  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Independence",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "62926",
      "recalling_firm": "Lucero Medical LLC",
      "address_1": "6100 Oak Tree Blvd Ste 200",
      "address_2": "N/A",
      "postal_code": "44131-6914",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2368-2012",
      "product_description": "Enduramesh Corpectomy Spacer System    Product Usage: The Enduramesh is a vertebral body replacement system intended for use in the thoracic and lumbar spine (T1-S1) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma/fracture. The device is intended for use with either allograft or autograft. The device may be used with a cleared supplemental fixation system based on surgeon discretion and preference.",
      "product_quantity": "169",
      "reason_for_recall": "The firm voluntarily recalled the device after they discovered that the sterilization recommendations on the product insert did not pass sterilization validation.",
      "recall_initiation_date": "20120715",
      "center_classification_date": "20120911",
      "termination_date": "20130220",
      "report_date": "20120919",
      "code_info": "Part #-Lot # (s): 762.010L-110738; 762.013L-110738; 762.017L-60611, 90611, and 512171; 762.019L-60611, 90611, 100308, 1261901, and 20121901; 762.021L-60611, 90611, 100308,  110106, 20122101, and 20122102; 762.023L- 90611, 100209, 100218, 00706, 10709, 20122301, and 20122302; 762.025L-90611, 100218, 110106, 20122501, and 20122502; 762.027L-60611, 90611, 100706, and 20122071; 762.029L-90611, 1262901, and 90611; 762.031L-1263101; 762.033L-90611, 91002, and 1263301; 762.041L-90611 and 1264101; 762.043L-90611 and 1264301; 762.045L-90611; 762.047L-90611; 762.049L-90611; 762.051L-90611 and 512511; 762.053L-90611; 762.090L-90611; 762.090L-2.0-100206 and 100342; 762.090L-2.5-100206; 762.159L-110505; 762.237L-91002; 762.243L-91018; 762.270L-91018; 762.290L-91018; 762.731L-91018; 762.735L-91002; 762.767L-91018; and 762.790L-91018."
    }
  ]
}