      "country": "United States",
      "city": "Mounds View",
      "address_1": "8200 Coral Sea St Ne",
      "reason_for_recall": "Potential for a device reset to occur in Percepta CRT-P MRI SureScan and Percepta Quad CRT-P MRI SureScan due to a timing interaction between the EffectivCRT Diagnostic and the Ventricular Safety Pacing feature (VSP).",
      "address_2": "",
      "product_quantity": "1842 units",
      "code_info": "GTIN/UPN:   (a) 00643169735637  Serial Numbers:   RNO201997H  RNO202010H  RNO202012H  RNO202015H  RNO202016H  RNO202018H  RNO202045H  RNO202049H  RNO202050H  RNO202051H  RNO202055H  RNO202056H  RNO202058H  RNO202062H  RNO202065H  RNO202067H  RNO202068H  RNO202069H  RNO202072H  RNO202075H  RNO202078H  RNO202083H  RNO202086H  RNO202087H  RNO202088H  RNO202092H  RNO202093H  RNO202094H  RNO202098H  RNO202099H  RNO202100H  RNO202101H  RNO201399H  RNO201985H  RNO201986H  RNO201989H  RNO201990H  RNO201994H  RNO201988H  RNO201995H  RNO202048H  RNO202090H  RNO202091H  RNO202011H  RNO201962H  RNO201966H  RNO201967H  RNO201970H  RNO201974H  RNO201977H  RNO201978H  RNO201980H  RNO201982H  RNO202013H  RNO202014H  RNO202017H  RNO202019H  RNO202037H  RNO202040H  RNO202042H  RNO202073H  RNO202079H  RNO202009H  RNO201922H  RNO201981H  RNO202034H  RNO200663H  RNO201582H  RNO201603H  RNO200042H  RNO200092H  RNO200202H  RNO200455H  RNO200550H  RNO200583H  RNO201033H  RNO201133H  RNO201837H  RNO201923H  RNO201925H  RNO201928H  RNO201930H  RNO201934H  RNO201936H  RNO201938H  RNO201939H  RNO201940H  RNO201944H  RNO201945H  RNO201947H  RNO201948H  RNO201950H  RNO201952H  RNO201953H  RNO201956H  RNO201960H  RNO201968H  RNO201969H  RNO201971H  RNO201973H  RNO200252H  RNO200269H  RNO201907H  RNO201912H  RNO201913H  RNO201918H  RNO201920H  RNO201924H  RNO201935H  RNO201946H  RNO201904H  RNO200829H  RNO200809H  RNO200823H  RNO200087H  RNO201066H  RNO201821H  RNO201358H  RNO200859H  RNO201768H  RNO201082H  RNO200571H  RNO200547H  RNO201286H  RNO200520H  RNO201181H  RNO201431H  RNO200544H  RNO201490H  RNO200860H  RNO201019H  RNO200977H  RNO200414H  RNO200864H  RNO200499H  RNO201424H  RNO201574H  RNO201813H  RNO201308H  RNO201751H  RNO201333H  RNO201794H  RNO201487H  RNO200097H  RNO201743H  RNO200480H  RNO201411H  RNO201330H  RNO200399H  RNO200759H  RNO201804H  RNO201707H  RNO201460H  RNO200738H  RNO201463H  RNO201024H  RNO200528H  RNO200281H  RNO201770H  RNO200412H  RNO201070H  RNO201724H  RNO201937H  RNO201831H  RNO200256H  RNO201818H  RNO200526H  RNO201758H  RNO201858H  RNO201864H  RNO201641H  RNO200772H  RNO201558H  RNO201884H  RNO200153H  RNO200646H  RNO200845H  RNO200888H  RNO201842H  RNO201887H  RNO201888H  RNO201889H  RNO200900H  RNO201843H  RNO201413H  RNO200881H  RNO200872H  RNO200868H  RNO201306H  RNO200392H  RNO201714H  RNO200467H  RNO201860H  RNO201929H  RNO201412H  RNO201801H  RNO201047H  RNO201338H  RNO201827H  RNO200108H  RNO200929H  RNO200931H  RNO201869H  RNO201760H  RNO201931H  