNT610576S  PNT610577S  PNT610578S  PNT610579S  PNT610580S  PNT610581S  PNT610582S  PNT610583S  PNT610584S  PNT610585S  PNT610586S  PNT610587S  PNT610588S  PNT610589S  PNT610590S  PNT610591S  PNT610592S  PNT610593S  PNT610594S  PNT610596S  PNT610597S  PNT610598S  PNT610599S  PNT610600S  PNT610601S  PNT610602S  PNT610603S  PNT610604S  PNT610617S  PNT610619S  PNT610621S  PNT610622S  PNT610623S  PNT610627S  PNT610628S  PNT610630S  PNT610631S  PNT610632S  PNT610634S  PNT610637S  PNT610638S  PNT610639S  PNT610640S  PNT610641S  PNT610642S  PNT610656S  PNT610657S  PNT610662S  PNT610664S  PNT610666S  PNT610667S  PNT610668S  PNT610669S  PNT610670S  PNT610671S  PNT610672S  PNT610673S  PNT610674S  PNT610675S  PNT610676S  PNT610677S  PNT610678S  PNT610679S  PNT610680S  PNT610681S  PNT610682S  PNT610683S  PNT610684S  PNT610686S  PNT610687S  PNT610688S  PNT610689S  PNT610690S  PNT610691S  PNT610692S  PNT610693S  PNT610694S  PNT610695S  PNT610696S  PNT610697S  PNT610698S  PNT610699S  PNT610700S  PNT610701S  PNT610702S  PNT610703S  PNT610704S  PNT610705S  PNT610706S  PNT610707S  PNT610708S  PNT610709S  PNT610710S  PNT610711S  PNT610712S  PNT610713S  PNT610714S  PNT610715S  PNT610716S  PNT610717S  PNT610718S  PNT610719S  PNT610720S  PNT610722S  PNT610723S  PNT610724S  PNT610725S  PNT610726S  PNT610727S  PNT610728S  PNT610729S  PNT610730S  PNT610744S  PNT610745S  PNT610746S  PNT610747S  PNT610748S  PNT610749S  PNT610750S  PNT610751S  PNT610752S  PNT610753S  PNT610754S  PNT610755S  PNT610756S  PNT610759S  PNT610760S  PNT610761S  PNT610762S  PNT610763S  PNT610764S  PNT610765S  PNT610768S  PNT610769S  PNT610770S  PNT610771S  PNT610772S  PNT610773S  PNT610774S  PNT610775S  PNT610776S  PNT610777S  PNT610778S  PNT610779S  PNT610780S  PNT610781S  PNT610782S  PNT610783S  PNT610784S  PNT610785S  PNT610786S  PNT610787S  PNT610788S  PNT610789S  PNT610790S  PNT610791S  PNT610792S  PNT610793S  PNT610794S  PNT610795S  PNT610796S  PNT610797S  PNT610798S  PNT610799S  PNT610800S  PNT610801S  PNT610802S  PNT610804S  PNT610805S  PNT610806S  PNT610807S  PNT610808S  PNT610809S  PNT610810S  PNT610831S  PNT610832S  PNT610833S  PNT610834S  PNT610835S  PNT610836S  PNT610837S  PNT610839S  PNT610841S  PNT610842S  PNT610845S  PNT610848S  PNT610849S  PNT610850S  PNT610852S  PNT610853S  PNT610854S  PNT610855S  PNT610856S  PNT610859S  PNT610862S  PNT610865S  PNT610867S  PNT610869S  PNT610870S  PNT610871S  PNT610872S  PNT610873S  PNT610875S  PNT610876S  PNT610877S  PNT610878S  PNT610879S  PNT610880S  PNT610881S  PNT610882S  PNT610883S  PNT610884S  PNT610885S  PNT610886S  PNT610887S  PNT610888S  PNT610889S  PNT610890S  PNT610891S  PNT610892S  PNT610893S  PNT610894S  PNT610895S  PNT610896S  PNT610897S  PNT610898S  PNT610899S  PNT610900S  PNT610901S  PNT610902S  PNT610903S  PNT610904S  PNT610905S  PNT610906S  PNT610907S  PNT610908S  PNT610909S  PNT610910S  PNT610911S  PNT610912S  PNT610913S  PNT610914S  PNT610915S  PNT610916S  PNT610917S  PNT610918S  PNT610919S  PNT610920S  PNT610921S  PNT610922S  PNT610923S  PNT610924S  PNT610925S  PNT610926S  PNT610927S  PNT610928S  PNT610929S  PNT610930S  PNT610931S  PNT610932S  PNT610933S  PNT610934S  PNT610935S  PNT610936S  PNT610937S  PNT610938S  PNT610939S  PNT610940S  PNT610941S  PNT610942S  PNT610944S  PNT610945S  PNT610946S  PNT610947S  PNT610948S  PNT610949S  PNT610950S  PNT610951S  PNT610952S  PNT610953S  PNT610954S  PNT610955S  PNT610956S  PNT610957S  PNT610958S  PNT610960S  PNT610961S  PNT610962S  PNT610963S  PNT610964S  PNT610965S  PNT610967S  PNT610968S  PNT610969S  PNT610970S  PNT610971S  PNT610972S  PNT610973S  PNT610974S  PNT610975S  PNT610976S  PNT610977S  PNT610978S  PNT610979S  PNT610980S  PNT610981S  PNT610982S  PNT610983S  PNT610984S  PNT610985S  PNT610986S  PNT610987S  PNT610988S  PNT610989S  PNT610990S  PNT610991S  PNT610992S  PNT610993S  PNT610994S  PNT610995S  PNT610996S  PNT610997S  PNT610998S  PNT610999S  PNT611000S  PNT611001S  PNT611002S  PNT611003S  PNT611004S  PNT611005S  PNT611006S  PNT611007S  PNT611008S  PNT611009S  PNT611010S  PNT611011S  PNT611012S  PNT611014S  PNT611015S  PNT611016S  PNT611017S  PNT611018S  PNT611019S  PNT611020S  PNT611021S  PNT611022S  PNT611023S  PNT611024S  PNT611025S  PNT611026S  PNT611027S  PNT611028S  PNT611029S  PNT611030S  PNT611031S  PNT611032S  PNT611033S  PNT611034S  PNT611035S  PNT611036S  PNT611037S  PNT611038S  PNT611039S  PNT611040S  PNT611041S  PNT611042S  PNT611043S  PNT611044S  PNT611045S  PNT611046S  PNT611047S  PNT611048S  PNT611049S  PNT611050S  PNT611051S  PNT611052S  PNT611053S  PNT611054S  PNT611055S  PNT611056S  PNT611057S  PNT611058S  PNT611059S  PNT611060S  PNT611061S  PNT611062S  PNT611063S  PNT611064S  PNT611065S  PNT611066S  PNT611067S  PNT611068S  PNT611069S  PNT611070S  PNT611071S  PNT611072S  PNT611073S  PNT611074S  PNT611075S  PNT611076S  PNT611077S  PNT611078S  PNT611079S  PNT611080S  PNT611081S  PNT611082S  PNT611083S  PNT611084S  PNT611085S  PNT611086S  PNT611087S  PNT611088S  PNT611089S  PNT611090S  PNT611091S  PNT611092S  PNT611093S  PNT611094S  PNT611095S  PNT611096S  PNT611097S  PNT611098S  PNT611099S  PNT611100S  PNT611101S  PNT611102S  PNT611103S  PNT611104S  PNT611105S  PNT611106S  PNT611107S  PNT611108S  PNT611109S  PNT611110S  PNT611111S  PNT611112S  PNT611113S  PNT611114S  PNT611115S  PNT611116S  PNT611117S  PNT611118S  PNT611119S  PNT611120S  PNT611121S  PNT611122S  PNT611123S  PNT611124S  PNT611125S  PNT611126S  PNT611127S  PNT611128S  PNT611129S  PNT611130S  PNT611131S  PNT611132S  PNT611133S  PNT611134S  PNT611135S  PNT611136S  PNT611137S  PNT611138S  PNT611139S  PNT611140S  PNT611141S  PNT611142S  PNT611143S  PNT611144S  PNT611145S  PNT611146S  PNT611147S  PNT611148S  PNT611149S  PNT611150S  PNT611151S  PNT611152S  PNT611153S  PNT611154S  PNT611155S  PNT611156S  PNT611157S  PNT611158S  PNT611159S  PNT611160S  PNT611161S  PNT611162S  PNT611163S  PNT611164S  PNT611165S  PNT611166S  PNT611168S  PNT611169S  PNT611170S  PNT611171S  PNT611172S  PNT611173S  PNT611174S  PNT611175S  PNT611176S  PNT611177S  PNT611178S  PNT611179S  PNT611180S  PNT611181S  PNT611182S  PNT611183S  PNT611184S  PNT611185S  PNT611187S  PNT611188S  PNT611189S  PNT611190S  PNT611191S  PNT611192S  PNT611193S  PNT611194S  PNT611195S  PNT611196S  PNT611197S  PNT611198S  PNT611199S  PNT611200S  PNT611201S  PNT611202S  PNT611203S  PNT611204S  PNT611205S  PNT611206S  PNT611207S  PNT611208S  PNT611209S  PNT611210S  PNT611211S  PNT611212S  PNT611213S  PNT611214S  PNT611215S  PNT611216S  PNT611217S  PNT611218S  PNT611219S  PNT611220S  PNT611221S  PNT611222S  PNT611223S  PNT611224S  PNT611225S  PNT611226S  PNT611227S  PNT611228S  PNT611229S  PNT611230S  PNT611231S  PNT611232S  PNT611233S  PNT611234S  PNT611235S  