  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Alameda",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92708",
      "recalling_firm": "Abbott Diabetes Care, Inc.",
      "address_1": "1360 S Loop Rd",
      "address_2": "",
      "postal_code": "94502-7000",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CA, FL, MI, NY, OH, TX, UT, VA.",
      "recall_number": "Z-2414-2023",
      "product_description": "FreeStyle Libre 2 Readers, REF: 71951-01 and 71953-01, used with FreeStyle Libre 2 System",
      "product_quantity": "1,722,645",
      "reason_for_recall": "If blood glucose monitoring system users attempt to start a new sensor, when the old sensor wasn't used for the full 14-day wear time, then the reader may display \"Incompatible Sensor\" message. If users don't have the previous sensor, or FreeStyle Libre 2 app, or the sensor's full wear time hasn't ended, users may be unable to start a new sensor, which could result in no or delayed glucose results",
      "recall_initiation_date": "20230712",
      "center_classification_date": "20230817",
      "report_date": "20230823",
      "code_info": "REF/UDI-DI/Lots: 71951-01/357599805005/02A289D, 03B082D, 02B082D, 02B109D, 03B251D, 04B251D, 01C073D, 01C084D, 02C087D, 01C143D, 03C152D, 03C166D, 01C174D, 03C222D, 01C256D, 02C313D, 01D018D, 02D093D, 02D129D;  71953-01/357599803001/01A198D, 02A198D, 03A198D, 04A198D, 05A198D, 01A202D, 04A202D, 01A204D, 02A203D, 02A205D, 03A205D, 01A211D, 02A211D, 03A210D, 02A212D, 04A218D, 05A218D, 06A218D, 01A219D, 04A225D, 05A225D, 01A231D, 03A231D, 02A232D, 06A232D, 02A240D, 02A248D, 01A259D, 01A260D, 04A262D, 03A266D, 01A267D, 02A268D, 03A268D, 02A280D, 01A281D, 01A283D, 02A283D, 01A290D, 02A297D, 02A298D, 03A296D, 02A303D, 02A304D, 04A325D, 01A328D, 03A325D, 02A340D, 03A344D, 02A351D, 03A350D, 01A352D, 03A352D, 02A358D, 01A364D, 02A363D, 01B007F, 02B007F, 01B008F, 01B011F, 03B012F, 01B012F, 01B014F, 05B013F, 01B015F, 04B013F, 02B013F, 11B021F, 02B034F, 02B035F, 04B034F, 04B035F, 06B034F, 07B034F, 02B036F, 05B035F, 01B036F, 01B039F, 03B039F, 03B036F, 03B041F, 05B041F, 02B042F, 03B043F, 02B046F, 05B043F, 06B046F, 09B046F, 01B071F, 05B071F, 06B074F, 04B074F, 02B076F, 05B075F, 01B077F, 05B076F, 02B078F, 05B078F, 02B081F, 02B082F, 04B082F, 02B083D, 05B082D, 03B109D, 02B203D, 03B203D, 01B205D, 01B207D, 01B208D, 02B208D, 02B211D, 02B216D, 02B229D, 01B288D, 02B289D, 01B289D, 02B305D, 01B306F, 02B306F, 01B306D, 02B308F, 03B306F, 01B308F, 01B309F, 01B309D, 01B310D, 02B312D, 01B313D, 03B312D, 03B313D, 05B322F, 07B322F, 01B323F, 01B326F, 06B322F, 01B305D, 01B327F, 03B326F, 04B326F, 02B328F, 01B308D, 01B333D, 01B334D, 01B335F, 03B335F, 02B333D, 01B336F, 02B336F, 01B337F, 02B337F, 01B302F, 01B340F, 02B340F, 03B302F, 04B301F, 01B337D, 01B341F, 02B341F, 01B341D, 02B341D, 01B342F, 02B342F, 02B343F, 01B343F, 04B302F, 02B344F, 06B349F, 03B351D, 04B352D, 04B354D, 02B356D, 01B356D, 04B356D, 05B356D, 01B362D, 02B362D, 02B363D, 01C006D, 03C005D, 02C005D, 02C010D, 01C011D, 01C012D, 03C013D, 02C013D, 02C025F, 03C028F, 01C028F, 02C031F, 03C031F, 01C034F, 02C032F, 01C074D, 03C074D, 02C073D, 02C074D, 03C075D, 01C075D, 02C080D, 02C112D, 03C117D, 03C123D, 03C126D, 02C130D, 03C129D, 02C133D, 03C133D, 01C133D, 01C136D, 02C136D, 01C137D, 01C138D, 04C138D, 01C140D, 02C140D, 03C140D, 02C143D, 01C144D, 02C145D, 05C146D, 01C146D, 01C159D, 03C159D, 01C161D, 03C165D, 05C165D, 02C172D, 01C181D, 02C181D, 01C182D, 01C186D, 01C188D, 02C188D, 01C194D, 02C193D, 03C196D, 03C199D, 05C199D, 01C201D, 02C199D, 03C207D, 01C207D, 01C209D, 02C208D, 03C213D, 02C223D, 01C223D, 01C224D, 02C224D, 01C227D, 01C228D, 02C229D, 03C224D, 02C236D, 01C237D, 03C236D, 02C237D, 03C238D, 04C241D, 01C242D, 01C243D, 03C242D, 02C242D, 02C243D, 04C243D, 01C245D, 02C245D, 01C249D, 02C250D, 03C251D, 02C255D, 03C256D, 01C257D, 02C256D, 02C257D, 03C257D, 02C262D, 02C265D, 01C266D, 03C270D, 08C271D, 01C272D, 01C276D, 01C273D, 04C276D, 02C278D, 03C279D, 01C283D, 02C287D, 04C292D, 02C294D, 03C294D, 06C292D, 01C298D, 01C299D, 02C298D, 01C305D, 02C305D, 03C305D, 01C307D, 01C308D, 01C313D, 02C312D, 01C317D, 01C318D, 03C320D, 02C321D, 01C322D, 01C325D, 02C325D, 01C326D, 01C332D, 01C327D, 01C334D, 02C334D, 04C334D, 01C340D, 02C340D, 01C341D, 03C340D, 01C346D, 02C355D, 01D010D, 01D011D, 02D011D, 01D012D, 02D018D, 02D017D, 02D023D, 02D028D, 02D019D, 03D024D, 01D025D, 02D025D, 01D027D, 03D027D, 01D033D, 03D034D, 04D034D, 05D033D, 01D040D, 02D040D, 02D038D, 06D055D, 03D041D, 02D045D, 02D053D, 03D058D, 02D056D, 02D060D, 03D067D, 04D068D, 03D065D, 02D065D, 02D066D, 02D069D, 01D072D, 03D069D, 02D072D, 01D076D, 02D075D, 03D075D, 01D095D, 03D083D, 01D081D, 02D095D, 02D081D, 02D082D, 03D080D, 02D087D, 06D103D, 01D102D, 02D107D, 01D107D, 01D104D, 03D107D, 02D104D, 01D111D, 01D115D, 02D115D, 02D116D, 03D115D, 01D128D, 01D129D, 05D130D, 04D125D, 02D128D, 03D136D, 02D136D, 03D142D, 01D139D, 03D144D, 02D125D, 01D150D, 03D151D, 02D151D, 03D152D, 01D156D, 02D157D, 02D160D, 02D156D, 01D159D, 01D177D, 02D177D",
      "more_code_info": ""
    }
  ]
}