RNO200390H  RNO201339H  RNO200828H  RNO201871H  RNO201094H  RNO201248H  RNO201793H  RNO200157H  RNO201996H  RNO201144H  RNO201328H  RNO200732H  RNO200698H  RNO201449H  RNO200170H  RNO201715H  RNO200953H  RNO200810H  RNO200733H  RNO201681H  RNO201701H  RNO201029H  RNO200180H  RNO200992H  RNO200551H  RNO201824H  RNO201757H  RNO201789H  RNO201742H  RNO201639H  RNO200374H  RNO200063H  RNO201478H  RNO201479H  RNO201217H  RNO201709H  RNO200815H  RNO201205H  RNO200442H  RNO200413H  RNO200093H  RNO200608H  RNO201346H  RNO201347H  RNO200835H  RNO201542H  RNO201999H  RNO200493H  RNO202002H  RNO201771H  RNO201514H  RNO201396H  RNO201426H  RNO200968H  RNO200973H  RNO201799H  RNO201026H  RNO200878H  RNO201876H  RNO201755H  RNO201214H  RNO201404H  RNO201387H  RNO200907H  RNO201236H  RNO201421H  RNO201430H  RNO200147H  RNO201456H  RNO201414H  RNO201521H  RNO201713H  RNO201722H  RNO201529H  RNO201592H  RNO200875H  RNO200990H  RNO201337H  RNO200273H  RNO200498H  RNO200970H  RNO201457H  RNO200381H  RNO200076H  RNO200485H  RNO201113H  RNO201020H  RNO201432H  RNO201545H  RNO200171H  RNO200957H  RNO201108H  RNO200963H  RNO200912H  RNO201765H  RNO202003H  RNO200386H  RNO201872H  RNO200744H  RNO201445H  RNO200040H  RNO201749H  RNO200588H  RNO201147H  RNO201671H  RNO201740H  RNO201519H  RNO201001H  RNO201422H  RNO200251H  RNO200843H  RNO201097H  RNO200344H  RNO201393H  RNO201077H  RNO201383H  RNO200837H  RNO201591H  RNO201000H  RNO201512H  RNO201752H  RNO201774H  RNO200006H  RNO201516H  RNO200549H  RNO200577H  RNO201076H  RNO201748H  RNO200574H  RNO200911H  RNO201640H  RNO201304H  RNO201853H  RNO200360H  RNO200848H  RNO201407H  RNO200517H  RNO200729H  RNO201064H  RNO201828H  RNO200213H  RNO201857H  RNO201866H  RNO201437H  RNO200206H  RNO200756H  RNO200816H  RNO201911H  RNO201915H  RNO201579H  RNO201737H  RNO200523H  RNO201777H  RNO201791H  RNO201820H  RNO201425H  RNO201921H  RNO201806H  RNO201061H  RNO201206H  RNO201229H  RNO201398H  RNO201808H  RNO200032H  RNO201427H  RNO201299H  RNO200039H  RNO201767H  RNO201509H  RNO201829H  RNO200890H  RNO201902H  RNO201441H  RNO201570H  RNO200260H  RNO200734H  RNO202008H  RNO200511H  RNO200821H  RNO201812H  RNO201350H  RNO200770H  RNO200494H  RNO200933H  RNO201372H  RNO201632H  RNO201377H  RNO201381H  RNO201833H  RNO201762H  RNO200754H  RNO200811H  RNO200905H  RNO201811H  RNO201505H  RNO201683H  RNO201687H  RNO201475H  RNO201665H  RNO201539H  RNO201226H  RNO201493H  RNO201482H  RNO200760H  RNO201677H  RNO200460H  RNO201018H  RNO200617H  RNO201137H  RNO200914H  RNO201746H  RNO201738H  RNO201795H  RNO200258H  RNO200259H  RNO200813H  RNO200448H  RNO201543H  RNO200605H  RNO200389H  RNO201785H  RNO201015H  RNO200942H  RNO200236H  RNO200819H  RNO201518H  RNO201819H  RNO200144H  RNO200403H  RNO200438H  RNO201208H  RNO200836H  RNO200779H  RNO200138H  RNO200013H  RNO201848H  RNO200993H  RNO201343H  RNO200156H  RNO200137H  RNO201559H  RNO201783H  RNO201069H  RNO200554H  RNO200473H  