PNT611236S  PNT611237S  PNT611238S  PNT611239S  PNT611240S  PNT611241S  PNT611242S  PNT611243S  PNT611244S  PNT611245S  PNT611246S  PNT611247S  PNT611248S  PNT611249S  PNT611250S  PNT611251S  PNT611252S  PNT611253S  PNT611254S  PNT611255S  PNT611256S  PNT611257S  PNT611258S  PNT611259S  PNT611260S  PNT611261S  PNT611262S  PNT611263S  PNT611264S  PNT611265S  PNT611266S  PNT611267S  PNT611268S  PNT611269S  PNT611270S  PNT611271S  PNT611272S  PNT611273S  PNT611274S  PNT611275S  PNT611276S  PNT611277S  PNT611278S  PNT611279S  PNT611280S  PNT611281S  PNT611282S  PNT611283S  PNT611284S  PNT611286S  PNT611287S  PNT611288S  PNT611289S  PNT611290S  PNT611291S  PNT611292S  PNT611293S  PNT611294S  PNT611295S  PNT611296S  PNT611297S  PNT611298S  PNT611299S  PNT611300S  PNT611301S  PNT611302S  PNT611303S  PNT611305S  PNT611306S  PNT611307S  PNT611308S  PNT611309S  PNT611310S  PNT611311S  PNT611312S  PNT611314S  PNT611315S  PNT611316S  PNT611317S  PNT611318S  PNT611319S  PNT611320S  PNT611321S  PNT611322S  PNT611323S  PNT611324S  PNT611325S  PNT611326S  PNT611327S  PNT611328S  PNT611329S  PNT611330S  PNT611331S  PNT611332S  PNT611333S  PNT611334S  PNT611335S  PNT611336S  PNT611337S  PNT611338S  PNT611339S  PNT611340S  PNT611341S  PNT611342S  PNT611343S  PNT611344S  PNT611345S  PNT611346S  PNT611348S  PNT611349S  PNT611350S  PNT611351S  PNT611352S  PNT611353S  PNT611354S  PNT611355S  PNT611356S  PNT611357S  PNT611358S  PNT611359S  PNT611360S  PNT611362S  PNT611363S  PNT611364S  PNT611365S  PNT611366S  PNT611367S  PNT611368S  PNT611369S  PNT611370S  PNT611371S  PNT611372S  PNT611373S  PNT611375S  PNT611376S  PNT611377S  PNT611378S  PNT611380S  PNT611381S  PNT611382S  PNT611383S  PNT611384S  PNT611385S  PNT611386S  PNT611387S  PNT611388S  PNT611389S  PNT611390S  PNT611391S  PNT611392S  PNT611393S  PNT611394S  PNT611395S  PNT611396S  PNT611397S  PNT611398S  PNT611399S  PNT611400S  PNT611401S  PNT611402S  PNT611403S  PNT611404S  PNT611405S  PNT611406S  PNT611407S  PNT611408S  PNT611409S  PNT611410S  PNT611411S  PNT611412S  PNT611413S  PNT611414S  PNT611415S  PNT611416S  PNT611417S  PNT611418S  PNT611419S  PNT611420S  PNT611421S  PNT611422S  PNT611423S  PNT611424S  PNT611425S  PNT611426S  PNT611427S  PNT611428S  PNT611429S  PNT611430S  PNT611431S  PNT611432S  PNT611433S  PNT611434S  PNT611435S  PNT611436S  PNT611437S  PNT611438S  PNT611439S  PNT611440S  PNT611441S  PNT611442S  PNT611443S  PNT611444S  PNT611445S  PNT611446S  PNT611447S  PNT611448S  PNT611449S  PNT611450S  PNT611451S  PNT611452S  PNT611453S  PNT611454S  PNT611455S  PNT611456S  PNT611457S  PNT611458S  PNT611459S  PNT611460S  PNT611461S  PNT611462S  PNT611463S  PNT611464S  PNT611465S  PNT611466S  PNT611467S  PNT611469S  PNT611470S  PNT611471S  PNT611472S  PNT611473S  PNT611474S  PNT611475S  PNT611476S  PNT611477S  PNT611478S  PNT611479S  PNT611480S  PNT611481S  PNT611482S  PNT611483S  PNT611484S  PNT611485S  PNT611486S  PNT611487S  PNT611488S  PNT611489S  PNT611490S  PNT611491S  PNT611492S  PNT611493S  PNT611494S  PNT611495S  PNT611496S  PNT611497S  PNT611499S  PNT611500S  PNT611501S  PNT611502S  PNT611503S  PNT611504S  PNT611505S  PNT611507S  PNT611508S  PNT611509S  PNT611510S  PNT611511S  PNT611512S  PNT611513S  PNT611514S  PNT611515S  PNT611516S  PNT611518S  PNT611519S  PNT611520S  PNT611521S  PNT611522S  PNT611523S  PNT611524S  PNT611525S  PNT611526S  PNT611527S  PNT611528S  PNT611529S  PNT611530S  PNT611532S  PNT611533S  PNT611534S  PNT611535S  PNT611536S  PNT611537S  PNT611538S  PNT611539S  PNT611540S  PNT611541S  PNT611542S  PNT611543S  PNT611544S  PNT611545S  PNT611546S  PNT611547S  PNT611548S  PNT611549S  PNT611550S  PNT611551S  PNT611552S  PNT611553S  PNT611554S  PNT611555S  PNT611556S  PNT611557S  PNT611558S  PNT611559S  PNT611560S  PNT611561S  PNT611562S  PNT611563S  PNT611564S  PNT611565S  PNT611566S  PNT611567S  PNT611568S  PNT611569S  PNT611570S  PNT611571S  PNT611572S  PNT611573S  PNT611574S  PNT611575S  PNT611576S  PNT611577S  PNT611578S  PNT611579S  PNT611580S  PNT611581S  PNT611582S  PNT611583S  PNT611584S  PNT611585S  PNT611586S  PNT611587S  PNT611588S  PNT611589S  PNT611590S  PNT611591S  PNT611592S  PNT611593S  PNT611594S  PNT611595S  PNT611596S  PNT611597S  PNT611598S  PNT611599S  PNT611600S  PNT611601S  PNT611602S  PNT611603S  PNT611604S  PNT611605S  PNT611607S  PNT611608S  PNT611609S  PNT611611S  PNT611612S  PNT611613S  PNT611614S  PNT611615S  PNT611616S  PNT611617S  PNT611618S  PNT611619S  PNT611620S  PNT611621S  PNT611622S  PNT611623S  PNT611627S  PNT611655S  PNT611656S  PNT611657S  PNT611658S  PNT611659S  PNT611660S  PNT611661S  PNT611662S  PNT611663S  PNT611664S  PNT611666S  PNT611667S  PNT611668S  PNT611670S  PNT611671S  PNT611677S  PNT611682S  PNT611683S  PNT611684S  PNT611685S  PNT611686S  PNT611687S  PNT611688S  PNT611689S  PNT611690S  PNT611691S  PNT611692S  PNT611693S  PNT611694S  PNT611695S  PNT611696S  PNT611697S  PNT611698S  PNT611699S  PNT611700S  PNT611701S  PNT611702S  PNT611703S  PNT611704S  PNT611706S  PNT611707S  PNT611708S  PNT611709S  PNT611711S  PNT611712S  PNT611713S  PNT611714S  PNT611715S  PNT611716S  PNT611717S  PNT611718S  PNT611719S  PNT611720S  PNT611721S  PNT611722S  PNT611723S  PNT611724S  PNT611725S  PNT611726S  PNT611727S  PNT611728S  PNT611733S  PNT611734S  PNT611736S  PNT611746S  PNT611747S  PNT611748S  PNT611767S  PNT611768S  PNT611769S  PNT611770S  PNT611771S  PNT611772S  PNT611773S  PNT611774S  PNT611775S  PNT611776S  PNT611777S  PNT611778S  PNT611779S  PNT611780S  PNT611781S  PNT611782S  PNT611783S  PNT611784S  PNT611785S  PNT611786S  PNT611787S  PNT611788S  PNT611789S  PNT611790S  PNT611791S  PNT611795S  PNT611796S  PNT611797S  PNT611798S  PNT611799S  PNT611800S  PNT611801S  PNT611802S  PNT611803S  PNT611804S  PNT611805S  PNT611806S  PNT611807S  PNT611808S  PNT611809S  PNT611810S  PNT611811S  PNT611812S  PNT611814S  PNT611815S  PNT611816S  PNT611817S  PNT611818S  PNT611819S  PNT611820S  PNT611821S  PNT611822S  PNT611823S  PNT611824S  PNT611825S  PNT611826S  PNT611827S  PNT611828S  PNT611829S  PNT611830S  PNT611831S  PNT611832S  PNT611833S  PNT611834S  PNT611835S  PNT611836S  PNT611837S  PNT611838S  PNT611839S  PNT611840S  PNT611841S  PNT611842S  PNT611843S  PNT611844S  PNT611845S  PNT611846S  PNT611847S  PNT611848S  PNT611850S  PNT611851S  PNT611852S  PNT611853S  PNT611854S  PNT611855S  PNT611856S  PNT611857S  PNT611858S  PNT611859S  PNT611860S  PNT611861S  PNT611862S  PNT611863S  PNT611864S  PNT611865S  PNT611866S  PNT611867S  PNT611868S  PNT611869S  PNT611871S  PNT611872S  PNT611873S  PNT611874S  PNT611875S  PNT611876S  PNT611877S  PNT611878S  PNT611879S  PNT611880S  PNT611881S  PNT611883S  PNT611884S  PNT611886S  PNT611888S  PNT611889S  