{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "West Chester",
      "address_1": "1301 Goshen Pkwy",
      "reason_for_recall": "It was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft.  This resulted from the neck feature being too close to the StarDrive feature leaving a thin wall between neck and drive.",
      "address_2": "",
      "product_quantity": "41 units",
      "code_info": "Lot Number: 9942217  Part Number: 412.812",
      "center_classification_date": "20160811",
      "distribution_pattern": "US Distribution to states of: AR, CA, CO, DE, FL, GA, IA, IL, KY, LA, MN, NJ, OK, PA, SD, TN, and WA.",
      "state": "PA",
      "product_description": "Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone, Orthopedic Devices    The 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm is a part of the following systems: - 2.4. mm Titanium LCP Distal Radius Plate System - Titanium LCP Compact Distal Radius System - 2.4 mm Titanium LCP Radial Head Plate Set - Titanium Modular Mini Fragment LCP System - 2.4 mm Titanium Variable Angle LCP Distal Radius System - 2.7 mm/3.5 mm Titanium Distal Fibula Plate System",
      "report_date": "20160817",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Synthes (USA) Products LLC",
      "recall_number": "Z-2389-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74719",
      "termination_date": "20170202",
      "more_code_info": "",
      "recall_initiation_date": "20160628",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}