  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83565",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution - AL, AR, KS, MI, MN, NC, ND, NY, OH, SC, TX,  WI",
      "recall_number": "Z-2367-2019",
      "product_description": "Artis zee and Artis Q/ Q.zen  Interventional Fluoroscopic X-Ray System with patient tables as follows:    Systems                   Material Number  Artis zee biplane      10094141;  Artis Q floor\t         10848280;  Artis Q biplane\t         10848282;  Artis Q.zen floor        10848353;  Artis Q.zen biplane   10848355;  Artis zee ceiling\t 10094137;  Artis Q ceiling           10848281;  Artis Q zeego           10848283  ArtisQ.zen ceiling     10848354    Product Usage:  Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. The Artis family can also support the acquisition of position triggered imaging for spatial data synthesis.",
      "product_quantity": "US 17",
      "reason_for_recall": "Artis zee/Q systems patient tables may potentially have cracks in the table mainframe, and  cause mechanical detachment of the upper part of the patient bed and lead to hazardous situations for patients and medical personnel due to falling parts",
      "recall_initiation_date": "20190730",
      "center_classification_date": "20190824",
      "termination_date": "20200429",
      "report_date": "20190904",
      "code_info": "Serial Numbers:  123245  103302  121501  148401  109589  148393  111129  148400  103310  148397  155235  121505  109563  121506  131235  121502  105125"
    }
  ]
}