PNT611890S  PNT611891S  PNT611892S  PNT611893S  PNT611894S  PNT611896S  PNT611897S  PNT611898S  PNT611900S  PNT611902S  PNT611903S  PNT611907S  PNT611908S  PNT611909S  PNT611911S  PNT611912S  PNT611914S  PNT611916S  PNT611917S  PNT611918S  PNT611919S  PNT611922S  PNT611924S  PNT611925S  PNT611926S  PNT611927S  PNT611929S  PNT611930S  PNT611931S  PNT611932S  PNT611934S  PNT611935S  PNT611936S  PNT611937S  PNT611938S  PNT611939S  PNT611940S  PNT611941S  PNT611942S  PNT611943S  PNT611944S  PNT611945S  PNT611946S  PNT611947S  PNT611948S  PNT611949S  PNT611950S  PNT611951S  PNT611952S  PNT611953S  PNT611954S  PNT611955S  PNT611956S  PNT611957S  PNT611958S  PNT611959S  PNT611960S  PNT611961S  PNT611973S  PNT611974S  PNT611975S  PNT611977S  PNT611978S  PNT611979S  PNT611980S  PNT611981S  PNT611982S  PNT611983S  PNT611984S  PNT611985S  PNT611986S  PNT611987S  PNT611988S  PNT611989S  PNT611990S  PNT611991S  PNT611992S  PNT611993S  PNT611994S  PNT611995S  PNT611996S  PNT611997S  PNT611998S  PNT611999S  PNT612000S  PNT612001S  PNT612002S  PNT612003S  PNT612004S  PNT612005S  PNT612006S  PNT612007S  PNT612008S  PNT612009S  PNT612010S  PNT612011S  PNT612012S  PNT612013S  PNT612014S  PNT612015S  PNT612016S  PNT612017S  PNT612018S  PNT612019S  PNT612020S  PNT612021S  PNT612022S  PNT612023S  PNT612024S  PNT612025S  PNT612026S  PNT612027S  PNT612028S  PNT612029S  PNT612030S  PNT612066S  PNT612067S  PNT612068S  PNT612069S  PNT612070S  PNT612071S  PNT612072S  PNT612073S  PNT612074S  PNT612075S  PNT612077S  PNT612078S  PNT612079S  PNT612080S  PNT612081S  PNT612082S  PNT612083S  PNT612084S  PNT612085S  PNT612086S  PNT612087S  PNT612088S  PNT612089S  PNT612090S  PNT612091S  PNT612092S  PNT612093S  PNT612094S  PNT612095S  PNT612096S  PNT612097S  PNT612098S  PNT612099S  PNT612101S  PNT612102S  PNT612103S  PNT612104S  PNT612105S  PNT612106S  PNT612107S  PNT612108S  PNT612109S  PNT612110S  PNT612111S  PNT612112S  PNT612113S  PNT612114S  PNT612115S  PNT612116S  PNT612117S  PNT612118S  PNT612119S  PNT612120S  PNT612121S  PNT612122S  PNT612123S  PNT612124S  PNT612125S  PNT612126S  PNT612127S  PNT612128S  PNT612129S  PNT612130S  PNT612131S  PNT612132S  PNT612133S  PNT612134S  PNT612135S  PNT612136S  PNT612137S  PNT612138S  PNT612139S  PNT612140S  PNT612142S  PNT612143S  PNT612144S  PNT612145S  PNT612146S  PNT612147S  PNT612148S  PNT612149S  PNT612150S  PNT612151S  PNT612152S  PNT612153S  PNT612154S  PNT612155S  PNT612156S  PNT612157S  PNT612158S  PNT612159S  PNT612160S  PNT612163S  PNT612164S  PNT612166S  PNT612169S  PNT612170S  PNT612171S  PNT612172S  PNT612173S  PNT612174S  PNT612175S  PNT612176S  PNT612177S  PNT612178S  PNT612179S  PNT612180S  PNT612181S  PNT612182S  PNT612183S  PNT612184S  PNT612185S  PNT612186S  PNT612187S  PNT612188S  PNT612189S  PNT612190S  PNT612192S  PNT612193S  PNT612194S  PNT612195S  PNT612196S  PNT612197S  PNT612198S  PNT612199S  PNT612201S  PNT612202S  PNT612203S  PNT612204S  PNT612205S  PNT612206S  PNT612207S  PNT612208S  PNT612209S  PNT612210S  PNT612211S  PNT612212S  PNT612213S  PNT612214S  PNT612215S  PNT612216S  PNT612217S  PNT612218S  PNT612219S  PNT612220S  PNT612221S  PNT612222S  PNT612224S  PNT612225S  PNT612226S  PNT612227S  PNT612228S  PNT612229S  PNT612230S  PNT612231S  PNT612232S  PNT612233S  PNT612234S  PNT612235S  PNT612236S  PNT612237S  PNT612238S  PNT612239S  PNT612240S  PNT612241S  PNT612242S  PNT612243S  PNT612244S  PNT612245S  PNT612246S  PNT612251S  PNT612253S  PNT612254S  PNT612255S  PNT612257S  PNT612258S  PNT612259S  PNT612260S  PNT612263S  PNT612264S  PNT612266S  PNT612269S  PNT612270S  PNT612271S  PNT612273S  PNT612275S  PNT612276S  PNT612278S  PNT612279S  PNT612280S  PNT612281S  PNT612282S  PNT612283S  PNT612285S  PNT612286S  PNT612287S  PNT612288S  PNT612291S  PNT612293S  PNT612294S  PNT612295S  PNT612296S  PNT612297S  PNT612299S  PNT612301S  PNT612304S  PNT612305S  PNT612307S  PNT612309S  PNT612310S  PNT612311S  PNT612312S  PNT612313S  PNT612314S  PNT612315S  PNT612316S  PNT612317S  PNT612318S  PNT612319S  PNT612320S  PNT612321S  PNT612322S  PNT612323S  PNT612324S  PNT612325S  PNT612326S  PNT612327S  PNT612328S  PNT612329S  PNT612330S  PNT612331S  PNT612332S  PNT612333S  PNT612334S  PNT612335S  PNT612336S  PNT612337S  PNT612338S  PNT612339S  PNT612340S  PNT612342S  PNT612343S  PNT612344S  PNT612345S  PNT612346S  PNT612347S  PNT612348S  PNT612349S  PNT612350S  PNT612351S  PNT612352S  PNT612353S  PNT612354S  PNT612355S  PNT612356S  PNT612357S  PNT612358S  PNT612359S  PNT612360S  PNT612361S  PNT612362S  PNT612363S  PNT612364S  PNT612365S  PNT612366S  PNT612367S  PNT612368S  PNT612369S  PNT612373S  PNT612376S  PNT612377S  PNT612378S  PNT612379S  PNT612380S  PNT612381S  PNT612382S  PNT612383S  PNT612385S  PNT612386S  PNT612388S  PNT612389S  PNT612390S  PNT612392S  PNT612393S  PNT612397S  PNT612399S  PNT612401S  PNT612402S  PNT612403S  PNT612404S  PNT612405S  PNT612407S  PNT612413S  PNT612414S  PNT612418S  PNT612419S  PNT612420S  PNT612421S  PNT612422S  PNT612423S  PNT612424S  PNT612426S  PNT612429S  PNT612439S  PNT612449S  PNT612452S  PNT612453S  PNT612455S  PNT612456S  PNT612457S  PNT612458S  PNT612459S  PNT612460S  PNT612461S  PNT612462S  PNT612463S  PNT612464S  PNT612465S  PNT612466S  PNT612467S  PNT612468S  PNT612469S  PNT612470S  PNT612471S  PNT612472S  PNT612473S  PNT612474S  PNT612475S  PNT612476S  PNT612477S  PNT612478S  PNT612479S  PNT612480S  PNT612481S  PNT612482S  PNT612483S  PNT612484S  PNT612485S  PNT612486S  PNT612487S  PNT612488S  PNT612489S  PNT612490S  PNT612491S  PNT612492S  PNT612493S  PNT612494S  PNT612495S  PNT612496S  PNT612497S  PNT612498S  PNT612499S  PNT612500S  PNT612501S  PNT612502S  PNT612503S  PNT612504S  PNT612505S  PNT612506S  PNT612507S  PNT612508S  PNT612509S  PNT612510S  PNT612511S  PNT612512S  PNT612513S  PNT612514S  PNT612515S  PNT612516S  PNT612517S  PNT612518S  PNT612519S  PNT612520S  PNT612521S  PNT612522S  PNT612523S  PNT612524S  PNT612525S  PNT612526S  PNT612527S  PNT612528S  PNT612529S  PNT612530S  PNT612531S  PNT612532S  PNT612533S  PNT612534S  PNT612535S  PNT612536S  PNT612537S  PNT612538S  PNT612539S  PNT612540S  PNT612541S  PNT612542S  PNT612543S  PNT612544S  PNT612545S  PNT612546S  PNT612547S  PNT612548S  PNT612549S  PNT612550S  PNT612551S  PNT612552S  PNT612553S  PNT612554S  PNT612555S  PNT612556S  PNT612557S  PNT612558S  PNT612559S  PNT612560S  PNT612561S  PNT612562S  PNT612563S  PNT612564S  PNT612565S  PNT612566S  PNT612567S  PNT612568S  PNT612569S  PNT612570S  PNT612571S  PNT612572S  PNT612573S  PNT612574S  PNT612575S  PNT612576S  PNT612577S  PNT612578S  PNT612579S  PNT612580S  PNT612581S  PNT612582S  PNT612583S  PNT612584S  PNT612585S  PNT612586S  PNT612587S  PNT612588S  PNT612589S  