  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85719",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "",
      "postal_code": "19355",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2388-2020",
      "product_description": "MAMMOMAT Revelation, model no. 11343300 - Product Usage: is intended to be used for mammography exams, screening, diagnostics, biopsies and dual energy procedures under the supervision of medical professionals.",
      "product_quantity": "230 worldwide (64 US)",
      "reason_for_recall": "If an examination is interrupted when the InSpect Projection View (PV) is active and the vacuum biopsy system is being used, the tube arm's manual movement range is decreased from +/-15 deg. to +/- 6 deg. This limitation may lead to difficulties in removing the vacuum biopsy system.",
      "recall_initiation_date": "20200507",
      "center_classification_date": "20200616",
      "termination_date": "20220929",
      "report_date": "20200624",
      "code_info": "Serial Numbers (distributed in the US):    1026  1057  1348  1426  1027  1274  1292  1360  1324  1728  1372  1744  1036  1038  1620  1235  1751  1423  1379  1425  1342  1740  1445  1650  1058  1409  1277  1025  1253  1225  1749  1747  1422  1275  1228  1449  1707  1767  1387  1391  1206  1217  1250  1082  1357  1354  1256  1031  1339  1034  1219  1437  1289  1359  1045  1429  1408  1717  1051    ***Updated 8/18/2020***  1775  1784  1785  1795  1797",
      "more_code_info": ""
    }
  ]
}