RNO201361H  RNO201336H  RNO201814H  RNO201323H  RNO200436H  RNO201816H  RNO201428H  RNO200176H  RNO200826H  RNO200578H  RNO201916H  RNO200098H  RNO201468H  RNO201769H  RNO201173H  RNO201186H  RNO201508H  RNO201014H  RNO201220H  RNO201098H  RNO201910H  RNO201792H  RNO201486H  RNO201176H  RNO201177H  RNO201809H  RNO200576H  RNO201838H  RNO201442H  RNO201091H  RNO200958H  RNO200926H  RNO200930H  RNO200355H  RNO200774H  RNO200806H  RNO201841H  RNO201865H  RNO201868H  RNO200934H  RNO200221H  RNO201886H  RNO200850H  RNO201786H  RNO201227H  RNO200272H  RNO201423H  RNO201071H  RNO200409H  RNO201247H  RNO201433H  RNO201734H  RNO201736H  RNO200192H  RNO201489H  RNO201676H  RNO201451H  RNO201101H  RNO201932H  RNO200871H  RNO201817H  RNO200364H  RNO201822H  RNO201237H  RNO200457H  RNO200264H  RNO200266H  RNO201784H  RNO200454H  RNO201778H  RNO201139H  RNO201859H  RNO200406H  RNO201025H  RNO200769H  RNO200208H  RNO201028H  RNO201520H  RNO201800H  RNO200211H  RNO200212H  RNO201476H  RNO201788H  RNO201852H  RNO201662H  RNO200988H  RNO201184H  RNO201341H  RNO201418H  RNO201782H  RNO201847H  RNO200187H  RNO200445H  RNO201011H  RNO201439H  RNO201334H  RNO201727H  RNO201316H  RNO200852H  RNO201088H  RNO200101H  RNO201894H  RNO201708H  RNO201530H  RNO201717H  RNO200915H  RNO200794H  RNO201606H  RNO201607H  RNO201496H  RNO201712H  RNO201209H  RNO201207H  RNO200398H  RNO200595H  RNO200664H  RNO201510H  RNO200052H  RNO200582H  RNO201896H  RNO200880H  RNO201002H  RNO201302H  RNO201587H  RNO200902H  RNO201327H  RNO200949H  RNO201189H  RNO200822H  RNO200941H  RNO201547H  RNO201434H  RNO200834H  RNO201875H  RNO201877H  RNO201917H  RNO201927H  RNO201761H  RNO201780H  RNO201332H  RNO201797H  RNO201826H  RNO200225H  RNO200228H  RNO200808H  RNO200100H  RNO200105H  RNO200488H  RNO200644H  RNO200803H  RNO201310H  RNO200271H  RNO200950H  RNO200766H  RNO200767H  RNO200496H  RNO201680H  RNO201105H  RNO201107H  RNO201845H  RNO200967H  RNO201440H  RNO200253H  RNO200416H  RNO200111H  RNO200112H  RNO200125H  RNO200661H  RNO200703H  RNO200705H  RNO200706H  RNO200707H  RNO200708H  RNO200920H  RNO201313H  RNO201526H  RNO201038H  RNO200158H  RNO200618H  RNO200690H  RNO200893H  RNO200894H  RNO200186H  RNO200796H  RNO200043H  RNO200046H  RNO200047H  RNO200067H  RNO200072H  RNO200073H  RNO200127H  RNO200210H  RNO200295H  RNO200296H  RNO200297H  RNO200298H  RNO200299H  RNO200671H  RNO200672H  RNO200673H  RNO200674H  RNO200675H  RNO200676H  RNO200679H  RNO200681H  RNO200682H  RNO200683H  RNO201086H  RNO201089H  RNO201090H  RNO201092H  RNO201095H  RNO201362H  RNO201363H  RNO201367H  RNO201368H  RNO201369H  RNO201854H  RNO201855H  RNO201483H  RNO200804H  RNO200012H  RNO200265H  RNO200276H  RNO200387H  RNO201004H  RNO201005H  RNO201834H  RNO200580H  RNO200775H  RNO200118H  RNO200921H  RNO200109H  RNO200059H  RNO200222H  RNO200223H  RNO200317H  RNO201148H  RNO200686H  RNO201035H  RNO200066H  RNO200183H  RNO200203H  RNO200363H  RNO200447H  RNO200565H  