  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "San Diego",
      "address_1": "9975 Summers Ridge Rd",
      "reason_for_recall": "Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Monitoring Systems (Test Strips and Meters). This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR Monitoring System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.",
      "address_2": "",
      "product_quantity": "",
      "code_info": "All Lots  All Serial Numbers",
      "center_classification_date": "20160819",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and foreign distribution to the countries of: Argentina, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Brunei Daruss, Canada, Chile, China, Colombia, Egypt, France, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Malaysia, Maldives, Netherlands, Russia, Rwanda, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad, Ukraine, Uruguay, and Vietnam.",
      "state": "CA",
      "product_description": "Alere INRatio PT/INR Test Strips  Model Number: 0100071, 0100139    Product Usage:  The INRatio/INRatio2 Test Strips perform a modified version of the one-stage Prothrombin Time test, using a recombinant human thromboplastin reagent. The clot formed in the Prothrombin Time reaction is detected by a change in the electrical impedance of the sample during the coagulation process. The system consists of a monitor and disposable test strips. The monitor measures impedance, heats the test strip to the proper reaction temperature, and provides a user interface. The blood sample is applied to the Test Strip and the clotting reaction occurs on the Test Strip.",
      "report_date": "20160831",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "Alere San Diego, Inc.",
      "recall_number": "Z-2361-2016",
      "initial_firm_notification": "Press Release",
      "product_type": "Devices",
      "event_id": "74665",
      "termination_date": "20170720",
      "more_code_info": "",
      "recall_initiation_date": "20160711",
      "postal_code": "92121-2997",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}