  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Twinsburg",
      "address_1": "1959 Summit Commerce Park",
      "reason_for_recall": "The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.",
      "address_2": "",
      "product_quantity": "33 units",
      "code_info": "Serial #'s: S5002-S5014; S5017-S5038",
      "center_classification_date": "20160811",
      "distribution_pattern": "US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.",
      "state": "OH",
      "product_description": "Hitachi Scenaria Computed Tomography System  Hitachi Medical Systems America, Twinsburg, OH 44087",
      "report_date": "20160817",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hitachi Medical Systems America Inc",
      "recall_number": "Z-2392-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74714",
      "termination_date": "20161101",
      "more_code_info": "",
      "recall_initiation_date": "20150630",
      "postal_code": "44087-2371",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}