  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-06-24",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sylmar",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98963",
      "recalling_firm": "Abbott Medical",
      "address_1": "15900 Valley View Ct",
      "address_2": "N/A",
      "postal_code": "91342-3577",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of  AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, \tGA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, \tNY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. OUS: Not provided yet",
      "recall_number": "Z-2405-2026",
      "product_description": "Merlin\" Patient Care System (PCS) software  Merlin\" PCS 3650 programmer Model 3330 software",
      "product_quantity": "11182 programs",
      "reason_for_recall": "Software malfunction may occur during pacing capture threshold test of pacemaker resulting in  programmer not successfully communicating the command to terminate the test.",
      "recall_initiation_date": "20260512",
      "center_classification_date": "20260611",
      "report_date": "20260617",
      "code_info": "Programmer Model 3650; Software Model 3330;  GTIN 05414734509725; Software Versions 25.0.1 - 28.7.1 in the United States (varies based on region)"
    }
  ]
}