RNO200768H  RNO200781H  RNO200895H  RNO201151H  RNO201636H  RNO200055H  RNO200099H  RNO200145H  RNO200172H  RNO200188H  RNO200333H  RNO200334H  RNO200615H  RNO200687H  RNO200688H  RNO200938H  RNO201156H  RNO201157H  RNO200366H  RNO200230H  RNO201008H  RNO200037H  RNO200546H  RNO200736H  RNO200653H  RNO201466H  RNO201192H  RNO200050H  RNO200051H  RNO200502H  RNO201690H  RNO201691H  RNO201409H  RNO201533H  RNO200887H  RNO200952H  RNO200961H  RNO201429H  RNO201498H  RNO201535H  RNO200162H  RNO201739H  RNO200591H  RNO201693H  RNO201694H  RNO200030H  RNO200031H  RNO200078H  RNO200154H  RNO200199H  RNO200200H  RNO200824H  RNO200913H  RNO201340H  RNO200478H  RNO200696H  RNO200697H  RNO201116H  RNO201117H  RNO201118H  RNO201628H  RNO200980H  RNO201867H  RNO201219H  RNO200132H  RNO200233H  RNO200731H  RNO201474H  RNO200817H  RNO200610H  RNO201684H  RNO200319H  RNO200320H  RNO200321H  RNO200322H  RNO200323H  RNO200324H  RNO200325H  RNO200326H  RNO200327H  RNO200330H  RNO200593H  RNO200263H  RNO200219H  RNO200560H  RNO200584H  RNO200780H  RNO201150H  RNO201153H  RNO201104H  RNO201130H  RNO201846H  RNO200463H  RNO201168H  RNO201187H  RNO201380H  RNO200422H  RNO200897H  RNO201221H  RNO201400H  RNO201504H  RNO201507H  RNO200238H  RNO200023H  RNO200566H  RNO200500H  RNO200343H  RNO200807H  RNO200979H  RNO200982H  RNO201348H  RNO201459H  RNO201541H  RNO201624H  RNO201661H  RNO201692H  RNO201873H  RNO201174H  RNO200529H  RNO200235H  RNO200487H  RNO200937H  RNO200976H  RNO201134H  RNO200025H  RNO200764H  RNO200475H  RNO201093H  RNO201138H  RNO200345H  RNO201317H  RNO200167H  RNO200383H  RNO200981H  RNO200609H  RNO201311H  RNO200778H  RNO200479H  RNO200484H  RNO200846H  RNO200166H  RNO200357H  RNO201178H  RNO201919H  RNO201705H  RNO201706H  RNO200243H  RNO200869H  RNO200870H  RNO201480H  RNO201488H  RNO200102H  RNO200339H  RNO200316H  RNO201218H  RNO200342H  RNO201075H  RNO201132H  RNO200028H  RNO201435H  RNO200245H  RNO201021H  RNO200370H  RNO200472H  RNO200477H  RNO200628H  RNO200694H  RNO200797H  RNO200936H  RNO200987H  RNO201357H  RNO201458H  RNO201678H  RNO200545H  RNO200978H  RNO201365H  RNO201653H  RNO200151H  RNO200899H  RNO201006H  RNO201499H  RNO201500H  RNO201501H  RNO201122H  RNO200140H  RNO201136H  RNO200254H  RNO200086H  RNO200088H  RNO200205H  RNO200405H  RNO201123H  RNO201324H  RNO200209H  RNO201764H  RNO201815H  RNO200074H  RNO200419H  RNO201844H  RNO200812H  RNO201180H  RNO201390H  RNO200277H  RNO200375H  RNO200737H  RNO200867H  RNO201410H  RNO200261H  RNO200262H  RNO200417H  RNO200430H  RNO200278H  RNO200516H  RNO200962H  RNO200331H  RNO200335H  RNO200336H  RNO200026H  RNO200373H  RNO200429H  RNO200597H  RNO201763H  RNO200805H  RNO200468H  RNO200562H  RNO200564H  RNO201068H  RNO201074H  RNO201190H  RNO201228H  RNO201231H  RNO201605H  RNO201608H  RNO201202H  RNO200691H  RNO200693H  RNO201359H  RNO200995H  RNO200064H  RNO200662H  RNO200146H  RNO200678H  RNO200865H  RNO201416H  RNO200348H  RNO200352H  