  "results": [
    {
      "country": "United States",
      "city": "Cleveland",
      "address_1": "595 Miner Road",
      "reason_for_recall": "On August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.",
      "address_2": "",
      "product_quantity": "328 units",
      "code_info": "System Code #, 728131, Serial #: EP16E090001-EP16E090004, EP16E090006-EP16E090015, EP16E100001- EP16E100004, EP16E100006, EP16E100008- EP16E100063, EP16E100067, EP16E100069- EP16E100113, EP16E110001- EP16E110065, EP16E110067-EP16E110070, EP16E110074-EP16E110075, EP16E110077-EP16E110078, EP16E110080, EP16E110083- EP16E110085, EP16E110088- EP16E110112, EP16E110115- EP16E110119, EP16E120001- EP16E120014, EP16E120016, EP16EB110003- EP16ER100007, EP16ER110003, EP16ER110006, P16C080001, P16C080002, P16C090001, P16C100001-P16C100015, P16C100017, P16C100018, P16C110001- P16C110008, P16C110011-P16C110037, P16C120001-P16C120003, P16C120006, P16C120007, P16E080005, P16E080007, P16E080008, P16E080012, P16E090002-P16E090021, P16E090024-P16E090028, P16E100002, P16EF80010, P16EM08003, and P16EM09002-P16EM09004.    System Code #, 728132, Serial #: EP16E100005, EP16E100007, EP16E110066, EP16ER100003-EP16ER100005, EP16ER100008, EP16ER110004, EP16ER110005, and EP16ER110007-EP16ER110024.",
      "center_classification_date": "20120914",
      "distribution_pattern": "Worldwide distribution, Nationwide (US) including the states AL, AR, CA, FL, GA, IL, IN, MS, NJ, NV, NY, OH, PA, and VA and the countries of Albania, Argentina, Australia, Belgium, Brazil, Chile, China, Colombia, Dominican Republic, Ecuador, France, Germany, Greece, India, Indonesia, Ireland, Italy, Jordan, Lebanon, Malawi, Mexico, Nepal, Netherlands, Oman, Palestine, Philippines, Poland, Romania, Russia, Saudi Arabia, South Africa, South Korea, Spain, Switzerland, Turkey, Uganda, and Venezuela.",
      "state": "OH",
      "product_description": "MX 16-slice CT Scanner System, models 728131 and 728132. The MX 16-slice CT Scanner System can be used as a whole body computed tomography X-ray system.",
      "report_date": "20120926",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "recall_number": "Z-2386-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "63014",
      "termination_date": "20130828",
      "more_code_info": "",
      "recall_initiation_date": "20120824",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}