PNT612590S  PNT612591S  PNT612592S  PNT612593S  PNT612594S  PNT612595S  PNT612596S  PNT612597S  PNT612598S  PNT612599S  PNT612600S  PNT612601S  PNT612602S  PNT612603S  PNT612604S  PNT612605S  PNT612606S  PNT612607S  PNT612608S  PNT612609S  PNT612611S  PNT612612S  PNT612613S  PNT612614S  PNT612615S  PNT612616S  PNT612617S  PNT612619S  PNT612620S  PNT612621S  PNT612622S  PNT612623S  PNT612624S  PNT612625S  PNT612626S  PNT612627S  PNT612628S  PNT612629S  PNT612630S  PNT612631S  PNT612632S  PNT612633S  PNT612634S  PNT612635S  PNT612636S  PNT612637S  PNT612638S  PNT612639S  PNT612640S  PNT612641S  PNT612642S  PNT612643S  PNT612644S  PNT612645S  PNT612646S  PNT612647S  PNT612648S  PNT612649S  PNT612650S  PNT612651S  PNT612652S  PNT612653S  PNT612654S  PNT612655S  PNT612656S  PNT612657S  PNT612658S  PNT612659S  PNT612660S  PNT612661S  PNT612662S  PNT612663S  PNT612664S  PNT612665S  PNT612666S  PNT612667S  PNT612668S  PNT612669S  PNT612670S  PNT612671S  PNT612672S  PNT612674S  PNT612675S  PNT612676S  PNT612677S  PNT612678S  PNT612679S  PNT612680S  PNT612681S  PNT612682S  PNT612683S  PNT612684S  PNT612685S  PNT612686S  PNT612687S  PNT612688S  PNT612689S  PNT612690S  PNT612691S  PNT612692S  PNT612693S  PNT612694S  PNT612695S  PNT612696S  PNT612697S  PNT612698S  PNT612699S  PNT612700S  PNT612701S  PNT612702S  PNT612703S  PNT612704S  PNT612705S  PNT612706S  PNT612707S  PNT612708S  PNT612709S  PNT612710S  PNT612711S  PNT612712S  PNT612714S  PNT612715S  PNT612716S  PNT612717S  PNT612720S  PNT612721S  PNT612722S  PNT612723S  PNT612724S  PNT612725S  PNT612726S  PNT612727S  PNT612728S  PNT612730S  PNT612731S  PNT612732S  PNT612733S  PNT612734S  PNT612735S  PNT612736S  PNT612737S  PNT612738S  PNT612739S  PNT612740S  PNT612741S  PNT612742S  PNT612743S  PNT612744S  PNT612745S  PNT612746S  PNT612747S  PNT612748S  PNT612749S  PNT612750S  PNT612751S  PNT612753S  PNT612754S  PNT612755S  PNT612756S  PNT612757S  PNT612758S  PNT612759S  PNT612760S  PNT612761S  PNT612762S  PNT612763S  PNT612764S  PNT612765S  PNT612766S  PNT612767S  PNT612768S  PNT612769S  PNT612770S  PNT612771S  PNT612772S  PNT612773S  PNT612774S  PNT612776S  PNT612777S  PNT612778S  PNT612779S  PNT612780S  PNT612781S  PNT612782S  PNT612783S  PNT612784S  PNT612785S  PNT612786S  PNT612787S  PNT612788S  PNT612789S  PNT612790S  PNT612791S  PNT612792S  PNT612793S  PNT612794S  PNT612795S  PNT612796S  PNT612797S  PNT612798S  PNT612799S  PNT612800S  PNT612801S  PNT612802S  PNT612803S  PNT612804S  PNT612805S  PNT612806S  PNT612807S  PNT612808S  PNT612809S  PNT612810S  PNT612811S  PNT612812S  PNT612813S  PNT612814S  PNT612816S  PNT612817S  PNT612819S  PNT612820S  PNT612821S  PNT612822S  PNT612823S  PNT612824S  PNT612825S  PNT612826S  PNT612827S  PNT612828S  PNT612829S  PNT612830S  PNT612831S  PNT612832S  PNT612833S  PNT612834S  PNT612835S  PNT612836S  PNT612837S  PNT612838S  PNT612839S  PNT612840S  PNT612841S  PNT612842S  PNT612843S  PNT612844S  PNT612845S  PNT612846S  PNT612847S  PNT612848S  PNT612849S  PNT612850S  PNT612851S  PNT612852S  PNT612853S  PNT612854S  PNT612855S  PNT612856S  PNT612857S  PNT612858S  PNT612859S  PNT612860S  PNT612861S  PNT612862S  PNT612863S  PNT612864S  PNT612865S  PNT612866S  PNT612867S  PNT612869S  PNT612870S  PNT612871S  PNT612872S  PNT612873S  PNT612874S  PNT612875S  PNT612876S  PNT612877S  PNT612878S  PNT612879S  PNT612880S  PNT612881S  PNT612882S  PNT612912S  PNT612913S  PNT612914S  PNT612915S  PNT612916S  PNT612918S  PNT612919S  PNT612920S  PNT612921S  PNT612923S  PNT612924S  PNT612925S  PNT612926S  PNT612927S  PNT612928S  PNT612929S  PNT612930S  PNT612931S  PNT612932S  PNT612933S  PNT612934S  PNT612935S  PNT612936S  PNT612937S  PNT612938S  PNT612939S  PNT612940S  PNT612941S  PNT612943S  PNT612944S  PNT612945S  PNT612947S  PNT612948S  PNT612949S  PNT612950S  PNT612951S  PNT612952S  PNT612953S  PNT612954S  PNT612955S  PNT612956S  PNT612957S  PNT612983S  PNT612985S  PNT612986S  PNT612987S  PNT612988S  PNT612989S  PNT612990S  PNT612991S  PNT612992S  PNT612993S  PNT612994S  PNT612995S  PNT612996S  PNT612997S  PNT612998S  PNT612999S  PNT613000S  PNT613001S  PNT613002S  PNT613003S  PNT613004S  PNT613005S  PNT613006S  PNT613007S  PNT613008S  PNT613009S  PNT613010S  PNT613011S  PNT613012S  PNT613013S  PNT613014S  PNT613015S  PNT613016S  PNT613017S  PNT613018S  PNT613019S  PNT613020S  PNT613021S  PNT613022S  PNT613023S  PNT613024S  PNT613025S  PNT613026S  PNT613033S  PNT613034S  PNT613035S  PNT613036S  PNT613037S  PNT613038S  PNT613039S  PNT613040S  PNT613041S  PNT613043S  PNT613044S  PNT613045S  PNT613046S  PNT613047S  PNT6",
      "center_classification_date": "20180710",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "state": "MN",
      "product_description": "Medtronic EnTrust Escudo D144VRC (OUS distribution)    Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.",
      "report_date": "20180718",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "recall_number": "Z-2420-2018",
      "initial_firm_notification": "Visit",
      "product_type": "Devices",
      "event_id": "80379",
      "termination_date": "20210616",
      "more_code_info": "13048S  PNT613049S  PNT613050S  PNT613051S  PNT613052S  PNT613053S  PNT613054S  PNT613055S  PNT613056S  PNT613057S  PNT613058S  PNT613059S  PNT613060S  PNT613061S  PNT613062S  PNT613063S  PNT613064S  PNT613065S  PNT613066S  PNT613067S  PNT613068S  PNT613069S  PNT613070S  PNT613071S  PNT613073S  PNT613074S  PNT613075S  PNT613076S  PNT613077S  PNT613078S  PNT613079S  PNT613080S  PNT613081S  PNT613082S  PNT613083S  PNT613084S  PNT613085S  PNT613086S  PNT613087S  PNT613088S  PNT613089S  PNT613090S  PNT613091S  PNT613092S  PNT613093S  PNT613094S  PNT613095S  PNT613096S  PNT613097S  PNT613098S  PNT613099S  PNT613100S  PNT613101S  PNT613102S  PNT613103S  PNT613104S  PNT613105S  PNT613106S  PNT613107S  PNT613108S  PNT613109S  PNT613110S  PNT613111S  PNT613112S  PNT613113S  PNT613114S  PNT613115S  PNT613116S  PNT613118S  PNT613119S  PNT613120S  PNT613121S  PNT613122S  PNT613123S  PNT613125S  PNT613126S  PNT613127S  PNT613129S  PNT613130S  PNT613131S  PNT613132S  PNT613133S  PNT613134S  PNT613135S  PNT613136S  PNT613137S  PNT613138S  PNT613139S  PNT613140S  PNT613141S  PNT613142S  PNT613143S  PNT613144S  PNT613145S  PNT613146S  PNT613148S  PNT613149S  PNT613150S  PNT613151S  PNT613152S  PNT613153S  PNT613154S  PNT613155S  PNT613156S  PNT613157S  PNT613158S  PNT613159S  PNT613160S  PNT613161S  PNT613162S  PNT613163S  PNT613164S  PNT613165S  PNT613166S  PNT613167S  PNT613168S  PNT613169S  PNT613170S  PNT613171S  PNT613172S  PNT613173S  PNT613174S  PNT613175S  PNT613176S  PNT613177S  PNT613178S  PNT613179S  PNT613180S  PNT613182S  PNT613183S  PNT613184S  