  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Vernon Hills",
      "address_1": "75 N Fairway Dr",
      "reason_for_recall": "Becton Dickinson and Company, is conducting a voluntary recall on the V. MuellerTM 3H Knife Handles catalog number listed in the table above. BD has been informed by our supplier that the external threads on some of the collets (Figure 1) contain a machining burr, potentially causing the collet to not fully seat in the knife handle (Figure 2), inhibiting complete grip of the inserted knife blade possibly resulting in a loose blade.",
      "address_2": "",
      "product_quantity": "358 units",
      "code_info": "Model Numbers: OP0702 and OP0703.  Model OP0702, Lot C20VAA ( 4 inch length)  Model OP0703, Lot D17VAA (6 inch length)",
      "center_classification_date": "20200618",
      "distribution_pattern": "US Nationwide distribution including in the states of CA, FL, VA and KY.",
      "state": "IL",
      "product_description": "Mueller Knife Handle, Hexagonal - Product Usage: Intended to hold surgical blades.",
      "report_date": "20200624",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Carefusion 2200 Inc",
      "recall_number": "Z-2400-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "85747",
      "termination_date": "20210425",
      "more_code_info": "",
      "recall_initiation_date": "20200518",
      "postal_code": "60061-1845",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}