  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Loudon",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80293",
      "recalling_firm": "Fort Defiance Industries, LLC",
      "address_1": "2411 Maremont Pkwy",
      "address_2": "",
      "postal_code": "37774-1066",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide Distribution in the states of PA. VA. TX",
      "recall_number": "Z-2411-2018",
      "product_description": "FDI FORT DEFIANCE INDUSTRIES P2131 AUTOMATED FIELD STEAM STERILIZER, NSN: 6530-01-641-4641    Product Usage:  The Fort Defiance Industries Automated Field Steam Sterilizer Model P2131 is designed for sterilization of porous and non-porous, heat- and moisture-stable materials (e.g., surgical instruments and textiles) used in healthcare facilities. The P2131 sterilizer is a transportable device designed for use in a variety of austere environments.",
      "product_quantity": "199 units",
      "reason_for_recall": "During long-term storage (i.e. military depot), the interaction of stagnant water and dissimilar metals in the plumbing of the P2131 device produces deposits in the water system that can potentially interfere with the rear sight gauge and / or the water liquid level switch.",
      "recall_initiation_date": "20180524",
      "center_classification_date": "20180710",
      "termination_date": "20230627",
      "report_date": "20180718",
      "code_info": "Serial Numbers:  AFS-0001 through AFS-0199",
      "more_code_info": ""
    }
  ]
}