PNT613185S  PNT613186S  PNT613187S  PNT613188S  PNT613189S  PNT613190S  PNT613192S  PNT613193S  PNT613194S  PNT613195S  PNT613196S  PNT613197S  PNT613198S  PNT613199S  PNT613200S  PNT613201S  PNT613202S  PNT613203S  PNT613204S  PNT613205S  PNT613206S  PNT613207S  PNT613208S  PNT613209S  PNT613210S  PNT613211S  PNT613212S  PNT613213S  PNT613214S  PNT613215S  PNT613216S  PNT613217S  PNT613218S  PNT613220S  PNT613221S  PNT613222S  PNT613223S  PNT613224S  PNT613225S  PNT613226S  PNT613227S  PNT613228S  PNT613229S  PNT613230S  PNT613231S  PNT613232S  PNT613233S  PNT613235S  PNT613236S  PNT613237S  PNT613238S  PNT613239S  PNT613240S  PNT613241S  PNT613242S  PNT613243S  PNT613244S  PNT613245S  PNT613246S  PNT613247S  PNT613248S  PNT613249S  PNT613250S  PNT613251S  PNT613252S  PNT613253S  PNT613254S  PNT613255S  PNT613256S  PNT613257S  PNT613258S  PNT613259S  PNT613260S  PNT613261S  PNT613262S  PNT613263S  PNT613264S  PNT613265S  PNT613266S  PNT613267S  PNT613268S  PNT613270S  PNT613271S  PNT613272S  PNT613273S  PNT613274S  PNT613275S  PNT613276S  PNT613277S  PNT613278S  PNT613279S  PNT613280S  PNT613281S  PNT613282S  PNT613283S  PNT613284S  PNT613285S  PNT613286S  PNT613287S  PNT613288S  PNT613289S  PNT613290S  PNT613291S  PNT613292S  PNT613293S  PNT613294S  PNT613295S  PNT613296S  PNT613297S  PNT613298S  PNT613299S  PNT613300S  PNT613301S  PNT613302S  PNT613303S  PNT613304S  PNT613305S  PNT613306S  PNT613307S  PNT613308S  PNT613309S  PNT613310S  PNT613311S  PNT613312S  PNT613313S  PNT613314S  PNT613315S  PNT613316S  PNT613317S  PNT613318S  PNT613319S  PNT613320S  PNT613321S  PNT613322S  PNT613323S  PNT613324S  PNT613325S  PNT613326S  PNT613327S  PNT613328S  PNT613329S  PNT613330S  PNT613331S  PNT613332S  PNT613333S  PNT613334S  PNT613335S  PNT613336S  PNT613337S  PNT613338S  PNT613339S  PNT613340S  PNT613341S  PNT613342S  PNT613343S  PNT613344S  PNT613345S  PNT613346S  PNT613347S  PNT613348S  PNT613349S  PNT613350S  PNT613351S  PNT613352S  PNT613353S  PNT613354S  PNT613355S  PNT613356S  PNT613357S  PNT613358S  PNT613359S  PNT613360S  PNT613361S  PNT613362S  PNT613363S  PNT613364S  PNT613365S  PNT613366S  PNT613367S  PNT613369S  PNT613370S  PNT613371S  PNT613372S  PNT613373S  PNT613374S  PNT613375S  PNT613377S  PNT613378S  PNT613379S  PNT613380S  PNT613381S  PNT613382S  PNT613383S  PNT613384S  PNT613385S  PNT613386S  PNT613387S  PNT613388S  PNT613389S  PNT613390S  PNT613391S  PNT613392S  PNT613393S  PNT613394S  PNT613395S  PNT613396S  PNT613398S  PNT613399S  PNT613401S  PNT613402S  PNT613403S  PNT613404S  PNT613405S  PNT613406S  PNT613407S  PNT613408S  PNT613409S  PNT613410S  PNT613411S  PNT613412S  PNT613413S  PNT613414S  PNT613415S  PNT613416S  PNT613417S  PNT613418S  PNT613419S  PNT613420S  PNT613421S  PNT613422S  PNT613423S  PNT613424S  PNT613425S  PNT613426S  PNT613427S  PNT613428S  PNT613429S  PNT613430S  PNT613431S  PNT613432S  PNT613433S  PNT613434S  PNT613435S  PNT613436S  PNT613437S  PNT613439S  PNT613440S  PNT613441S  PNT613442S  PNT613443S  PNT613444S  PNT613445S  PNT613446S  PNT613447S  PNT613448S  PNT613449S  PNT613450S  PNT613451S  PNT613452S  PNT613453S  PNT613454S  PNT613455S  PNT613456S  PNT613457S  PNT613458S  PNT613460S  PNT613461S  PNT613462S  PNT613463S  PNT613464S  PNT613465S  PNT613466S  PNT613468S  PNT613469S  PNT613470S  PNT613471S  PNT613472S  PNT613473S  PNT613474S  PNT613475S  PNT613476S  PNT613477S  PNT613479S  PNT613480S  PNT613481S  PNT613482S  PNT613483S  PNT613484S  PNT613485S  PNT613486S  PNT613487S  PNT613488S  PNT613489S  PNT613492S  PNT613493S  PNT613494S  PNT613496S  PNT613498S  PNT613499S  PNT613500S  PNT613531S  PNT613532S  PNT613533S  PNT613534S  PNT613535S  PNT613536S  PNT613537S  PNT613538S  PNT613539S  PNT613540S  PNT613541S  PNT613542S  PNT613543S  PNT613544S  PNT613545S  PNT613546S  PNT613547S  PNT613548S  PNT613549S  PNT613550S  PNT613551S  PNT613552S  PNT613553S  PNT613554S  PNT613556S  PNT613557S  PNT613558S  PNT613559S  PNT613560S  PNT613561S  PNT613563S  PNT613564S  PNT613565S  PNT613566S  PNT613567S  PNT613568S  PNT613569S  PNT613570S  PNT613571S  PNT613572S  PNT613573S  PNT613574S  PNT613575S  PNT613576S  PNT613577S  PNT613578S  PNT613579S  PNT613580S  PNT613581S  PNT613582S  PNT613583S  PNT613584S  PNT613585S  PNT613586S  PNT613587S  PNT613588S  PNT613589S  PNT613590S  PNT613592S  PNT613593S  PNT613594S  PNT613595S  PNT613596S  PNT613597S  PNT613598S  PNT613599S  PNT613600S  PNT613601S  PNT613602S  PNT613603S  PNT613605S  PNT613606S  PNT613607S  PNT613608S  PNT613609S  PNT613610S  PNT613611S  PNT613612S  PNT613613S  PNT613614S  PNT613615S  PNT613616S  PNT613617S  PNT613618S  PNT613619S  PNT613620S  PNT613621S  PNT613622S  PNT613623S  PNT613624S  PNT613628S  PNT613629S  PNT613630S  PNT613631S  PNT613632S  PNT613633S  PNT613634S  PNT613635S  PNT613636S  PNT613637S  PNT613638S  PNT613639S  PNT613640S  PNT613641S  PNT613642S  PNT613643S  PNT613644S  PNT613646S  PNT613647S  PNT613648S  PNT613649S  PNT613650S  PNT613651S  PNT613652S  PNT613653S  PNT613654S  PNT613656S  PNT613657S  PNT613658S  PNT613659S  PNT613660S  PNT613661S  PNT613662S  PNT613663S  PNT613664S  PNT613665S  PNT613666S  PNT613667S  PNT613668S  PNT613669S  PNT613670S  PNT613671S  PNT613672S  PNT613674S  PNT613675S  PNT613677S  PNT613678S  PNT613679S  PNT613680S  PNT613681S  PNT613682S  PNT613683S  PNT613684S  PNT613686S  PNT613687S  PNT613688S  PNT613689S  PNT613690S  PNT613691S  PNT613692S  PNT613693S  PNT613694S  PNT613695S  PNT613696S  PNT613697S  PNT613698S  PNT613699S  PNT613700S  PNT613701S  PNT613702S  PNT613703S  PNT613704S  PNT613705S  PNT613706S  PNT613707S  PNT613708S  PNT613709S  PNT613710S  PNT613711S  PNT613712S  PNT613713S  PNT613714S  PNT613716S  PNT613717S  PNT613718S  PNT613719S  PNT613720S  PNT613721S  PNT613722S  PNT613723S  PNT613724S  PNT613726S  PNT613727S  PNT613728S  PNT613729S  PNT613731S  PNT613732S  PNT613733S  PNT613734S  PNT613735S  PNT613736S  PNT613737S  PNT613738S  PNT613739S  PNT613740S  PNT613741S  PNT613743S  PNT613744S  PNT613745S  PNT613746S  PNT613747S  PNT613748S  PNT613749S  PNT613750S  PNT613751S  PNT613752S  PNT613753S  PNT613754S  PNT613755S  PNT613756S  PNT613757S  PNT613758S  PNT613759S  PNT613760S  PNT613762S  PNT613763S  PNT613764S  PNT613765S  PNT613766S  PNT613767S  PNT613768S  PNT613769S  PNT613770S  PNT613771S  PNT613773S  PNT613775S  PNT613776S  PNT613777S  PNT613778S  PNT613779S  PNT613780S  PNT613781S  PNT613782S  PNT613783S  PNT613784S  PNT613786S  PNT613787S  PNT613788S  PNT613789S  PNT613790S  PNT613791S  PNT613792S  PNT613793S  PNT613794S  PNT613795S  PNT613796S  PNT613798S  PNT613799S  PNT613800S  PNT613801S  PNT613802S  PNT613803S  PNT613804S  PNT613805S  PNT613806S  PNT613807S  PNT613808S  