  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Branchburg",
      "address_1": "1080 Us Highway 202 S",
      "reason_for_recall": "When using a Laboratory Information system (LIS) with automated systems utilizing AMPLILINK software v3.2 series (AL v3.2) there is a potential for wrong results to be assigned to an order.",
      "address_2": "",
      "product_quantity": "1292 instruments",
      "code_info": "05666678001",
      "center_classification_date": "20120918",
      "distribution_pattern": "Nationwide (USA) Distribution",
      "state": "NJ",
      "product_description": "Amplilink (AL) Software 3.2.3 for use with COBAS AMPLICOR, CAP/CA, COBAS TaqMan,and COBAS AmpliPrep/COBAS TaqMan. The software can be used with a Laboratory Information System.",
      "report_date": "20120926",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Roche Molecular Systems, Inc.",
      "recall_number": "Z-2403-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62738",
      "termination_date": "20140212",
      "more_code_info": "",
      "recall_initiation_date": "20120423",
      "postal_code": "08876-3733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}