{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88522",
      "recalling_firm": "DePuy Orthopaedics, Inc.",
      "address_1": "700 Orthopaedic Dr",
      "address_2": "N/A",
      "postal_code": "46582-3994",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution to US states of CA, CO, FL, IL, IN, NY, MI, MN, TX, and WA, Poland and UK.",
      "recall_number": "Z-2394-2021",
      "product_description": "TRUMATCH CT PIN GUIDE KIT L-  TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting  Part Number: 420578",
      "product_quantity": "4 cases (1/case)",
      "reason_for_recall": "Software:  A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.",
      "recall_initiation_date": "20210810",
      "center_classification_date": "20210901",
      "termination_date": "20230517",
      "report_date": "20210908",
      "code_info": "Lot Numbers: 27932 , 27806, 27515, 27466,  GTIN: 10603295430797"
    }
  ]
}