PNT613809S  PNT613810S  PNT613811S  PNT613812S  PNT613813S  PNT613815S  PNT613816S  PNT613817S  PNT613818S  PNT613819S  PNT613820S  PNT613821S  PNT613822S  PNT613823S  PNT613824S  PNT613825S  PNT613826S  PNT613827S  PNT613828S  PNT613829S  PNT613830S  PNT613832S  PNT613833S  PNT613834S  PNT613835S  PNT613836S  PNT613837S  PNT613838S  PNT613839S  PNT613840S  PNT613841S  PNT613842S  PNT613843S  PNT613844S  PNT613845S  PNT613846S  PNT613847S  PNT613848S  PNT613849S  PNT613850S  PNT613851S  PNT613852S  PNT613853S  PNT613854S  PNT613855S  PNT613856S  PNT613857S  PNT613858S  PNT613859S  PNT613860S  PNT613862S  PNT613863S  PNT613864S  PNT613865S  PNT613866S  PNT613867S  PNT613868S  PNT613869S  PNT613870S  PNT613871S  PNT613872S  PNT613873S  PNT613874S  PNT613875S  PNT613876S  PNT613877S  PNT613878S  PNT613879S  PNT613880S  PNT613881S  PNT613882S  PNT613883S  PNT613884S  PNT613885S  PNT613886S  PNT613887S  PNT613888S  PNT613889S  PNT613890S  PNT613891S  PNT613892S  PNT613893S  PNT613894S  PNT613895S  PNT613896S  PNT613897S  PNT613898S  PNT613899S  PNT613900S  PNT613901S  PNT613902S  PNT613903S  PNT613904S  PNT613905S  PNT613906S  PNT613907S  PNT613908S  PNT613909S  PNT613910S  PNT613911S  PNT613912S  PNT613913S  PNT613914S  PNT613915S  PNT613916S  PNT613917S  PNT613918S  PNT613919S  PNT613921S  PNT613922S  PNT613923S  PNT613924S  PNT613925S  PNT613926S  PNT613927S  PNT613928S  PNT613929S  PNT613930S  PNT613931S  PNT613932S  PNT613933S  PNT613934S  PNT613935S  PNT613936S  PNT613937S  PNT613938S  PNT613939S  PNT613940S  PNT613941S  PNT613942S  PNT613943S  PNT613944S  PNT613945S  PNT613946S  PNT613947S  PNT613948S  PNT613949S  PNT613950S  PNT613951S  PNT613952S  PNT613953S  PNT613954S  PNT613955S  PNT613956S  PNT613957S  PNT613958S  PNT613959S  PNT613960S  PNT613961S  PNT613962S  PNT613963S  PNT613964S  PNT613965S  PNT613966S  PNT613967S  PNT613968S  PNT613969S  PNT613970S  PNT613971S  PNT613972S  PNT613973S  PNT613974S  PNT613975S  PNT613976S  PNT613977S  PNT613978S  PNT613979S  PNT613980S  PNT613981S  PNT613982S  PNT613983S  PNT613984S  PNT613985S  PNT613986S  PNT613987S  PNT613988S  PNT613989S  PNT613990S  PNT613991S  PNT613993S  PNT613994S  PNT613997S  PNT613998S  PNT613999S  PNT614000S  PNT614001S  PNT614002S  PNT614003S  PNT614004S  PNT614005S  PNT614006S  PNT614007S  PNT614008S  PNT614009S  PNT614010S  PNT614011S  PNT614012S  PNT614013S  PNT614014S  PNT614016S  PNT614017S  PNT614018S  PNT614019S  PNT614020S  PNT614021S  PNT614022S  PNT614023S  PNT614024S  PNT614026S  PNT614027S  PNT614028S  PNT614029S  PNT614030S  PNT614031S  PNT614032S  PNT614033S  PNT614035S  PNT614037S  PNT614038S  PNT614039S  PNT614041S  PNT614042S  PNT614043S  PNT614044S  PNT614045S  PNT614047S  PNT614048S  PNT614050S  PNT614052S  PNT614054S  PNT614055S  PNT614056S  PNT614057S  PNT614058S  PNT614059S  PNT614060S  PNT614061S  PNT614063S  PNT614064S  PNT614065S  PNT614066S  PNT614067S  PNT614070S  PNT614071S  PNT614072S  PNT614073S  PNT614075S  PNT614076S  PNT614077S  PNT614078S  PNT614079S  PNT614080S  PNT614081S  PNT614082S  PNT614083S  PNT614084S  PNT614085S  PNT614086S  PNT614088S  PNT614089S  PNT614090S  PNT614091S  PNT614092S  PNT614093S  PNT614094S  PNT614095S  PNT614096S  PNT614097S  PNT614098S  PNT614099S  PNT614100S  PNT614101S  PNT614102S  PNT614103S  PNT614104S  PNT614105S  PNT614106S  PNT614107S  PNT614108S  PNT614109S  PNT614110S  PNT614111S  PNT614112S  PNT614113S  PNT614114S  PNT614115S  PNT614116S  PNT614117S  PNT614118S  PNT614119S  PNT614120S  PNT614121S  PNT614122S  PNT614123S  PNT614124S  PNT614125S  PNT614126S  PNT614127S  PNT614128S  PNT614129S  PNT614130S  PNT614132S  PNT614133S  PNT614134S  PNT614135S  PNT614136S  PNT614137S  PNT614138S  PNT614139S  PNT614140S  PNT614141S  PNT614142S  PNT614143S  PNT614144S  PNT614145S  PNT614146S  PNT614147S  PNT614148S  PNT614149S  PNT614150S  PNT614151S  PNT614152S  PNT614154S  PNT614155S  PNT614156S  PNT614157S  PNT614158S  PNT614159S  PNT614160S  PNT614161S  PNT614162S  PNT614163S  PNT614164S  PNT614165S  PNT614166S  PNT614167S  PNT614168S  PNT614169S  PNT614171S  PNT614172S  PNT614173S  PNT614174S  PNT614176S  PNT614179S  PNT614181S  PNT614182S  PNT614183S  PNT614184S  PNT614185S  PNT614186S  PNT614187S  PNT614188S  PNT614191S  PNT614193S  PNT614194S  PNT614195S  PNT614196S  PNT614197S  PNT614198S  PNT614199S  PNT614200S  PNT614201S  PNT614202S  PNT614203S  PNT614204S  PNT614205S  PNT614206S  PNT614210S  PNT614212S  PNT614213S  PNT614214S  PNT614215S  PNT614216S  PNT614217S  PNT614218S  PNT614219S  PNT614220S  PNT614221S  PNT614222S  PNT614223S  PNT614224S  PNT614225S  PNT614226S  PNT614228S  PNT614229S  PNT614230S  PNT614231S  PNT614232S  PNT614233S  PNT614234S  PNT614235S  PNT614236S  PNT614237S  PNT614238S  PNT614239S  PNT614240S  PNT614241S  PNT614242S  PNT614243S  PNT614244S  PNT614245S  PNT614246S  PNT614247S  PNT614248S  PNT614249S  PNT614250S  PNT614251S  PNT614252S  PNT614253S  PNT614254S  PNT614255S  PNT614256S  PNT614257S  PNT614259S  PNT614260S  PNT614261S  PNT614262S  PNT614263S  PNT614264S  PNT614265S  PNT614266S  PNT614267S  PNT614268S  PNT614269S  PNT614270S  PNT614271S  PNT614272S  PNT614273S  PNT614274S  PNT614275S  PNT614276S  PNT614277S  PNT614278S  PNT614279S  PNT614280S  PNT614281S  PNT614282S  PNT614283S  PNT614284S  PNT614285S  PNT614286S  PNT614287S  PNT614288S  PNT614289S  PNT614290S  PNT614291S  PNT614292S  PNT614293S  PNT614294S  PNT614295S  PNT614297S  PNT614298S  PNT614299S  PNT614300S  PNT614301S  PNT614302S  PNT614303S  PNT614304S  PNT614305S  PNT614306S  PNT614307S  PNT614308S  PNT614309S  PNT614310S  PNT614311S  PNT614312S  PNT614313S  PNT614314S  PNT614315S  PNT614316S  PNT614317S  PNT614318S  PNT614319S  PNT614320S  PNT614321S  PNT614322S  PNT614323S  PNT614324S  PNT614325S  PNT614326S  PNT614327S  PNT614328S  PNT614329S  PNT614330S  PNT614331S  PNT614332S  PNT614333S  PNT614334S  PNT614335S  PNT614336S  PNT614337S  PNT614338S  PNT614339S  PNT614340S  PNT614341S  PNT614342S  PNT614343S  PNT614344S  PNT614345S  PNT614346S  PNT614347S  PNT614348S  PNT614349S  PNT614350S  PNT614351S  PNT614352S  PNT614353S  PNT614354S  PNT614355S  PNT614357S  PNT614358S  PNT614359S  PNT614360S  PNT614361S  PNT614362S  PNT614363S  PNT614364S  PNT614365S  PNT614366S  PNT614367S  PNT614368S  PNT614369S  PNT614370S  PNT614371S  PNT614372S  PNT614373S  PNT614374S  PNT614375S  PNT614376S  PNT614377S  PNT614378S  PNT614379S  PNT614380S  PNT614381S  PNT614382S  PNT614383S  PNT614384S  PNT614385S  PNT614386S  PNT614387S  PNT614388S  PNT614389S  PNT614390S  PNT614391S  PNT614393S  PNT614394S  PNT614395S  PNT614396S  PNT614397S  PNT614399S  PNT614400S  PNT614401S  PNT614402S  PNT614403S  PNT614404S  PNT614405S  