RNO200194H  RNO200312H  RNO200313H  RNO201374H  RNO201375H  RNO200841H  RNO200077H  RNO200082H  RNO200084H  RNO200561H  RNO200741H  RNO200742H  RNO200743H  RNO200179H  RNO200397H  RNO200415H  RNO200862H  RNO200886H  RNO201354H  RNO201455H  RNO200274H  RNO200275H  RNO200665H  RNO200667H  RNO200668H  RNO200669H  RNO200670H  RNO201042H  RNO201444H  RNO200314H  RNO200315H  RNO201540H  RNO201729H  RNO201172H  RNO201835H  RNO200548H  RNO200075H  RNO200858H  RNO201366H  RNO201370H  RNO201371H  RNO201839H  RNO200218H  RNO200898H  RNO200903H  RNO200904H  RNO201642H  RNO200054H  RNO201169H  RNO201531H  RNO201376H  RNO200404H  RNO201182H  RNO200818H  RNO201629H  RNO200080H  RNO201166H  RNO200466H  RNO200533H  RNO200873H  RNO200453H  RNO201471H  RNO200725H  RNO200727H  RNO200927H  RNO200169H  RNO201517H  RNO200525H  RNO200621H  RNO200255H  RNO200481H  RNO201331H  RNO200378H  RNO200505H  RNO201175H  RNO200038H  RNO200428H  RNO200495H  RNO200585H  RNO200740H  RNO200996H  RNO201119H  RNO201513H  RNO200426H  RNO200989H  RNO200514H  RNO201300H  RNO200279H  RNO200234H  RNO200410H  RNO200198H  RNO200270H  RNO200371H  RNO200538H  RNO200637H  RNO200882H  RNO200999H  RNO201043H  RNO201054H  RNO201057H  RNO201232H  RNO201309H  RNO201723H  RNO201725H  RNO200883H  RNO200972H  RNO201110H  RNO201307H  RNO200214H  RNO201392H  RNO200061H  RNO200503H  RNO200048H  RNO201225H  RNO201453H  RNO200569H  RNO200106H  RNO200393H  RNO201408H  RNO200318H  RNO200539H  RNO201224H  RNO200116H  RNO200611H  RNO200612H  RNO200955H  RNO200966H  RNO200969H  RNO201145H  RNO200411H  RNO200142H  RNO200820H  RNO200939H  RNO200945H  RNO201072H  RNO201238H  RNO200614H  RNO201899H  RNO200010H  RNO200149H  RNO200191H  RNO200855H  RNO200379H  RNO200586H  RNO200647H  RNO200795H  RNO201720H  RNO201721H  RNO200110H  RNO200216H  RNO200518H  RNO200555H  RNO200619H  RNO200626H  RNO201041H  RNO201160H  RNO201197H  RNO201446H  RNO201502H  RNO200800H  RNO200964H  RNO201546H  RNO200309H  RNO200616H  RNO200103H  RNO200510H  RNO200739H  RNO200943H  RNO200782H  RNO200029H  RNO200346H  RNO200771H  RNO201469H  RNO200268H  RNO200354H  RNO200367H  RNO200368H  RNO200777H  RNO200786H  RNO201325H  RNO201406H  RNO201549H  RNO201550H  RNO201551H  RNO201552H  RNO201553H  RNO201554H  RNO201555H  RNO200959H  RNO201044H  RNO200563H  RNO201779H  RNO200115H  RNO200531H  RNO201121H  RNO201481H  RNO201631H  RNO201790H  RNO200490H  RNO200603H  RNO200986H  RNO200177H  RNO200536H  RNO200598H  RNO200851H  RNO200431H  RNO201385H  RNO201386H  RNO201527H  RNO201528H  RNO201537H  RNO201544H  RNO201563H  RNO201567H  RNO201654H  RNO201660H  RNO200575H  RNO200638H  RNO200960H  RNO200085H  RNO200908H  RNO200974H  RNO201726H  RNO200224H  RNO200730H  RNO201861H  RNO200139H  RNO200215H  RNO200394H  RNO201415H  RNO201787H  RNO201349H  RNO200121H  RNO200748H  RNO200649H  RNO200650H  RNO200651H  RNO200654H  RNO200655H  RNO200190H  RNO200421H  RNO200524H  RNO200600H  RNO201557H  RNO200035H  