  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "A change in the concentration of N-acetylcysteine (NAC) that may cause interference for the ADVIA¿ Chemistry Systems Acetaminophen assay for certain lots.",
      "address_2": "PO BOX 6101",
      "product_quantity": "1032",
      "code_info": "Catalog Number 07989138, Siemens Material Number (SMN) 10327381, Lot Numbers: 46251, 46564, 46369, 46370, 46619, 46782, 47001, 47157, 47558; with expiration dates June 30, 2015 through April 30, 2016.",
      "center_classification_date": "20150819",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and to the countries of : Canada, Netherlands, Great Britain, Norway, Sweden, Spain, Italy, Germany, France, Finland, Belgium, Portugal, Czech Republic, Israel, Poland, Turkey and Bahrain.",
      "state": "DE",
      "product_description": "ADVIA Chemistry Acetaminophen reagents; for in vitro diagnostic use in the quantitative determination of acetaminophen in human serum and plasma (lithium heparin) on ADVIA Chemistry systems.",
      "report_date": "20150826",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-2416-2015",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "71710",
      "termination_date": "20160901",
      "more_code_info": "",
      "recall_initiation_date": "20150630",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}