PNT614406S  PNT614407S  PNT614408S  PNT614409S  PNT614410S  PNT614411S  PNT614412S  PNT614413S  PNT614414S  PNT614415S  PNT614416S  PNT614417S  PNT614418S  PNT614419S  PNT614420S  PNT614421S  PNT614422S  PNT614423S  PNT614424S  PNT614425S  PNT614426S  PNT614427S  PNT614428S  PNT614429S  PNT614430S  PNT614431S  PNT614432S  PNT614433S  PNT614434S  PNT614435S  PNT614436S  PNT614437S  PNT614438S  PNT614439S  PNT614441S  PNT614442S  PNT614443S  PNT614444S  PNT614445S  PNT614446S  PNT614447S  PNT614448S  PNT614449S  PNT614450S  PNT614451S  PNT614452S  PNT614453S  PNT614454S  PNT614455S  PNT614456S  PNT614457S  PNT614458S  PNT614459S  PNT614460S  PNT614461S  PNT614462S  PNT614463S  PNT614466S  PNT614467S  PNT614468S  PNT614469S  PNT614470S  PNT614471S  PNT614472S  PNT614473S  PNT614474S  PNT614475S  PNT614476S  PNT614477S  PNT614478S  PNT614479S  PNT614480S  PNT614481S  PNT614482S  PNT614483S  PNT614484S  PNT614485S  PNT614486S  PNT614487S  PNT614488S  PNT614489S  PNT614490S  PNT614491S  PNT614492S  PNT614494S  PNT614495S  PNT614496S  PNT614497S  PNT614498S  PNT614500S  PNT614501S  PNT614502S  PNT614503S  PNT614504S  PNT614505S  PNT614506S  PNT614507S  PNT614508S  PNT614509S  PNT614510S  PNT614511S  PNT614512S  PNT614513S  PNT614514S  PNT614515S  PNT614516S  PNT614517S  PNT614518S  PNT614519S  PNT614520S  PNT614521S  PNT614522S  PNT614524S  PNT614525S  PNT614526S  PNT614527S  PNT614528S  PNT614529S  PNT614530S  PNT614531S  PNT614532S  PNT614533S  PNT614534S  PNT614535S  PNT614536S  PNT614537S  PNT614538S  PNT614539S  PNT614540S  PNT614541S  PNT614542S  PNT614543S  PNT614544S  PNT614545S  PNT614546S  PNT614547S  PNT614548S  PNT614549S  PNT614550S  PNT614551S  PNT614552S  PNT614553S  PNT614554S  PNT614555S  PNT614556S  PNT614557S  PNT614558S  PNT614559S  PNT614560S  PNT614561S  PNT614562S  PNT614563S  PNT614565S  PNT614566S  PNT614567S  PNT614568S  PNT614569S  PNT614570S  PNT614571S  PNT614572S  PNT614573S  PNT614574S  PNT614575S  PNT614576S  PNT614577S  PNT614578S  PNT614579S  PNT614580S  PNT614581S  PNT614582S  PNT614583S  PNT614584S  PNT614585S  PNT614587S  PNT614588S  PNT614589S  PNT614590S  PNT614591S  PNT614592S  PNT614593S  PNT614594S  PNT614595S  PNT614596S  PNT614597S  PNT614598S  PNT614599S  PNT614600S  PNT614601S  PNT614602S  PNT614603S  PNT614604S  PNT614605S  PNT614606S  PNT614607S  PNT614608S  PNT614609S  PNT614610S  PNT614611S  PNT614612S  PNT614613S  PNT614614S  PNT614615S  PNT614616S  PNT614617S  PNT614618S  PNT614619S  PNT614620S  PNT614621S  PNT614622S  PNT614623S  PNT614624S  PNT614625S  PNT614626S  PNT614627S  PNT614628S  PNT614630S  PNT614631S  PNT614632S  PNT614633S  PNT614635S  PNT614636S  PNT614637S  PNT614638S  PNT614639S  PNT614640S  PNT614641S  PNT614642S  PNT614643S  PNT614644S  PNT614645S  PNT614646S  PNT614647S  PNT614648S  PNT614649S  PNT614650S  PNT614651S  PNT614652S  PNT614653S  PNT614654S  PNT614655S  PNT614656S  PNT614657S  PNT614659S  PNT614660S  PNT614661S  PNT614662S  PNT614663S  PNT614664S  PNT614665S  PNT614666S  PNT614667S  PNT614668S  PNT614669S  PNT614670S  PNT614671S  PNT614672S  PNT614673S  PNT614674S  PNT614675S  PNT614676S  PNT614677S  PNT614678S  PNT614679S  PNT614680S  PNT614681S  PNT614682S  PNT614683S  PNT614684S  PNT614685S  PNT614686S  PNT614687S  PNT614689S  PNT614690S  PNT614691S  PNT614692S  PNT614693S  PNT614694S  PNT614695S  PNT614696S  PNT614697S  PNT614698S  PNT614699S  PNT614700S  PNT614701S  PNT614702S  PNT614703S  PNT614704S  PNT614705S  PNT614706S  PNT614707S  PNT614708S  PNT614709S  PNT614710S  PNT614711S  PNT614712S  PNT614713S  PNT614714S  PNT614715S  PNT614716S  PNT614717S  PNT614718S  PNT614719S  PNT614720S  PNT614721S  PNT614723S  PNT614724S  PNT614725S  PNT614726S  PNT614727S  PNT614728S  PNT614729S  PNT614730S  PNT614731S  PNT614732S  PNT614733S  PNT614734S  PNT614735S  PNT614736S  PNT614737S  PNT614738S  PNT614739S  PNT614740S  PNT614742S  PNT614743S  PNT614744S  PNT614745S  PNT614746S  PNT614747S  PNT614748S  PNT614749S  PNT614750S  PNT614751S  PNT614752S  PNT614753S  PNT614754S  PNT614755S  PNT614756S  PNT614757S  PNT614758S  PNT614759S  PNT614760S  PNT614761S  PNT614762S  PNT614763S  PNT614764S  PNT614765S  PNT614766S  PNT614767S  PNT614768S  PNT614769S  PNT614770S  PNT614771S  PNT614772S  PNT614773S  PNT614774S  PNT614775S  PNT614776S  PNT614777S  PNT614778S  PNT614779S  PNT614780S  PNT614781S  PNT614782S  PNT614783S  PNT614784S  PNT614785S  PNT614786S  PNT614787S  PNT614788S  PNT614789S  PNT614790S  PNT614791S  PNT614792S  PNT614793S  PNT614794S  PNT614795S  PNT614796S  PNT614797S  PNT614798S  PNT614799S  PNT614800S  PNT614801S  PNT614802S  PNT614803S  PNT614804S  PNT614805S  PNT614806S  PNT614807S  PNT614808S  PNT614809S  PNT614811S  PNT614812S  PNT614813S  PNT614814S  PNT614815S  PNT614816S  PNT614817S  PNT614818S  PNT614819S  PNT614820S  PNT614821S  PNT614822S  PNT614823S  PNT614824S  PNT614825S  PNT614826S  PNT614827S  PNT614828S  PNT614830S  PNT614831S  PNT614832S  PNT614833S  PNT614834S  PNT614835S  PNT614836S  PNT614837S  PNT614838S  PNT614839S  PNT614840S  PNT614841S  PNT614842S  PNT614843S  PNT614844S  PNT614845S  PNT614846S  PNT614847S  PNT614848S  PNT614849S  PNT614851S  PNT614852S  PNT614853S  PNT614854S  PNT614855S  PNT614856S  PNT614857S  PNT614858S  PNT614859S  PNT614860S  PNT614861S  PNT614862S  PNT614863S  PNT614864S  PNT614865S  PNT614866S  PNT614867S  PNT614868S  PNT614870S  PNT614873S  PNT614875S  PNT614877S  PNT614878S  PNT614879S  PNT614880S  PNT614881S  PNT614882S  PNT614883S  PNT614884S  PNT614885S  PNT614886S  PNT614887S  PNT614888S  PNT614889S  PNT614890S  PNT614891S  PNT614892S  PNT614893S  PNT614894S  PNT614895S  PNT614896S  PNT614897S  PNT614898S  PNT614899S  PNT614900S  PNT614901S  PNT614902S  PNT614903S  PNT614904S  PNT614905S  PNT614906S  PNT614907S  PNT614908S  PNT614909S  PNT614910S  PNT614911S  PNT614912S  PNT614913S  PNT614914S  PNT614915S  PNT614916S  PNT614917S  PNT614918S  PNT614919S  PNT614920S  PNT614921S  PNT614922S  PNT614923S  PNT614924S  PNT614925S  PNT614926S  PNT614927S  PNT614928S  PNT614929S  PNT614930S  PNT614931S  PNT614932S  PNT614933S  PNT614934S  PNT614935S  PNT614936S  PNT614937S  PNT614938S  PNT614939S  PNT614940S  PNT614941S  PNT614942S  PNT614943S  PNT614944S  PNT614945S  PNT614946S  PNT614947S  PNT614948S  PNT614949S  PNT614950S  PNT614951S  PNT614952S  PNT614953S  PNT614954S  PNT614955S  PNT614957S  PNT614958S  PNT614959S  PNT614960S  PNT614961S  PNT614962S  PNT614963S  PNT614964S  PNT614965S  PNT614966S  PNT614967S  PNT614968S  PNT614969S  PNT614971S  PNT614972S  PNT614973S  PNT614974S  PNT614975S  PNT614976S  PNT614977S  PNT614978S  PNT614979S  PNT614982S  PNT614983S  PNT614984S  PNT614985S  PNT614986S  PNT614987S  PNT614988S  