RNO200148H  RNO200181H  RNO200437H  RNO200482H  RNO200763H  RNO200201H  RNO200377H  RNO200461H  RNO200849H  RNO201465H  RNO200196H  RNO200062H  RNO200361H  RNO200440H  RNO200594H  RNO200633H  RNO201593H  RNO201599H  RNO201401H  RNO200027H  RNO200241H  RNO200242H  RNO200340H  RNO200692H  RNO200785H  RNO201719H  RNO200396H  RNO200425H  RNO200910H  RNO201073H  RNO201135H  RNO201164H  RNO201211H  RNO200946H  RNO200304H  RNO200395H  RNO200877H  RNO201007H  RNO201185H  RNO201344H  RNO201759H  RNO200512H  RNO200509H  RNO200728H  RNO200917H  RNO200919H  RNO200380H  RNO201125H  RNO201146H  RNO201802H  RNO200527H  RNO201030H  RNO200069H  RNO200530H  RNO200793H  RNO201436H  RNO201525H  RNO200711H  RNO200712H  RNO200713H  RNO200307H  RNO200879H  RNO200444H  RNO200465H  RNO201685H  RNO201686H  RNO200636H  RNO200648H  RNO200790H  RNO200866H  RNO200515H  RNO200570H  RNO200540H  RNO201165H  RNO201167H  RNO201754H  RNO200197H  RNO200844H  RNO201037H  RNO200462H  RNO200164H  RNO200053H  RNO200123H  RNO200237H  RNO200401H  RNO200402H  RNO200783H  RNO201319H  RNO201322H  RNO201352H  RNO201356H  RNO200735H  RNO200750H  RNO200752H  RNO200753H  RNO200762H  RNO200541H  RNO200984H  RNO200831H  RNO200441H  RNO201776H  RNO200435H  RNO200058H  RNO200423H  RNO200522H  RNO201364H  RNO201515H  RNO201314H  RNO201773H  RNO200506H  RNO200231H  RNO200247H  RNO200652H  RNO201697H  RNO201744H  RNO200189H  RNO200601H  RNO200784H  RNO201538H  RNO200349H  RNO201114H  RNO201115H  RNO201710H  RNO201140H  RNO200090H  RNO200091H  RNO200175H  RNO200388H  RNO200840H  RNO201728H  RNO200427H  RNO200723H  RNO200724H  RNO201126H  RNO201127H  RNO201452H  RNO201536H  RNO201556H  RNO201679H  RNO201823H  RNO200521H  RNO200557H  RNO200559H  RNO200553H  RNO200607H  RNO200956H  RNO200483H  RNO200666H  RNO200994H  RNO201083H  RNO201084H  RNO201085H  RNO201087H  RNO200044H  RNO200587H  RNO200758H  RNO201303H  RNO200492H  RNO200983H  RNO200922H  RNO200924H  RNO200857H  RNO200885H  RNO201171H  RNO201756H  RNO200439H  RNO200573H  RNO201548H  RNO200792H  RNO200928H  RNO201215H  RNO201522H  RNO201523H  RNO201532H  RNO201702H  RNO201703H  RNO201731H  RNO200684H  RNO200765H  RNO201080H  RNO201194H  RNO201315H  RNO201270H  RNO200178H  RNO200356H  RNO201163H  RNO200141H  RNO200182H  RNO200310H  RNO200353H  RNO200382H  RNO200384H  RNO200709H  RNO200710H  RNO201016H  RNO201154H  RNO201155H  RNO201159H  RNO201161H  RNO201695H  RNO201698H  RNO201699H  RNO201700H  RNO200133H  RNO201079H  RNO201470H  RNO201732H  RNO201027H  RNO201141H  RNO200185H  RNO200838H  RNO200909H  RNO200391H  RNO200160H  RNO200842H  RNO200832H  RNO201402H  RNO200627H  RNO201561H  RNO200311H  RNO200751H  RNO200787H  RNO200359H  RNO201249H  RNO201251H  RNO201326H  RNO200036H  RNO200420H  RNO200596H  RNO201143H  RNO201503H  RNO201704H  RNO201448H  RNO201003H  RNO201109H  RNO200163H  RNO200081H  RNO201031H  RNO200458H  RNO201213H  RNO201234H  RNO201477H  RNO200568H  RNO200358H  RNO200385H  RNO201212H  