  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Elyria",
      "address_1": "1 Invacare Way",
      "reason_for_recall": "The rear gradation scale was screen printed upside down on the flowmeter gauge.",
      "address_2": "",
      "product_quantity": "395 units",
      "code_info": "Serial numbers for Model IRCPF16AW:     16CF024370, 16CF024373, 16CF024376, 16CF024379, 16CF024382,16CF024368,16CF024369, 16CF024371,16CF024372, 16CF024374,16CF024375, 16CF024377,16CF024378, 16CF024380,16CF024381, 16CF037743.",
      "center_classification_date": "20160808",
      "distribution_pattern": "Worldwide Distribution - US to CA, CO, CT, FL, GA, IA, ID, IL,IN,KS,KY,LA,MA,MD,MI,MS,MT,NC,NE,NJ,NM,NY,OH,PA,SC,  TN,TX,UT,VA,WA and Internationally to Canada, Germany and New Zealand.",
      "state": "OH",
      "product_description": "Invacare Precise Rx Pediatric Flowmeter Model Number: IRCPF16 AW. Packed one per carton.  The lnvacare Precise Rx\" Pediatric Flowmeter adapts to lnvacare Concentrators for low-flow usage.The gauge has a scale printed on the front side and the back side. The front scale is the primary source for monitoring oxygen flow. The rear scale is an aide for monitoring oxygen flow.",
      "report_date": "20160817",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Invacare Corporation",
      "recall_number": "Z-2365-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74581",
      "termination_date": "20170207",
      "more_code_info": "",
      "recall_initiation_date": "20160621",
      "postal_code": "44035-4190",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}