PNT614989S  PNT614990S  PNT614991S  PNT614992S  PNT614993S  PNT614994S  PNT614995S  PNT614996S  PNT614997S  PNT614998S  PNT614999S  PNT615000S  PNT615001S  PNT615002S  PNT615003S  PNT615005S  PNT615006S  PNT615007S  PNT615008S  PNT615009S  PNT615010S  PNT615011S  PNT615012S  PNT615014S  PNT615015S  PNT615016S  PNT615017S  PNT615018S  PNT615019S  PNT615020S  PNT615021S  PNT615022S  PNT615023S  PNT615024S  PNT615026S  PNT615027S  PNT615028S  PNT615030S  PNT615031S  PNT615032S  PNT615033S  PNT615034S  PNT615035S  PNT615036S  PNT615037S  PNT615038S  PNT615039S  PNT615040S  PNT615041S  PNT615042S  PNT615043S  PNT615044S  PNT615045S  PNT615046S  PNT615047S  PNT615048S  PNT615049S  PNT615050S  PNT615051S  PNT615052S  PNT615053S  PNT615054S  PNT615055S  PNT615056S  PNT615057S  PNT615058S  PNT615059S  PNT615060S  PNT615061S  PNT615062S  PNT615063S  PNT615064S  PNT615065S  PNT615066S  PNT615067S  PNT615068S  PNT615069S  PNT615070S  PNT615071S  PNT615072S  PNT615073S  PNT615074S  PNT615075S  PNT615076S  PNT615077S  PNT615078S  PNT615079S  PNT615080S  PNT615081S  PNT615082S  PNT615083S  PNT615084S  PNT615085S  PNT615086S  PNT615087S  PNT615088S  PNT615089S  PNT615090S  PNT615091S  PNT615092S  PNT615093S  PNT615094S  PNT615095S  PNT615096S  PNT615097S  PNT615098S  PNT615099S  PNT615100S  PNT615101S  PNT615103S  PNT615104S  PNT615105S  PNT615106S  PNT615107S  PNT615108S  PNT615109S  PNT615110S  PNT615111S  PNT615112S  PNT615113S  PNT615114S  PNT615115S  PNT615116S  PNT615117S  PNT615118S  PNT615119S  PNT615120S  PNT615121S  PNT615122S  PNT615123S  PNT615124S  PNT615125S  PNT615126S  PNT615127S  PNT615128S  PNT615129S  PNT615130S  PNT615131S  PNT615132S  PNT615133S  PNT615134S  PNT615135S  PNT615136S  PNT615137S  PNT615138S  PNT615139S  PNT615140S  PNT615141S  PNT615142S  PNT615143S  PNT615145S  PNT615146S  PNT615147S  PNT615148S  PNT615150S  PNT615151S  PNT615152S  PNT615153S  PNT615154S  PNT615156S  PNT615157S  PNT615158S  PNT615159S  PNT615160S  PNT615161S  PNT615162S  PNT615163S  PNT615164S  PNT615165S  PNT615166S  PNT615167S  PNT615168S  PNT615169S  PNT615170S  PNT615171S  PNT615172S  PNT615173S  PNT615174S  PNT615175S  PNT615176S  PNT615177S  PNT615178S  PNT615179S  PNT615180S  PNT615181S  PNT615182S  PNT615183S  PNT615184S  PNT615185S  PNT615186S  PNT615187S  PNT615188S  PNT615189S  PNT615190S  PNT615191S  PNT615192S  PNT615193S  PNT615194S  PNT615195S  PNT615196S  PNT615197S  PNT615198S  PNT615199S  PNT615200S  PNT615201S  PNT615202S  PNT615203S  PNT615204S  PNT615205S  PNT615206S  PNT615207S  PNT615208S  PNT615209S  PNT615210S  PNT615211S  PNT615212S  PNT615213S  PNT615214S  PNT615215S  PNT615216S  PNT615217S  PNT615218S  PNT615219S  PNT615220S  PNT615221S  PNT615222S  PNT615224S  PNT615225S  PNT615226S  PNT615227S  PNT615228S  PNT615229S  PNT615230S  PNT615231S  PNT615232S  PNT615233S  PNT615234S  PNT615235S  PNT615236S  PNT615237S  PNT615238S  PNT615239S  PNT615240S  PNT615241S  PNT615242S  PNT615243S  PNT615244S  PNT615245S  PNT615246S  PNT615247S  PNT615248S  PNT615249S  PNT615250S  PNT615251S  PNT615252S  PNT615253S  PNT615254S  PNT615255S  PNT615256S  PNT615257S  PNT615258S  PNT615259S  PNT615261S  PNT615262S  PNT615263S  PNT615264S  PNT615265S  PNT615266S  PNT615267S  PNT615268S  PNT615270S  PNT615271S  PNT615272S  PNT615274S  PNT615275S  PNT615276S  PNT615277S  PNT615278S  PNT615279S  PNT615280S  PNT615281S  PNT615282S  PNT615283S  PNT615284S  PNT615285S  PNT615286S  PNT615287S  PNT615288S  PNT615289S  PNT615290S  PNT615291S  PNT615293S  PNT615294S  PNT615295S  PNT615296S  PNT615297S  PNT615298S  PNT615299S  PNT615300S  PNT615301S  PNT615302S  PNT615304S  PNT615305S  PNT615306S  PNT615307S  PNT615308S  PNT615309S  PNT615310S  PNT615311S  PNT615312S  PNT615314S  PNT615316S  PNT615317S  PNT615318S  PNT615319S  PNT615320S  PNT615321S  PNT615322S  PNT615323S  PNT615324S  PNT615325S  PNT615326S  PNT615327S  PNT615328S  PNT615329S  PNT615330S  PNT615331S  PNT615332S  PNT615333S  PNT615334S  PNT615335S  PNT615336S  PNT615337S  PNT615338S  PNT615339S  PNT615340S  PNT615341S  PNT615342S  PNT615343S  PNT615344S  PNT615345S  PNT615346S  PNT615347S  PNT615348S  PNT615349S  PNT615350S  PNT615351S  PNT615352S  PNT615353S  PNT615354S  PNT615355S  PNT615356S  PNT615357S  PNT615358S  PNT615359S  PNT615360S  PNT615361S  PNT615362S  PNT615363S  PNT615364S  PNT615365S  PNT615366S  PNT615367S  PNT615368S  PNT615369S  PNT615370S  PNT615371S  PNT615372S  PNT615373S  PNT615374S  PNT615375S  PNT615376S  PNT615377S  PNT615378S  PNT615379S  PNT615380S  PNT615381S  PNT615382S  PNT615383S  PNT615384S  PNT615385S  PNT615386S  PNT615387S  PNT615388S  PNT615389S  PNT615390S  PNT615391S  PNT615392S  PNT615393S  PNT615394S  PNT615395S  PNT615396S  PNT615397S  PNT615398S  PNT615399S  PNT615400S  PNT615402S  PNT615403S  PNT615404S  PNT615405S  PNT615407S  PNT615408S  PNT615409S  PNT615410S  PNT615411S  PNT615412S  PNT615413S  PNT615414S  PNT615416S  PNT615417S  PNT615419S  PNT615420S  PNT615421S  PNT615423S  PNT615425S  PNT615426S  PNT615427S  PNT615428S  PNT615429S  PNT615430S  PNT615431S  PNT615432S  PNT615433S  PNT615434S  PNT615435S  PNT615436S  PNT615438S  PNT615439S  PNT615441S  PNT615442S  PNT615443S  PNT615444S  PNT615445S  PNT615446S  PNT615447S  PNT615448S  PNT615449S  PNT615450S  PNT615451S  PNT615452S  PNT615453S  PNT615454S  PNT615455S  PNT615456S  PNT615457S  PNT615458S  PNT615459S  PNT615460S  PNT615461S  PNT615462S  PNT615463S  PNT615464S  PNT615465S  PNT615466S  PNT615467S  PNT615468S  PNT615469S  PNT615470S  PNT615471S  PNT615473S  PNT615474S  PNT615475S  PNT615476S  PNT615477S  PNT615478S  PNT615479S  PNT615480S  PNT615481S  PNT615482S  PNT615483S  PNT615484S  PNT615485S  PNT615486S  PNT615487S  PNT615488S  PNT615489S  PNT615490S  PNT615491S  PNT615492S  PNT615493S  PNT615494S  PNT615495S  PNT615496S  PNT615497S  PNT615498S  PNT615499S  PNT615500S  PNT615501S  PNT615502S  PNT615504S  PNT615505S  PNT615506S  PNT615507S  PNT615508S  PNT615509S  PNT615510S  PNT615511S  PNT615512S  PNT615513S  PNT615514S  PNT615520S  PNT615521S  PNT615522S  PNT615523S  PNT615525S  PNT615526S  PNT615527S  PNT615528S  PNT615529S  PNT615531S  PNT615532S  PNT615533S  PNT615534S  PNT615535S  PNT615536S  PNT615537S  PNT615538S  PNT615539S  PNT615540S  PNT615541S  PNT615542S  PNT615543S  PNT615545S  PNT615546S  PNT615547S  PNT615548S  PNT615549S  PNT615550S  PNT615551S  PNT615552S  PNT615553S  PNT615554S  PNT615555S  PNT615556S  PNT615557S  PNT615558S  PNT615559S  PNT615560S  PNT615561S  PNT615562S  PNT615563S  PNT615629S  PNT615630S    Product UPN/GTIN:  00885074300252  Serial Numbers:  PNT601839S",
      "recall_initiation_date": "20180621",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}