  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the polyethylene bag containing the implant. The black residue is a combination of the titanium plasma spray used to create the porous surface of the implant and polyethylene bag containing the implant. The recalling firm states the residue is biocompatible. Risks include delay of surgery while locating a replacement device, and surgical intervention due to wear caused by debris.",
      "address_2": "",
      "product_quantity": "101",
      "code_info": "00111778  60940196  61025425  61063904  61097911  61120383  61134445  61619526  62420837  62798037",
      "center_classification_date": "20180705",
      "distribution_pattern": "Products were distributed solely to Japan.",
      "state": "IN",
      "product_description": "Zimmer¿ M/L Taper Hip Prosthesis With Kinective¿ Technology Cementless, Size 22.5; 00-7713-022-00",
      "report_date": "20180711",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-2373-2018",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "80188",
      "termination_date": "20200528",
      "more_code_info": "",
      "recall_initiation_date": "20180530",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}