      "status": "Ongoing",
      "city": "Blue Ash",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88434",
      "recalling_firm": "Ethicon Endo-Surgery Inc",
      "address_1": "4545 Creek Rd",
      "address_2": "N/A",
      "postal_code": "45242-2803",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Austria, Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czechia,  Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Ireland,  Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico,  Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden,  Switzerland, Taiwan, Thailand, Turkey, United  Kingdom, United Arab Emirates.",
      "recall_number": "Z-2355-2021",
      "product_description": "HARMONIC HD 1000i Shears (20cm Shaft Length)-indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired.  Product Code: HARHD20",
      "product_quantity": "7870 units",
      "reason_for_recall": "An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns",
      "recall_initiation_date": "20210720",
      "center_classification_date": "20210826",
      "report_date": "20210901",
      "code_info": "Lot Codes: U95126 U9571N U9526Z U95A80 U9543P U95T0X U9550G  GTIN: 10705036015048"
    }
  ]
}