RNO201438H  RNO200916H  RNO201078H  RNO201305H  RNO201329H  RNO201856H  RNO201885H  RNO200535H  RNO200537H  RNO200606H  RNO201017H  RNO201388H  RNO200159H  RNO200350H  RNO200351H  RNO200407H  RNO200443H  RNO200620H  RNO200755H  RNO200248H  RNO200249H  RNO200519H  RNO200567H  RNO200622H  RNO200623H  RNO200624H  RNO200625H  RNO200699H  RNO200700H  RNO200701H  RNO201657H  RNO201658H  RNO201659H  RNO201663H  RNO201664H  RNO201666H  RNO201667H  RNO201669H  RNO201670H  RNO200229H  RNO201389H  RNO200579H  RNO200459H  RNO200631H  RNO200825H  RNO200951H  RNO200954H  RNO201045H  RNO201420H  RNO201447H  RNO201682H  RNO201689H  RNO201766H  RNO200592H  RNO200332H  RNO200372H  RNO200250H  RNO201825H  RNO200504H  RNO200630H  RNO201199H  RNO200581H  RNO201023H  RNO200418H  RNO201188H  RNO201467H  RNO201620H  RNO200246H  RNO201013H  RNO201131H  RNO200532H  RNO200542H  RNO200935H  RNO201417H  RNO200280H  RNO200839H  RNO201643H  RNO201644H  RNO201645H  RNO201648H  RNO201649H  RNO201650H  RNO201651H  RNO201652H  RNO201655H  RNO201656H  RNO200150H  RNO201403H  RNO201405H  RNO200749H  RNO200761H  RNO200590H  RNO200704H  RNO200718H  RNO200122H  RNO200134H  RNO200161H  RNO200165H  RNO200433H  RNO200923H  RNO201223H  RNO201450H  RNO201506H  RNO200613H  RNO201149H  RNO200629H  RNO201201H  RNO200065H  RNO200282H  RNO200283H  RNO200284H  RNO200286H  RNO200287H  RNO200288H  RNO200290H  RNO200291H  RNO200292H  RNO200293H  RNO200365H  RNO200369H  RNO200497H  RNO201850H  RNO201772H  RNO200745H  RNO201022H  RNO200024H  RNO200173H  RNO200469H  RNO200470H  RNO200471H  RNO200892H  RNO201203H  RNO201473H  RNO201741H  RNO200446H  RNO201461H  RNO201462H  RNO201494H  RNO201637H  RNO201638H  RNO200830H  RNO200558H  RNO201335H  RNO201391H  RNO201630H  RNO200128H  RNO200131H  RNO200543H  RNO200776H  RNO200932H  RNO201081H  RNO200136H  RNO200552H  RNO200940H  RNO200155H  RNO200702H  RNO201111H  RNO201112H  RNO201301H  RNO200464H  RNO200604H  RNO200501H  RNO200997H  RNO201472H  RNO201805H  RNO201832H  RNO201443H  RNO200833H  RNO201106H  RNO201492H  RNO200079H  RNO200432H  RNO201497H  RNO201562H  RNO201564H  RNO200220H  RNO200508H  RNO201191H  RNO200257H  RNO201454H  RNO201798H  RNO200814H  RNO200965H  RNO200802H  RNO201810H  RNO200117H  RNO200507H  RNO200474H  RNO200476H  RNO200489H  RNO200861H  RNO201032H  RNO200041H  RNO200239H  RNO200294H  RNO200456H  RNO200491H  RNO200434H  RNO200677H  RNO200975H  RNO200129H  RNO200135H  RNO200143H  RNO200193H  RNO200207H  RNO200534H  RNO200556H  RNO200572H  RNO200602H  RNO200746H  RNO200827H  RNO201100H  RNO201216H  RNO200240H  RNO200174H  RNO201355H  RNO200060H  RNO200719H  RNO200789H  RNO201382H  RNO200070H  RNO201321H  RNO200773H  RNO201711H  RNO200947H  RNO201222H  RNO200798H  RNO201012H  RNO200925H  RNO200998H  RNO201036H  RNO200056H  RNO200948H  RNO201096H  RNO201534H  RNO201796H  RNO200408H  RNO201807H  RNO200096H  RNO200204H  RNO200341H  RNO201495H  RNO200788H  RNO200791H  RNO201102H  