  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Warsaw",
      "address_1": "1800 W Center St",
      "reason_for_recall": "M/L Taper Hip Prosthesis and M/L Taper Hip Prosthesis with Kinectiv Technology manufactured prior to October 2015 are being recalled due to the potential for black residue to be observed within the polyethylene bag containing the implant. The black residue is a combination of the titanium plasma spray used to create the porous surface of the implant and polyethylene bag containing the implant. The recalling firm states the residue is biocompatible. Risks include delay of surgery while locating a replacement device, and surgical intervention due to wear caused by debris.",
      "address_2": "",
      "product_quantity": "294",
      "code_info": "62107667  62410983  62410984  62445410  62449677  62450635  62469393  62478848  62478852  62478855  62503181  62547038  62554172  62559999  62560152  62577929  62630007  62636746  62652981  62659334  62667678  62753395  62770876  62777941  62791158  62806022  62897465",
      "center_classification_date": "20180705",
      "distribution_pattern": "Products were distributed solely to Japan.",
      "state": "IN",
      "product_description": "Zimmer¿ M/L Taper Hip Prosthesis Plasma Sprayed, Size 4 Standard Offset; 00-7711-004-10",
      "report_date": "20180711",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Zimmer Biomet, Inc.",
      "recall_number": "Z-2362-2018",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "80188",
      "termination_date": "20200528",
      "more_code_info": "",
      "recall_initiation_date": "20180530",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}