{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Trumbull",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83110",
      "recalling_firm": "CooperSurgical, Inc.",
      "address_1": "95 Corporate Dr",
      "address_2": "N/A",
      "postal_code": "06611-1350",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in all 50 states. International distribution to Australia, Brazil, Canada, Ecuador, Germany, Gibraltar, India, Ireland, Kuwait, Macedonia, New Zealand, Oman, Poland, Puerto Rico, Qatar, Romania, Saudi Arabia, South Korea, Switzerland, United Arab Emirates, & United Kingdom.",
      "recall_number": "Z-2364-2019",
      "product_description": "TransWarmer Warming Infant Transport Mattress, Model no. 20421    Product  Usage:  The TransWarmer¿ Warming Infant Transport Mattress is gel-filled, disposable mattress that provides heat for up to two hours when cold stress is a concern. The patented WarmGel heats up to 104 (when starting from 75 ) in 60 seconds or less. The product is sold in boxes of 6 Infant Transport Mattresses per box. The intended use of the product is to provide warmth during infant transport within the hospital or between hospitals.",
      "product_quantity": "49813",
      "reason_for_recall": "The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.",
      "recall_initiation_date": "20190708",
      "center_classification_date": "20190823",
      "termination_date": "20200528",
      "report_date": "20190904",
      "code_info": "IJ780, IJ782, IJ785, IJ789, IJ794, IJ796, IJ804, IJ807, IJ817, IJ824, IJ832, IJ834, IJ841, IJ853, IJ865, IJ874, IJ878, IJ884, IJ886, IJ892, IJ895, IJ897, IJ900, IJ902, IJ909, IJ917, IJ920, IJ924, IJ933, IJ946, IJ948, IJ957, IJ962, IJ973, IJ976, IJ982, IJ985, IJ999, IK007, IK009, IK015, IK028, IK038, IK041, IK053, IK060, IK066, IK095, IK102, IK107, IK127, IK135, IK147, IK162, IK164, IK171, IK178, IK184, IK201, IK214, IK238, IK252, IK258, IK265, IK270, IK276, IK298, IK318, IK322, IK330, IK332, IK339, IK353, IK361, IK378, IK384, & IK390"
    }
  ]
}