RNO200068H  RNO200071H  RNO200889H  RNO200896H  RNO201584H  RNO200184H  RNO201688H  RNO200695H  RNO200034H  RNO201009H  RNO201010H  RNO201384H  RNO201840H  RNO201312H  RNO201485H  RNO201373H  RNO200656H  RNO200657H  RNO200658H  RNO200659H  RNO200660H  RNO200376H  RNO200918H  RNO201419H  RNO200449H  RNO200450H  RNO200451H  RNO200452H  RNO201672H  RNO201673H  RNO201674H  RNO201675H  RNO200168H  RNO200300H  RNO200301H  RNO200302H  RNO200303H  RNO200305H  RNO200306H  RNO200632H  RNO200634H  RNO200635H  RNO200639H  RNO200640H  RNO200641H  RNO200642H  RNO200643H  RNO200645H  RNO200680H  RNO200685H  RNO200689H  RNO200714H  RNO200715H  RNO200716H  RNO200717H  RNO200720H  RNO200721H  RNO200722H  RNO200847H  RNO200874H  RNO201062H  RNO201063H  RNO201162H  RNO201183H  RNO201193H  RNO201195H  RNO201196H  RNO201345H  RNO201351H  RNO201353H  RNO201360H  RNO201394H  RNO201395H  RNO201397H  RNO201464H  RNO201566H  RNO201568H  RNO201569H  RNO201571H  RNO201572H  RNO201573H  RNO201575H  RNO201576H  RNO201577H  RNO201578H  RNO201581H  RNO201583H  RNO201586H  RNO201588H  RNO201589H  RNO201590H  RNO201594H  RNO201595H  RNO201596H  RNO201597H  RNO201600H  RNO201601H  RNO201602H  RNO201604H  RNO201610H  RNO201611H  RNO201612H  RNO201614H  RNO201615H  RNO201616H  RNO201618H  RNO201623H  RNO201625H  RNO201626H  RNO201627H  RNO200853H  RNO200854H  RNO200863H  RNO201633H  RNO201836H  RNO201862H  RNO201863H  RNO201874H  RNO200991H  RNO201239H  RNO200891H  RNO200971H  RNO201342H  RNO201318H  RNO201622H  RNO201099H  RNO201646H  RNO201781H  RNO201951H  RNO201955H  RNO201957H  RNO201959H  RNO201961H  RNO201963H  RNO201964H  RNO201965H  RNO201975H  RNO201979H    (b)00643169891401  Serial Numbers:  RNO600230S  RNO600231S  RNO600232S  RNO600213S  RNO600221S  RNO600219S  RNO600228S  RNO600222S  RNO600210S  RNO600208S  RNO600207S  RNO600212S  RNO600214S  RNO600216S  RNO600226S  RNO600215S  RNO600209S  RNO600211S  RNO600229S  RNO600227S  RNO600225S  RNO600223S  RNO600224S  RNO600220S  RNO600217S  RNO600218S",
      "center_classification_date": "20180710",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "state": "MN",
      "product_description": "Percepta CRT-P MRI SureScan, REF W1TR01    Product Usage:  The CRT-P models are implantable pulse generators with cardiac resynchronization therapy (CRT-P). These devices monitor and regulate the patient s heart rate by providing single or dual chamber rate responsive bradycardia pacing, sequential biventricular pacing, and atrial tachyarrhythmia therapies. These devices are indicated for heart failure patients who may have atrial arrhythmias or AV block.",
      "report_date": "20180718",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "recall_number": "Z-2412-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80433",
      "termination_date": "20200819",
      "more_code_info": "",
      "recall_initiation_date": "20180621",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}